510(k) Report

510(k) · Cleared

Recalled · 2008

Bipap Focus

K053168510(k) TraditionalCleared Jan 4, 200651-day reviewClass II21 CFR 868.5895Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Respironics (Ireland) Limited · MNS Ventilator, Continuous, Non-Life-Supporting

Cleared on Jan 4, 2006 as substantially equivalent to K982454 (Respironics, Inc., 1998) and 2 other predicates after a 51-day review. Its family tree reaches back 4 generations to 1983. 11 later devices descend from it. It has been recalled once.

Decided
Jan 4, 2006
Review
51 days
Class
Class II
Product code
MNS
Recalls
1
Predicates
3

Its longest line of descent runs back 4 generations to K833786 (1983). 15 ancestors in all, the oldest K811102 (1981). 11 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor1999: 1 other ancestor2001: 2 other ancestors1997: 1 other ancestor2000: 2 other ancestors1995: 1 other ancestor1991: 1 other ancestor1988: 1 other ancestor1981: 1 other ancestor2008: 1 descendant cleared2010: 1 descendant cleared2012: 2 descendants cleared2009: 1 descendant cleared2018: 1 descendant cleared2020: 2 descendants cleared2013: 1 descendant cleared2014: 1 descendant cleared2023: 1 descendant clearedK833786 · Microprocessor Ventilator 7200 · Puritan Bennett Corp. · 1983×2: 2 generations cleared close together, K951264 (1996), K955324 (1996) K982454 · Bipap Vision Ventilatory Support · Respironics, Inc. · 1998 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 3 other ancestors3 generations back: 6 other ancestors 1 generation forward: 1 descendant2 generations forward: 3 descendants3 generations forward: 5 descendants4 generations forward: 1 descendant5 generations forward: 1 descendant K982454 · Bipap Vision Ventilatory Support · Respironics, Inc. · 1998 · recalledK955324 · Bipap S/T-D 30 System · Respironics, Inc. · 1996K951264 · Bipap Model S/T-D · Respironics, Inc. · 1996K833786 · Microprocessor Ventilator 7200 · Puritan Bennett Corp. · 19830+5
  1. Root · 4 generations backK833786 Microprocessor Ventilator 7200Puritan Bennett Corp. · 1983
  2. 3 generations backK951264 Bipap Model S/T-DRespironics, Inc. · 1996
  3. 2 generations backK955324 Bipap S/T-D 30 SystemRespironics, Inc. · 1996
  4. 1 generation backK982454 Bipap Vision Ventilatory SupportRespironics, Inc. · 1998 · Recalled
  5. This deviceK053168 Bipap FocusRespironics (Ireland) Limited · 2006
  6. Descendant · cites it directly · 10 of its ownK071560 Provent Nasal Dilator, Model BR2Ventus Medical, Inc. · 2008
  7. Descendant · 2 generations on · 5 of its ownK111549 Attune Sleep Apnea SystemApnicure, Inc. · 2012
  8. Descendant · 3 generations on · 3 of its ownK122130 Winx Sleep Therapy SystemApnicure, Inc. · 2012
  9. Descendant · 2 generations on · 3 of its ownK090398 Provent Professional Sleep Apnea TherapyVentus Medical, Inc. · 2009 · Recalled
  10. Descendant · 3 generations on · 2 of its ownK130538 Winx Sleep Therapy SystemApnicure, Inc. · 2013

Showing 5 of 11 descendants. See all descendants

Descendants cleared per year

2008: 12009: 12010: 12012: 22013: 12014: 12018: 12020: 22023: 1
2008high 2 (2012) · low 0 (2011)2026

