Magellan Diagnostics · DOF Lead, Atomic Absorption
Cleared on Jul 7, 2015 as substantially equivalent to K123563 (Magellan Diagnostics, 2013) after a 288-day review. Its family tree reaches back 3 generations to 1997. 4 of the 4 devices in its family, counting itself, have a Class I recall. It has been recalled 4 times, including 2 Class I recalls.
- Decided
- Jul 7, 2015
- Review
- 288 days
- Class
- Class II
- Product code
- DOF
- Recalls
- 4 (2 Class I)
- Predicates
- 1
Its longest line of descent runs back 3 generations to K971640 (1997). No later device cites it as a predicate yet. 4 of the 4 devices in this family, counting this one, have a Class I recall.
- Root · 3 generations backK971640 Leadcare Blood Lead Testing SystemEsa, Inc. · 1997 · Recalled (Class I)
- 2 generations backK052549 Leadcare II Blood Lead Testing SystemEsa Biosciences, Inc. · 2005 · Recalled (Class I)
- 1 generation backK123563 Leadcare (R) Ultra (TM) Blood Lead Testing SystemMagellan Diagnostics · 2013 · Recalled (Class I)
- This deviceK142705 LeadCare Plus Blood Lead Testing SystemMagellan Diagnostics · 2015
Class I recalls in this family
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K971640 | Leadcare Blood Lead Testing System | Esa, Inc. | 1997 | Class I |
| −2 | K052549 | Leadcare II Blood Lead Testing System | Esa Biosciences, Inc. | 2005 | Class I |
| −1 | K123563 | Leadcare (R) Ultra (TM) Blood Lead Testing System | Magellan Diagnostics | 2013 | Class I |
| 0 (this device) | K142705 | LeadCare Plus Blood Lead Testing System | Magellan Diagnostics | 2015 | — |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K123563 | Leadcare (R) Ultra (TM) Blood Lead Testing SystemSummary page 6…ellandx.com Predicate Device: K123563 LeadCare Ultra® Substantial Equivalence Information: The LeadCare Plus Blood Lead Testing Sy… | 2013 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (4 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Mar 13, 2026 | Class II | Z-1894-2026 | Device Design | Customers reported observing unexpectedly elevated results when using certain third-party micro-collection devices to collect capillary samples for testing with the LeadCare Test Systems. False positive lead results may cause delayed… |
| May 7, 2021 | Class I | Z-1953-2021 | Nonconforming Material/Component | Magellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples. |
| Jan 9, 2018 | Class II | Z-1296-2018 | Device Design | The incorrect placement of AA batteries into certain LeadCare Plus Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot. |
| Nov 29, 2014 | Class I | Z-1926-2017 | Device Design | LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the… |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K123563Leadcare (R) Ultra (TM) Blood Lead Testing SystemMagellan Diagnostics · 2013 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
- K052549Leadcare II Blood Lead Testing SystemEsa Biosciences, Inc. · 2005 · 2 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
- K971640Leadcare Blood Lead Testing SystemEsa, Inc. · 1997 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 18 years and passes through 3 companies; a new generation was cleared every 7.9 years on median.
- 3 of the 3 ancestors have a recall of any class; 3 have a Class I recall.
How does its category compare?
All DOF devicesDOF Lead, Atomic Absorption: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.
FDA decisions per year · DOF
1in 2015
2026 to Oct 25, outlined.
Median 510(k) review, days · DOF
Too few DOF 510(k)s in any year since 2007 to give a median.
Recalls per year · DOF
4so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · DOF
55in 2025
MAUDE reports naming product code DOF, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 2 | 0 |
| 2008 | 0 | — | 0 | 1 |
| 2009 | 0 | — | 0 | 0 |
| 2010 | 0 | — | 0 | 1 |
| 2011 | 0 | — | 0 | 0 |
| 2012 | 0 | — | 0 | 0 |
| 2013 | 1 | — | 0 | 1 |
| 2014 | 0 | — | 0 | 0 |
| 2015 | 1 | — | 0 | 1 |
| 2016 | 0 | — | 0 | 0 |
| 2017 | 0 | — | 0 | 39 |
| 2018 | 0 | — | 1 | 40 |
| 2019 | 0 | — | 0 | 63 |
| 2020 | 0 | — | 0 | 2 |
| 2021 | 0 | — | 0 | 98 |
| 2022 | 0 | — | 0 | 1 |
| 2023 | 0 | — | 1 | 0 |
| 2024 | 0 | — | 1 | 0 |
| 2025 | 0 | — | 0 | 55 |
| 2026 | 0 | — | 4 | — |
What is it indicated for?
The LeadCare® Plus™ Blood Lead Testing System is intended for the quantitative measurement of lead in a whole blood sample. The LeadCare Plus Blood Lead Testing System is intended for in vitro (external) use only. The test kit components are for use with both the LeadCare Plus and LeadCare Ultra® Blood Lead Testing Systems.
This test system is for prescription use only. This system is not intended for point of care use.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K142705 · printed Nov 6, 2026
Cite this page
510(k) Report. “K142705: LeadCare Plus Blood Lead Testing System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K142705
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026