LeadCare Plus Blood Lead Testing System

K142705510(k) TraditionalCleared Jul 7, 2015288-day reviewClass II21 CFR 862.3550Recalls 4 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Magellan Diagnostics · DOF Lead, Atomic Absorption

Cleared on Jul 7, 2015 as substantially equivalent to K123563 (Magellan Diagnostics, 2013) after a 288-day review. Its family tree reaches back 3 generations to 1997. 4 of the 4 devices in its family, counting itself, have a Class I recall. It has been recalled 4 times, including 2 Class I recalls.

Decided
Jul 7, 2015
Review
288 days
Class
Class II
Product code
DOF
Recalls
4 (2 Class I)
Predicates
1

Its longest line of descent runs back 3 generations to K971640 (1997). No later device cites it as a predicate yet. 4 of the 4 devices in this family, counting this one, have a Class I recall.

  1. Root · 3 generations backK971640 Leadcare Blood Lead Testing SystemEsa, Inc. · 1997 · Recalled (Class I)
  2. 2 generations backK052549 Leadcare II Blood Lead Testing SystemEsa Biosciences, Inc. · 2005 · Recalled (Class I)
  3. 1 generation backK123563 Leadcare (R) Ultra (TM) Blood Lead Testing SystemMagellan Diagnostics · 2013 · Recalled (Class I)
  4. This deviceK142705 LeadCare Plus Blood Lead Testing SystemMagellan Diagnostics · 2015

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K971640Leadcare Blood Lead Testing SystemEsa, Inc.1997Class I
−2K052549Leadcare II Blood Lead Testing SystemEsa Biosciences, Inc.2005Class I
−1K123563Leadcare (R) Ultra (TM) Blood Lead Testing SystemMagellan Diagnostics2013Class I
0 (this device)K142705LeadCare Plus Blood Lead Testing SystemMagellan Diagnostics2015—
FIG. 1 — Lineage of K142705, 1997–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K123563Leadcare (R) Ultra (TM) Blood Lead Testing System
Summary page 6
…ellandx.com    Predicate Device: K123563 LeadCare Ultra® Substantial Equivalence Information: The LeadCare Plus Blood Lead Testing Sy…
Magellan Diagnostics2013Class I

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 13, 2026Class IIZ-1894-2026Device DesignCustomers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed…
May 7, 2021Class IZ-1953-2021Nonconforming Material/ComponentMagellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.
Jan 9, 2018Class IIZ-1296-2018Device DesignThe incorrect placement of AA batteries into certain LeadCare Plus Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.
Nov 29, 2014Class IZ-1926-2017Device DesignLeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K123563Leadcare (R) Ultra (TM) Blood Lead Testing SystemMagellan Diagnostics · 2013 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
  • K052549Leadcare II Blood Lead Testing SystemEsa Biosciences, Inc. · 2005 · 2 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
  • K971640Leadcare Blood Lead Testing SystemEsa, Inc. · 1997 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 18 years and passes through 3 companies; a new generation was cleared every 7.9 years on median.
  • 3 of the 3 ancestors have a recall of any class; 3 have a Class I recall.

How does its category compare?

All DOF devices

DOF Lead, Atomic Absorption: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · DOF

1in 2015

2013: 12015: 1
2007high 1 (2013) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DOF

Too few DOF 510(k)s in any year since 2007 to give a median.

Recalls per year · DOF

4so far in 2026, the most since 2007

2007: 22018: 12023: 12024: 12026 (partial year): 4
2007high 4 (2026) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DOF

55in 2025

2008: 12010: 12013: 12015: 12017: 392018: 402019: 632020: 22021: 982022: 12025: 55
2007high 98 (2021) · low 0 (2007)2025

MAUDE reports naming product code DOF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—20
20080—01
20090—00
20100—01
20110—00
20120—00
20131—01
20140—00
20151—01
20160—00
20170—039
20180—140
20190—063
20200—02
20210—098
20220—01
20230—10
20240—10
20250—055
20260—4—

What is it indicated for?

The LeadCare® Plus™ Blood Lead Testing System is intended for the quantitative measurement of lead in a whole blood sample. The LeadCare Plus Blood Lead Testing System is intended for in vitro (external) use only. The test kit components are for use with both the LeadCare Plus and LeadCare Ultra® Blood Lead Testing Systems.

This test system is for prescription use only. This system is not intended for point of care use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K142705: LeadCare Plus Blood Lead Testing System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K142705

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026