510(k) Report

510(k) · Cleared

Recalled · Class I · 2019

Dual-Sided Interhemispheric Subdural Electrodes, Grid Subdural Electrodes, Intraoperative Subdural Electrodes

K053363510(k) AbbreviatedCleared May 30, 2006179-day reviewClass II21 CFR 882.1310Recalls 2 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ad-Tech Medical Instrument Corp · GYC Electrode, Cortical

Cleared on May 30, 2006 as substantially equivalent to K970587 (Ad-Tech Medical Instrument Corp, 1997) and 2 other predicates after a 179-day review. Its family tree reaches back 2 generations to 1985. 1 of the 14 devices in its family, counting itself, has a Class I recall. 9 later devices descend from it. It has been recalled 2 times, including a Class I recall.

Decided
May 30, 2006
Review
179 days
Class
Class II
Product code
GYC
Recalls
2 (1 Class I)
Predicates
3

Its longest line of descent runs back 2 generations to K850342 (1985). 4 ancestors in all. 9 later devices descend from it, 5 citing it directly. 1 of the 14 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1997: 1 other ancestor1992: 1 other ancestor2009: 1 descendant cleared2019: 2 descendants cleared2020: 2 descendants cleared2022: 2 descendants cleared2025: 1 descendant cleared2021: 1 descendant clearedK850342 · Wyler Subdural Strip Electrode · Ad-Tech Medical Instrument Corp · 1985K021144 · Subdural Strip Electrode · Nicolet Biomedical, Inc. · 20021990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 5 descendants2 generations forward: 4 descendants K021144 · Subdural Strip Electrode · Nicolet Biomedical, Inc. · 2002K850342 · Wyler Subdural Strip Electrode · Ad-Tech Medical Instrument Corp · 19850
  1. Root · 2 generations backK850342 Wyler Subdural Strip ElectrodeAd-Tech Medical Instrument Corp · 1985
  2. 1 generation backK021144 Subdural Strip ElectrodeNicolet Biomedical, Inc. · 2002
  3. This deviceK053363 Dual-Sided Interhemispheric Subdural Electrodes, Grid Subdural…Ad-Tech Medical Instrument Corp · 2006
  4. Descendant · cites it directly · 4 of its ownK191186 Subdural Strip/Intraoperative Split Electrode, Subdural…Ad-Tech Medical Instrument Corporation · 2020
  5. Descendant · cites it directly · 1 of its ownK082474 PMT Subdural Cortical ElectrodesPmt Corp. · 2009
  6. Descendant · 2 generations on · 0 of its ownK242618 Layer 7-TPrecision Neuroscience, Corp. · 2025
  7. Descendant · cites it directly · 0 of its ownK221123 WISE Cortical Strip (WCS)Wise S.R.L. · 2022
  8. Descendant · 2 generations on · 0 of its ownK211954 Subdural Electrode, Strip/Intraoperative Strip, Grid/Intraoperative…Spes Medica Srl · 2022

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2009: 12019: 22020: 22021: 12022: 22025: 1
2009high 2 (2019) · low 0 (2010)2026

Most influential descendants

Showing 5 of 9 · See all
  • K191186Subdural Strip/Intraoperative Split Electrode, Subdural Grid/Intraoperative Grid…Ad-Tech Medical Instrument Corporation · 2020 · cites it directly4descendants
  • K082474PMT Subdural Cortical ElectrodesPmt Corp. · 2009 · cites it directly1descendant
  • K242618Layer 7-TPrecision Neuroscience, Corp. · 2025 · 2 generations on0descendants
  • K221123WISE Cortical Strip (WCS)Wise S.R.L. · 2022 · cites it directly0descendants
  • K211954Subdural Electrode, Strip/Intraoperative Strip, Grid/Intraoperative Grid, Multi-Strip…Spes Medica Srl · 2022 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K850342Wyler Subdural Strip ElectrodeAd-Tech Medical Instrument Corp1985—
−1K021144Subdural Strip ElectrodeNicolet Biomedical, Inc.2002—
0 (this device)K053363Dual-Sided Interhemispheric Subdural Electrodes, Grid Subdural Electrodes…Ad-Tech Medical Instrument Corp2006—
+1K191186Subdural Strip/Intraoperative Split Electrode, Subdural Grid/Intraoperative…Ad-Tech Medical Instrument Corporation2020—
+1K082474PMT Subdural Cortical ElectrodesPmt Corp.2009—
+2K242618Layer 7-TPrecision Neuroscience, Corp.2025—
+1K221123WISE Cortical Strip (WCS)Wise S.R.L.2022—
+2K211954Subdural Electrode, Strip/Intraoperative Strip, Grid/Intraoperative Grid…Spes Medica Srl2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199210
199710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201920
202020
202110
202220
202510
FIG. 1 — Lineage of K053363, 1985–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GYC device

