510(k) Report

510(k) · Cleared

Cleared

Subdural Strip/Intraoperative Split Electrode, Subdural Grid/Intraoperative Grid Electrode, Dual-Sided Interhemispheric Subdural Electrode, Multi-Strip and Slit Grid Subdural Electrode, Intraoperative Mapping Grid Subdural Electrode

K191186510(k) TraditionalCleared Jan 25, 2020267-day reviewClass II21 CFR 882.1310

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ad-Tech Medical Instrument Corporation · GYC Electrode, Cortical

Cleared on Jan 25, 2020 as substantially equivalent to K053363 (Ad-Tech Medical Instrument Corp, 2006) after a 267-day review. Its family tree reaches back 3 generations to 1985. 1 of the 10 devices in its family, counting itself, has a Class I recall. 4 later devices descend from it.

Decided
Jan 25, 2020
Review
267 days
Class
Class II
Product code
GYC
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K850342 (1985). 5 ancestors in all. 4 later devices descend from it, 4 citing it directly. 1 of the 10 devices in this family, counting this one, has a Class I recall.

  1. Root · 3 generations backK850342 Wyler Subdural Strip ElectrodeAd-Tech Medical Instrument Corp · 1985
  2. 2 generations backK021144 Subdural Strip ElectrodeNicolet Biomedical, Inc. · 2002
  3. 1 generation backK053363 Dual-Sided Interhemispheric Subdural Electrodes, Grid Subdural…Ad-Tech Medical Instrument Corp · 2006 · Recalled (Class I)
  4. This deviceK191186 Subdural Strip/Intraoperative Split Electrode, Subdural…Ad-Tech Medical Instrument Corporation · 2020
  5. Descendant · cites it directly · 0 of its ownK242618 Layer 7-TPrecision Neuroscience, Corp. · 2025
  6. Descendant · cites it directly · 0 of its ownK221123 WISE Cortical Strip (WCS)Wise S.R.L. · 2022
  7. Descendant · cites it directly · 0 of its ownK211954 Subdural Electrode, Strip/Intraoperative Strip, Grid/Intraoperative…Spes Medica Srl · 2022
  8. Descendant · cites it directly · 0 of its ownK201931 Dixi Medical Intraoperative Subdural Electrodes (Strips and Grids)Dixi Medical · 2021

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K242618Layer 7-TPrecision Neuroscience, Corp. · 2025 · cites it directly0descendants
  • K221123WISE Cortical Strip (WCS)Wise S.R.L. · 2022 · cites it directly0descendants
  • K211954Subdural Electrode, Strip/Intraoperative Strip, Grid/Intraoperative Grid, Multi-Strip…Spes Medica Srl · 2022 · cites it directly0descendants
  • K201931Dixi Medical Intraoperative Subdural Electrodes (Strips and Grids)Dixi Medical · 2021 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K850342Wyler Subdural Strip ElectrodeAd-Tech Medical Instrument Corp1985—
−2K021144Subdural Strip ElectrodeNicolet Biomedical, Inc.2002—
−1K053363Dual-Sided Interhemispheric Subdural Electrodes, Grid Subdural Electrodes…Ad-Tech Medical Instrument Corp2006Class I
0 (this device)K191186Subdural Strip/Intraoperative Split Electrode, Subdural Grid/Intraoperative…Ad-Tech Medical Instrument Corporation2020—
+1K242618Layer 7-TPrecision Neuroscience, Corp.2025—
+1K221123WISE Cortical Strip (WCS)Wise S.R.L.2022—
+1K211954Subdural Electrode, Strip/Intraoperative Strip, Grid/Intraoperative Grid…Spes Medica Srl2022—
+1K201931Dixi Medical Intraoperative Subdural Electrodes (Strips and Grids)Dixi Medical2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199210
199710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
202220
202510
FIG. 1 — Lineage of K191186, 1985–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GYC device

  1. #1 K250094 Atlas Stim Headbox (NK) (31-0601-0077); Atlas…score 11.3 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K053363Dual-Sided Interhemispheric Subdural Electrodes, Grid Subdural Electrodes…
Summary page 4
…lassification Name: Cortical Electrode 21 CFR 882.1310 Product Code: GYC Class II Predicate Devices: K053363 Subdural Electrode -Dual-Sided Interhemispheric, -Grid,…
Ad-Tech Medical Instrument Corp2006Class I
†K152769Fremap ElectrodeAd-Tech Medical Instrument Corporation2016—

Also cleared against K053363: 4 other devices →

What descends from it? (4)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K242618Layer 7-TPrecision Neuroscience, Corp.20250
K221123WISE Cortical Strip (WCS)Wise S.R.L.20220
K211954Subdural Electrode, Strip/Intraoperative Strip, Grid/Intraoperative Grid, Multi-Strip…Spes Medica Srl20220
K201931Dixi Medical Intraoperative Subdural Electrodes (Strips and Grids)Dixi Medical20210

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K053363Dual-Sided Interhemispheric Subdural Electrodes, Grid Subdural Electrodes, Intraoperative Subdural ElectrodesAd-Tech Medical Instrument Corp · 2006 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 35 years and passes through 3 companies; a new generation was cleared every 13.7 years on median.
  • 1 of the 5 ancestors has a recall of any class; 1 has a Class I recall.
  • Of its descendants, K242618 (Precision Neuroscience, Corp., 2025) has gone furthest, with 0 descendants of its own.
  • 4 other devices also cite its predicate K053363.

How does its category compare?

All GYC devices

GYC Electrode, Cortical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · GYC

2in 2025

2008: 22009: 22010: 12011: 12016: 12018: 22019: 22020: 22021: 12022: 22025: 2
2007high 2 (2008) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GYC

Too few GYC 510(k)s in any year since 2007 to give a median.

Recalls per year · GYC

1in 2013

2011: 32012: 1, 1 Class I2013: 1
2007high 3 (2011) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · GYC

2in 2025

2007: 22008: 22009: 12010: 32011: 62012: 172013: 52014: 142015: 92016: 22017: 62018: 32019: 12020: 12021: 12022: 22023: 22024: 32025: 2
2007high 17 (2012) · low 1 (2009)2025

MAUDE reports naming product code GYC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—02
20082—02
20092—01
20101—03
20111—36
20120—1 (1)17
20130—15
20140—014
20150—09
20161—02
20170—06
20182—03
20192—01
20202—01
20211—01
20222—02
20230—02
20240—03
20252—02
20260—0—

What is it indicated for?

The Ad-Tech Subdural Electrodes (Strip/Intraoperative Strip, Grid/Intraoperative Grid, Dual-Sided Interhemispheric,

Multi-Strip and Split Grid, Intraoperative Mapping Grid) are intended for temporary (< 30 days) use with recording,

monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals on the surface

of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain

mapping.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K191186: Subdural Strip/Intraoperative Split Electrode, Subdural Grid/Intraoperative Grid Electrode, Dual-Sided Interhemispheric Subdural Electrode, Multi-Strip and Slit Grid Subdural Electrode, Intraoperative Mapping Grid Subdural Electrode.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K191186

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)