510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

Triage Tox Drug Screen, Model 94400

K060791510(k) TraditionalCleared Jun 22, 200691-day reviewClass II21 CFR 862.3620Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biosite Incorporated · DJR Enzyme Immunoassay, Methadone

Cleared on Jun 22, 2006 as substantially equivalent to K973784 (Biosite Incorporated, 1997) and 1 other predicate after a 91-day review. Its family tree reaches back 2 generations to 2002. It has been recalled 5 times.

Decided
Jun 22, 2006
Review
91 days
Class
Class II
Product code
DJR
Recalls
5
Predicates
2

Its longest line of descent runs back 2 generations to K012745 (2002). 3 ancestors in all, the oldest K973784 (1997). No later device cites it as a predicate yet.

Other ancestors, by year cleared 1997: 1 other ancestorK012745 · Triage Tox Drug Screen, Catalog #94000 · Biosite Incorporated · 2002K043242 · Triage Tox Drug Screen · Biosite Incorporated · 2005 · recalled1996199820002002200420062008
Other ancestors, by generation 1 generation back: 1 other ancestor K043242 · Triage Tox Drug Screen · Biosite Incorporated · 2005 · recalledK012745 · Triage Tox Drug Screen, Catalog #94000 · Biosite Incorporated · 20020
  1. Root · 2 generations backK012745 Triage Tox Drug Screen, Catalog #94000Biosite Incorporated · 2002
  2. 1 generation backK043242 Triage Tox Drug ScreenBiosite Incorporated · 2005 · Recalled
  3. This deviceK060791 Triage Tox Drug Screen, Model 94400Biosite Incorporated · 2006
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K012745Triage Tox Drug Screen, Catalog #94000Biosite Incorporated2002—
−1K043242Triage Tox Drug ScreenBiosite Incorporated20051
0 (this device)K060791Triage Tox Drug Screen, Model 94400Biosite Incorporated2006—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199710
FIG. 1 — Lineage of K060791, 2002–2006Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DJR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K973784Triage Panel for Drugs of Abuse, Plus Tricyclic Antidepressants and Other Triage Drug…
Summary page 1
…: 03/22/06 B. Device Names Triage® TOX Drug Screen C. Predicate Devices Biosite Triage® 8 Panel for Drugs of Abuse [FDA file number K973784] Biosite Triage® TOX Drug Screen [FDA file number K043242] 0. Device Description and Intended Use The Triage®…
Biosite Incorporated19974
*K043242Triage Tox Drug Screen
Summary page 1
…Drug Screen C. Predicate Devices Biosite Triage® 8 Panel for Drugs of Abuse [FDA file number K973784] Biosite Triage® TOX Drug Screen [FDA file number K043242] 0. Device Description and Intended Use The Triage® TOX Drug Screen Methadone assay is a fluorescence immunoassay in…
Biosite Incorporated20051

Also cleared against K973784: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 5, 2014Class IIZ-1214-2014Process controlAlere San Diego, Inc. is initiating a recall of the Alere Triage¿ TOX Drug Screen, Catalog Number 94400, Lots W56013B and W56035B because Alere's investigation has shown that these lots have a frequency of false negative results for PCP…
Jul 12, 2012Class IIZ-2460-2012Nonconforming Material/ComponentAlere San Diego is recalling Alere Triage TOX Drug Screen because there is a possibility that after installing the code chip provided with the affected lots: AMP, BAR, and PCP, the threshold ranges may not be consistently set correctly in…
May 14, 2012Class IIZ-2039-2012Nonconforming Material/ComponentA recall was initiated because Alere San Diego has confirmed that the Triage TOX Drug Screen may have significantly decreased precision and accuracy than claimed. An increased frequency of false positive and/or false negative results has…
Mar 12, 2010Not classifiedZ-1446-2012Nonconforming Material/ComponentThe recall was initiated because Biosite Inc. received an increase in customer calls regarding unexpected positive THC results when using the Triage TOX Drug Screen.
May 28, 2009Class IIZ-0029-2013Error in labelingBiosite is recalling the Triage TOX Drug Screen kits where the box containing test devices because it was mislabeled.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 4 years and passes through 1 company; a new generation was cleared every 3.1 years on median.
  • 2 of the 3 ancestors have a recall of any class.
  • 1 other device also cites its predicate K973784.

How does its category compare?

All DJR devices

DJR Enzyme Immunoassay, Methadone: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DJR

1in 2019

2009: 12010: 32013: 12015: 32016: 12017: 12018: 12019: 1
2007high 3 (2010) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DJR

59in 2015

2010: 469 days median2015: 59 days median
2007high 469 (2010) · low 59 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DJR

1in 2010

2008: 22009: 12010: 1
2007high 2 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DJR

1in 2025

2010: 32011: 62012: 12014: 22015: 22016: 12025: 1
2007high 6 (2011) · low 0 (2007)2025

MAUDE reports naming product code DJR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—20
20091—10
2010346913
20110—06
20120—01
20131—00
20140—02
201535902
20161—01
20171—00
20181—00
20191—00
20200—00
20210—00
20220—00
20230—00
20240—00
20250—01
20260—0—

What is it indicated for?

The Triage TOX Drug Screen is a fluorescence immunoassay intended to be used with the Triage Meters for the point-of-care qualitative determination of the presence of drug and/or the major metabolites above the threshold concentrations of up to 10 distinct drug classes, including assays for acetaminophen/paracetamol, amphetamines, methamphetamines, barbiturates, benzodiazepines, cocaine, methadone, opiates, phencyclidine, THC and tricyclic…

Read the full statement (196 words)

The Triage TOX Drug Screen is a fluorescence immunoassay intended to be used with the Triage Meters for the point-of-care qualitative determination of the presence of drug and/or the major metabolites above the threshold concentrations of up to 10 distinct drug classes, including assays for acetaminophen/paracetamol, amphetamines, methamphetamines, barbiturates, benzodiazepines, cocaine, methadone, opiates, phencyclidine, THC and tricyclic antidepressants in urine.

The acetaminophen/paracetamol assay will yield positive results when acetaminophen/paracetamol is ingested at or above therapeutic doses.

The threshold concentrations are provided below:

Acetaminophen/Paracetamol APAP 5 pg/mL

Amphetamines AMP 1000 ng/mL

Methamphetamines mAMP 1000 ng/mL

Barbiturates BAR 300 ng/mL

Benzodiazepines BZO 300 ng/mL

Cocaine COC 300 ng/mL

Methadone MTD 300 ng/mL

Opiates OPI 300 ng/mL

Phencyclidine PCP 25 ng/mL

THC THC 50 ng/mL

Tricyclic Antidepressants TCA 1000 ng/mL

This test provides only preliminary test results. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectroscopy (GC/MS) is the preferred confirmatory method.

A quantitative serum acetaminophen/paracetamol measurement is the common confirmatory method for preliminary positive acetaminophen/paracetamol results.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K060791: Triage Tox Drug Screen, Model 94400.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K060791

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026