Leadcare II Blood Lead Testing System

K052549510(k) SpecialCleared Oct 6, 200520-day reviewClass II21 CFR 862.3550Recalls 13 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Esa Biosciences, Inc. · DOF Lead, Atomic Absorption

Cleared on Oct 6, 2005 as substantially equivalent to K971640 (Esa, Inc., 1997) after a 20-day review. Its family tree reaches back 1 generation to 1997. 4 of the 4 devices in its family, counting itself, have a Class I recall. 2 later devices descend from it. It has been recalled 13 times, including 2 Class I recalls.

Decided
Oct 6, 2005
Review
20 days
Class
Class II
Product code
DOF
Recalls
13 (2 Class I)
Predicates
1

Its longest line of descent runs back 1 generation to K971640 (1997). 2 later devices descend from it, 1 citing it directly. 4 of the 4 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2013: 1 descendant cleared, 1 with a Class I recall12015: 1 descendant cleared, 1 with a Class I recall1K971640 · Leadcare Blood Lead Testing System · Esa, Inc. · 1997 · Class I recall1995200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 1 descendant, 1 with a Class I recall1 K971640 · Leadcare Blood Lead Testing System · Esa, Inc. · 1997 · Class I recall0
  1. Root · 1 generation backK971640 Leadcare Blood Lead Testing SystemEsa, Inc. · 1997 · Recalled (Class I)
  2. This deviceK052549 Leadcare II Blood Lead Testing SystemEsa Biosciences, Inc. · 2005
  3. Descendant · cites it directly · 1 of its ownK123563 Leadcare (R) Ultra (TM) Blood Lead Testing SystemMagellan Diagnostics · 2013 · Recalled (Class I)
  4. Descendant · 2 generations on · 0 of its ownK142705 LeadCare Plus Blood Lead Testing SystemMagellan Diagnostics · 2015 · Recalled (Class I)

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K123563Leadcare (R) Ultra (TM) Blood Lead Testing SystemMagellan Diagnostics · 2013 · cites it directly1descendant
  • K142705LeadCare Plus Blood Lead Testing SystemMagellan Diagnostics · 2015 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K971640Leadcare Blood Lead Testing SystemEsa, Inc.1997Class I
0 (this device)K052549Leadcare II Blood Lead Testing SystemEsa Biosciences, Inc.2005—
+1K123563Leadcare (R) Ultra (TM) Blood Lead Testing SystemMagellan Diagnostics2013Class I
+2K142705LeadCare Plus Blood Lead Testing SystemMagellan Diagnostics2015Class I

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201311
201511
FIG. 1 — Lineage of K052549, 1997–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K971640Leadcare Blood Lead Testing System
Summary page 1
…Substantial Equivalence to ESA Bioscience Inc. (formerly ESA Umed Saee Inc.) LeadCare® Blood Lead Testing System. (K971640) (800) 235 2436 Fax (785) 832 0406 wwwcypresssystemscorm 4. Device Description…
Esa, Inc.1997Class I

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K123563Leadcare (R) Ultra (TM) Blood Lead Testing System Class IMagellan Diagnostics20131

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 4, 2026Class IIIZ-0162-2027Labeling Change ControlIn kit sub-lot 2530M-01 the date of expiration printed on the outer labeling may indicate the wrong expiration date of July 22, 2026 while the correct expiration date is July 13, 2026. If used the system will display an error message and…
Mar 13, 2026Class IIZ-1893-2026Device DesignCustomers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed…
Mar 13, 2026Class IIZ-1895-2026Device DesignCustomers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed…
Feb 23, 2024Class IIZ-0584-2026Labeling mix-upsCalibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.
Sep 19, 2023Class IIZ-0219-2024Nonconforming Material/ComponentDuring an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with…
May 25, 2022Class IIZ-1597-2022Environmental controlFacility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
May 7, 2021Class IZ-1952-2021Nonconforming Material/ComponentMagellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.
Aug 8, 2018Class IIZ-0285-2019Process controlDefect with the calibration button for lot 1808M. Level 1 and Level 2 Quality Control (QC) values will generate results of HIGH (>65 g/dL). Results for patient testing with this lot will be displayed as HIGH.
Jan 9, 2018Class IIZ-1295-2018Device DesignThe incorrect placement of AA batteries into certain LeadCare II Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.
May 23, 2017Class IZ-2226-2017Device DesignUnderestimates the lead concentration of venous blood samples when the sample is analyzed with the LeadCare systems
Jan 20, 2016Class IIZ-3035-2017Device DesignLeadCare II Test Kits controls out of range
Sep 20, 2007Not classifiedZ-0042-2008Process designLevel 1 and Level 2 Control values were incorrectly assigned.
Sep 20, 2007Not classifiedZ-0043-2008Process designLevel 1 and Level 2 Control values were incorrectly assigned.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K971640Leadcare Blood Lead Testing SystemEsa, Inc. · 1997 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K123563Leadcare (R) Ultra (TM) Blood Lead Testing SystemMagellan Diagnostics · 2013 · Cites it directly · ◆ in the 2013 descendants bar of FIG. 12 Class I recalls
  • K142705LeadCare Plus Blood Lead Testing SystemMagellan Diagnostics · 2015 · 2 generations on · ◆ in the 2015 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Its one ancestor has a recall, Class I.
  • Of its descendants, K123563 (Magellan Diagnostics, 2013) has gone furthest, with 1 descendant of its own.

How does its category compare?

All DOF devices

DOF Lead, Atomic Absorption: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DOF

1in 2015

2013: 12015: 1
2007high 1 (2013) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DOF

Too few DOF 510(k)s in any year since 2007 to give a median.

Recalls per year · DOF

4so far in 2026, the most since 2007

2007: 22018: 12023: 12024: 12026 (partial year): 4
2007high 4 (2026) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DOF

55in 2025

2008: 12010: 12013: 12015: 12017: 392018: 402019: 632020: 22021: 982022: 12025: 55
2007high 98 (2021) · low 0 (2007)2025

MAUDE reports naming product code DOF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—20
20080—01
20090—00
20100—01
20110—00
20120—00
20131—01
20140—00
20151—01
20160—00
20170—039
20180—140
20190—063
20200—02
20210—098
20220—01
20230—10
20240—10
20250—055
20260—4—

What is it indicated for?

For In Vitro Diagnostic Use Only

The LeadCare® II Blood Lead Testing System is an instrumented assay to be used in the quantitation of lead in human whole blood. The Leadcare® II System is suitable for use in a physician's office laboratory environment (POL).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K052549: Leadcare II Blood Lead Testing System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K052549

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026