Most influential descendants

Showing 5 of 11 · See all
  • K071560Provent Nasal Dilator, Model BR2Ventus Medical, Inc. · 2008 · cites it directly10descendants
  • K111549Attune Sleep Apnea SystemApnicure, Inc. · 2012 · 2 generations on5descendants
  • K122130Winx Sleep Therapy SystemApnicure, Inc. · 2012 · 3 generations on3descendants
  • K090398Provent Professional Sleep Apnea TherapyVentus Medical, Inc. · 2009 · 2 generations on3descendants
  • K130538Winx Sleep Therapy SystemApnicure, Inc. · 2013 · 3 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K833786Microprocessor Ventilator 7200Puritan Bennett Corp.1983—
−3K951264Bipap Model S/T-DRespironics, Inc.1996—
−2K955324Bipap S/T-D 30 SystemRespironics, Inc.1996—
−1K982454Bipap Vision Ventilatory SupportRespironics, Inc.19982
0 (this device)K053168Bipap FocusRespironics (Ireland) Limited2006—
+1K071560Provent Nasal Dilator, Model BR2Ventus Medical, Inc.2008—
+2K111549Attune Sleep Apnea SystemApnicure, Inc.2012—
+3K122130Winx Sleep Therapy SystemApnicure, Inc.2012—
+2K090398Provent Professional Sleep Apnea TherapyVentus Medical, Inc.20091
+3K130538Winx Sleep Therapy SystemApnicure, Inc.2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198810
199110
199510
199710
199910
200020
200120
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
200910
201010
201220
201310
201410
201810
202020
202310
FIG. 1 — Lineage of K053168, 1983–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MNS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MNS device

  1. #1 K251661 Mariana Minerva; Mariana Nimbusscore 14.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K982454Bipap Vision Ventilatory Support
Summary page 4
…y use (MNT) Device: Model - BiPAP Focus Predicate Devices: Respironics - Vision - K982454 Harmony K984407 and K031656 Device Description: The BiPAP Focus is a non-invasive ventilat…
Respironics, Inc.19982
*K984407Bipap Harmony S/T, Model 1001445
Summary page 4
…ice: Model - BiPAP Focus Predicate Devices: Respironics - Vision - K982454 Harmony K984407 and K031656 Device Description: The BiPAP Focus is a non-invasive ventilator which has two (2) ventilation modes: 1. Conti…
Respironics, Inc.1999—
*K031656Bipap Harmony Ventilatory Support System
Summary page 4
…Model - BiPAP Focus Predicate Devices: Respironics - Vision - K982454 Harmony K984407 and K031656 Device Description: The BiPAP Focus is a non-invasive ventilator which has two (2) ventilation modes: 1. Continuous positi…
Respironics, Inc.2003—

Also cleared against K982454: 6 other devices →

What descends from it? (11)

See all 11

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K071560Provent Nasal Dilator, Model BR2Ventus Medical, Inc.200810

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 22, 2008Not classifiedZ-1606-2009Component design/selectionPower Supply failures have occurred on some ventilators. A discrepancy between the power supply strain relief clip and the power cord connector may allow unwanted movement of the cord and the unit's power supply. A disconnection or…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 3 companies; a new generation was cleared every 7.2 years on median.
  • 3 of the 15 ancestors have a recall of any class.
  • Of its descendants, K071560 (Ventus Medical, Inc., 2008) has gone furthest, with 10 descendants of its own.
  • 6 other devices also cite its predicate K982454.

How does its category compare?

All MNS devices

MNS Ventilator, Continuous, Non-Life-Supporting: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MNS

1in 2024

2007: 42009: 42010: 42011: 32012: 52013: 32016: 22017: 12018: 12022: 12023: 12024: 12026 (partial year): 3
2007high 5 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MNS

115in 2013

2007: 90 days median2009: 92 days median2010: 124 days median2011: 140 days median2012: 111 days median2013: 115 days median2026 (partial year): 269 days median
2007high 269 (2026) · low 90 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MNS

1in 2008

2007: 22008: 12026 (partial year): 1, 1 Class I
2007high 2 (2007) · low 0 (2009)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · MNS

1,790in 2025

2007: 92008: 72009: 82010: 92011: 92012: 122013: 182014: 52015: 502016: 162017: 202018: 222019: 2752020: 642021: 2562022: 2,5142023: 1,3202024: 1,5712025: 1,790
2007high 2,514 (2022) · low 5 (2014)2025

MAUDE reports naming product code MNS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200749029
20080—17
200949208
2010412409
2011314009
20125111012
20133115018
20140—05
20150—050
20162—016
20171—020
20181—022
20190—0275
20200—064
20210—0256
20221—02,514
20231—01,320
20241—01,571
20250—01,790
202632691 (1)—

What is it indicated for?

The BiPAP Focus ventilator is a non-invasive, pressure support ventilator used to augment the breathing of patients suffering from acute or chronic respiratory insufficiency, or respiratory failure, or to maintain airway patency and provide ventilatory support to patients who experience obstructive sleep apnea. It is not intended to provide the total ventilatory requirements of the patient.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K053168: Bipap Focus.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K053168

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026