  1. #1 K191186 Subdural Strip/Intraoperative Split Electrode…score 11.3 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K970587AD-Tech Subdural Electrode
Summary page 1
…mmon Name: Cortical Electrodes Classification Name: 21 CFR 882.1310, Cortical Electrode Predicate Device(s): 510(k) Number: K970587, K923803 Manufacture: AD-TECH Medical Instrument Corporation Trade Name:…
Ad-Tech Medical Instrument Corp1997—
*K923803AD-Tech Sub Dural Electrodeds
Summary page 1
…: Cortical Electrodes Classification Name: 21 CFR 882.1310, Cortical Electrode Predicate Device(s): 510(k) Number: K970587, K923803 Manufacture: AD-TECH Medical Instrument Corporation Trade Name: Subdural…
Ad-Tech Medical Instrument Corp1992—
*K021144Subdural Strip Electrode
Summary page 1
…Manufacture: AD-TECH Medical Instrument Corporation Trade Name: Subdural Electrode 510(k) Number: K021144 Manufacture: Nicolet Biomedical Corporation Trade Name: Subdural Strip E…
Nicolet Biomedical, Inc.2002—

Also cleared against K970587: 1 other device →

What descends from it? (9)

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K191186Subdural Strip/Intraoperative Split Electrode, Subdural Grid/Intraoperative Grid…Ad-Tech Medical Instrument Corporation20204
K082474PMT Subdural Cortical ElectrodesPmt Corp.20091
K221123WISE Cortical Strip (WCS)Wise S.R.L.20220
K191346Blackrock NeuroCoG Subdural Cortical Electrodes (Strips and Grids)Blackrock Microsystems20200
K183437AirRay Subdural Cortical ElectrodesCortec GmbH20190

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 18, 2019Class IIZ-2406-2019Employee errorSupplemental information provided with devices indicates that the subdural electrodes, depth electrodes, and anchor bolts are approved for use in the MR environment, for which they are not approved for such use.
Dec 18, 2012Class IZ-0892-2013Device DesignThe microelectrodes are not flush with the silastic substrate, may contain sharp edges capable of abrading the cortical surface, and may have remaining pieces which may fracture and remain in the cortical tissue with the electrode is…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 2 companies; a new generation was cleared every 16.9 years on median.
  • None of the 4 ancestors has a recall.
  • Of its descendants, K191186 (Ad-Tech Medical Instrument Corporation, 2020) has gone furthest, with 4 descendants of its own.
  • 1 other device also cites its predicate K970587.

How does its category compare?

All GYC devices

GYC Electrode, Cortical: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GYC

2in 2025

2008: 22009: 22010: 12011: 12016: 12018: 22019: 22020: 22021: 12022: 22025: 2
2007high 2 (2008) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GYC

Too few GYC 510(k)s in any year since 2007 to give a median.

Recalls per year · GYC

1in 2013

2011: 32012: 1, 1 Class I2013: 1
2007high 3 (2011) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · GYC

2in 2025

2007: 22008: 22009: 12010: 32011: 62012: 172013: 52014: 142015: 92016: 22017: 62018: 32019: 12020: 12021: 12022: 22023: 22024: 32025: 2
2007high 17 (2012) · low 1 (2009)2025

MAUDE reports naming product code GYC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—02
20082—02
20092—01
20101—03
20111—36
20120—1 (1)17
20130—15
20140—014
20150—09
20161—02
20170—06
20182—03
20192—01
20202—01
20211—01
20222—02
20230—02
20240—03
20252—02
20260—0—

What is it indicated for?

The AD-TECH Subdural Electrodes (Dual-Sided Interhemispheric, Grid, Intraoperative, Strip, Wyler) are intended for temporary (< 30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals on the surface of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain mapping.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K053363: Dual-Sided Interhemispheric Subdural Electrodes, Grid Subdural Electrodes, Intraoperative Subdural Electrodes.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K053363

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026