C2000 Automated Manifold Kit

K052744510(k) SpecialCleared Dec 16, 200577-day reviewClass II21 CFR 870.1650Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Acist Medical Systems, Inc. · DXT Injector And Syringe, Angiographic

Cleared on Dec 16, 2005 as substantially equivalent to K984231 (Invasatec, 2000) after a 77-day review. Its family tree reaches back 3 generations to 1991. 14 later devices descend from it. It has been recalled once.

Decided
Dec 16, 2005
Review
77 days
Class
Class II
Product code
DXT
Recalls
1
Predicates
1

Its longest line of descent runs back 3 generations to K903493 (1991). 5 ancestors in all, the oldest K822536 (1982). 14 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1997: 1 other ancestor1982: 1 other ancestor2013: 1 descendant cleared2014: 1 descendant cleared2015: 4 descendants cleared2016: 2 descendants cleared2018: 2 descendants cleared2017: 3 descendants cleared2019: 1 descendant clearedK903493 · Namic Contrast Savings Delivery System · North American Instrument Corp. · 1991K961794 · Fluid Administration Set - Contrast Manager · Merit Medical Systems, Inc. · 1996K984231 · Angiographic Contrast Management System, Model Part a and Part B · Invasatec · 2000 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 4 descendants4 generations forward: 1 descendant5 generations forward: 1 descendant6 generations forward: 3 descendants7 generations forward: 3 descendants K984231 · Angiographic Contrast Management System, Model Part a and Part B · Invasatec · 2000 · recalledK961794 · Fluid Administration Set - Contrast Manager · Merit Medical Systems, Inc. · 1996K903493 · Namic Contrast Savings Delivery System · North American Instrument Corp. · 19910+5
  1. Root · 3 generations backK903493 Namic Contrast Savings Delivery SystemNorth American Instrument Corp. · 1991
  2. 2 generations backK961794 Fluid Administration Set - Contrast ManagerMerit Medical Systems, Inc. · 1996
  3. 1 generation backK984231 Angiographic Contrast Management System, Model Part a and Part BInvasatec · 2000 · Recalled
  4. This deviceK052744 C2000 Automated Manifold KitAcist Medical Systems, Inc. · 2005
  5. Descendant · cites it directly · 13 of its ownK131478 Contrast Conservation SystemOsprey Medical, Inc. · 2013
  6. Descendant · 2 generations on · 12 of its ownK140425 Contrast Modulation SystemOsprey Medical, Inc. · 2014
  7. Descendant · 3 generations on · 10 of its ownK150485 AVERT Contrast Modulation SystemOsprey Medical, Inc. · 2015
  8. Descendant · 3 generations on · 8 of its ownK151300 AVERT Contrast Modulation SystemOsprey Medical, Inc. · 2015
  9. Descendant · 3 generations on · 8 of its ownK151746 Dyevert Contrast Modulation SystemOsprey Medical, Inc. · 2015

Showing 5 of 14 descendants. See all descendants

Descendants cleared per year

2013: 12014: 12015: 42016: 22017: 32018: 22019: 1
2013high 4 (2015) · low 0 (2020)2026

Most influential descendants

Showing 5 of 14 · See all
  • K131478Contrast Conservation SystemOsprey Medical, Inc. · 2013 · cites it directly13descendants
  • K140425Contrast Modulation SystemOsprey Medical, Inc. · 2014 · 2 generations on12descendants
  • K150485AVERT Contrast Modulation SystemOsprey Medical, Inc. · 2015 · 3 generations on10descendants
  • K151300AVERT Contrast Modulation SystemOsprey Medical, Inc. · 2015 · 3 generations on8descendants
  • K151746Dyevert Contrast Modulation SystemOsprey Medical, Inc. · 2015 · 3 generations on8descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K903493Namic Contrast Savings Delivery SystemNorth American Instrument Corp.1991—
−2K961794Fluid Administration Set - Contrast ManagerMerit Medical Systems, Inc.1996—
−1K984231Angiographic Contrast Management System, Model Part a and Part BInvasatec20004
0 (this device)K052744C2000 Automated Manifold KitAcist Medical Systems, Inc.2005—
+1K131478Contrast Conservation SystemOsprey Medical, Inc.2013—
+2K140425Contrast Modulation SystemOsprey Medical, Inc.2014—
+3K150485AVERT Contrast Modulation SystemOsprey Medical, Inc.2015—
+3K151300AVERT Contrast Modulation SystemOsprey Medical, Inc.2015—
+3K151746Dyevert Contrast Modulation SystemOsprey Medical, Inc.2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198210
199710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201410
201540
201620
201730
201820
201910
FIG. 1 — Lineage of K052744, 1991–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DXT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DXT device

  1. #1 K253048 High Pressure Syringe; Pressure Connecting…score 13.6 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K984231Angiographic Contrast Management System, Model Part a and Part B
Summary page 1
…ially, equivalent to thei predicate device, ACIST Anogroaphic Coiuuaat Management System (cleared under K98423 1) This conclusion is based uponi the similarities of the devices in terms of functional design, i…
Invasatec20004

Also cleared against K984231: 1 other device →

What descends from it? (14)

See all 14

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K131478Contrast Conservation SystemOsprey Medical, Inc.201313

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 21, 2015Class IIZ-0744-2016Nonconforming Material/ComponentReports related to ingress of air into the manifold.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 14 years and passes through 4 companies; a new generation was cleared every 5.4 years on median.
  • 1 of the 5 ancestors has a recall of any class.
  • Of its descendants, K131478 (Osprey Medical, Inc., 2013) has gone furthest, with 13 descendants of its own.
  • 1 other device also cites its predicate K984231.

How does its category compare?

All DXT devices

DXT Injector And Syringe, Angiographic: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DXT

1in 2025

2007: 32008: 22009: 62010: 12011: 42013: 22014: 62015: 72016: 82017: 72018: 92019: 42020: 32021: 12023: 12024: 42025: 12026 (partial year): 7
2007high 9 (2018) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DXT

138in 2024

2007: 182 days median2009: 137 days median2011: 69 days median2014: 144 days median2015: 153 days median2016: 172 days median2017: 183 days median2018: 93 days median2019: 96 days median2020: 154 days median2024: 138 days median2026 (partial year): 264 days median
2007high 264 (2026) · low 69 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DXT

10in 2024

2007: 42008: 32009: 72010: 32011: 32012: 72015: 12018: 22020: 12021: 12023: 22024: 102026 (partial year): 6
2007high 10 (2024) · low 0 (2013)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DXT

97in 2025

2007: 1072008: 1682009: 1882010: 1242011: 1332012: 1352013: 1412014: 972015: 1352016: 1522017: 1282018: 1292019: 1212020: 932021: 1322022: 1112023: 1172024: 1182025: 97
2007high 188 (2009) · low 93 (2020)2025

MAUDE reports naming product code DXT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731824107
20082—3168
200961377188
20101—3124
20114693133
20120—7135
20132—0141
20146144097
201571531135
201681720152
201771830128
20189932129
20194960121
20203154193
20211—1132
20220—0111
20231—2117
2024413810118
20251—097
202672646—

What is it indicated for?

The ACIST Angiographic Injection System is intended to be used for the controlled infusion of radiopaque contrast media for angiographic procedures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K052744: C2000 Automated Manifold Kit.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K052744

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026