510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Coaguchek XS System

K060978510(k) TraditionalCleared Aug 11, 2006123-day reviewClass II21 CFR 864.7750Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Diagnostics · GJS Test, Time, Prothrombin

Cleared on Aug 11, 2006 as substantially equivalent to K020831 (Roche Diagnostics Corp., 2002) after a 123-day review. Its family tree reaches back 2 generations to 2001. 4 of the 19 devices in its family, counting itself, have a Class I recall. 16 later devices descend from it. It has been recalled 3 times.

Decided
Aug 11, 2006
Review
123 days
Class
Class II
Product code
GJS
Recalls
3
Predicates
1

Its longest line of descent runs back 2 generations to K010599 (2001). 16 later devices descend from it, 5 citing it directly. 4 of the 19 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2007: 2 descendants cleared, 1 with a Class I recall12019: 1 descendant cleared2016: 1 descendant cleared2024: 1 descendant cleared2017: 1 descendant cleared, 1 with a Class I recall12018: 2 descendants cleared, 1 with a Class I recall12023: 2 descendants cleared2010: 2 descendants cleared, 1 with a Class I recall12025: 3 descendants cleared2020: 1 descendant clearedK010599 · Protime Microcoagulation System / Protime 3 Cuvette · International Technidyne Corp. · 2001 · recalledK020831 · PT S Test Strips and Controls for the Coaguchek S System, Models 2032422 and 3271404 · Roche Diagnostics Corp. · 2002200020052010201520202025
Descendants, by generation 1 generation forward: 5 descendants, 1 with a Class I recall12 generations forward: 8 descendants, 3 with a Class I recall33 generations forward: 3 descendants K020831 · PT S Test Strips and Controls for the Coaguchek S System, Models 2032422 and 3271404 · Roche Diagnostics Corp. · 2002K010599 · Protime Microcoagulation System / Protime 3 Cuvette · International Technidyne Corp. · 2001 · recalled0
  1. Root · 2 generations backK010599 Protime Microcoagulation System / Protime 3 CuvetteInternational Technidyne Corp. · 2001 · Recalled
  2. 1 generation backK020831 PT S Test Strips and Controls for the Coaguchek S System, Models…Roche Diagnostics Corp. · 2002
  3. This deviceK060978 Coaguchek XS SystemRoche Diagnostics · 2006
  4. Descendant · cites it directly · 6 of its ownK071041 Coaguchek XS Plus SystemRoche Diagnostics · 2007 · Recalled (Class I)
  5. Descendant · cites it directly · 4 of its ownK180780 microINR SystemIline Microsystems, S.L. · 2019
  6. Descendant · 2 generations on · 3 of its ownK201185 microINR SystemIline Microsystems, S.L. · 2020
  7. Descendant · 3 generations on · 2 of its ownK231711 microINR SystemIline Microsystems, S.L. · 2023
  8. Descendant · 2 generations on · 1 of its ownK243543 microINR SystemIline Microsystems, S.L. · 2025

Showing 5 of 16 descendants. See all descendants

Descendants cleared per year

2007: 22010: 22016: 12017: 12018: 22019: 12020: 12023: 22024: 12025: 3
2007high 3 (2025) · low 0 (2008)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 16 · See all
  • K071041Coaguchek XS Plus SystemRoche Diagnostics · 2007 · cites it directly6descendants
  • K180780microINR SystemIline Microsystems, S.L. · 2019 · cites it directly4descendants
  • K201185microINR SystemIline Microsystems, S.L. · 2020 · 2 generations on3descendants
  • K231711microINR SystemIline Microsystems, S.L. · 2023 · 3 generations on2descendants
  • K243543microINR SystemIline Microsystems, S.L. · 2025 · 2 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K010599Protime Microcoagulation System / Protime 3 CuvetteInternational Technidyne Corp.20013
−1K020831PT S Test Strips and Controls for the Coaguchek S System, Models 2032422 and…Roche Diagnostics Corp.2002—
0 (this device)K060978Coaguchek XS SystemRoche Diagnostics2006—
+1K071041Coaguchek XS Plus SystemRoche Diagnostics2007Class I
+1K180780microINR SystemIline Microsystems, S.L.2019—
+2K201185microINR SystemIline Microsystems, S.L.2020—
+3K231711microINR SystemIline Microsystems, S.L.2023—
+2K243543microINR SystemIline Microsystems, S.L.2025—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200721
201021
201610
201711
201821
201910
202010
202320
202410
202530
FIG. 1 — Lineage of K060978, 2001–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GJS device

  1. #1 K243543 microINR Systemscore 13.4 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K020831PT S Test Strips and Controls for the Coaguchek S System, Models 2032422 and 3271404
Summary page 1
…ercial distribution intended for similar use. Most notably, it is substantially equivalent to the Roche Diagnostics CoaguChek S System (K02083 1). 4) Device The CoaguChek XS is a 3 a generation CoaguChek meter which measures Description prothrombin time in fre…
Roche Diagnostics Corp.2002—

Also cleared against K020831: 1 other device →

What descends from it? (16)

See all 16

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K071041Coaguchek XS Plus System Class IRoche Diagnostics20076
K180780microINR SystemIline Microsystems, S.L.20194
K062925Coaguchek XS SystemRoche Diagnostics Corp.20071
K230802Xprecia Prime Coagulation SystemUniversal Biosensors Pty, Ltd.20240
K151964Xprecia Stride Coagulation System, Xprecia System PT ControlsSiemens Healthcare Diagnostics20160

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 2, 2020Class IIZ-0803-2021Device DesignPossible Misinterpretation of Displayed Measuring Units from INR to %Quick or Seconds; providing instructions to users for ensuring the meter is displaying INR results
May 16, 2013Class IIZ-1584-2013Software designRoche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.
Jul 22, 2009Not classifiedZ-2002-2009Device DesignThe recommended cleaning procedures for the meter may cause a malfunction and falsely elevated test result reports.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K071041Coaguchek XS Plus SystemRoche Diagnostics · 2007 · Cites it directly · ◆ in the 2007 descendants bar of FIG. 11 Class I recall
  • K170960CoaguChek Vantus SystemRoche Diagnostics · 2017 · 2 generations on · ◆ in the 2017 descendants bar of FIG. 11 Class I recall
  • K180684CoaguChek XS Plus SystemRoche Diagnostics · 2018 · 2 generations on · ◆ in the 2018 descendants bar of FIG. 11 Class I recall
  • K093460Coaguchek XS Pro SystemRoche Diagnostics · 2010 · 2 generations on · ◆ in the 2010 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 5 years and passes through 3 companies; a new generation was cleared every 3.8 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K071041 (Roche Diagnostics, 2007) has gone furthest, with 6 descendants of its own.
  • 1 other device also cites its predicate K020831.

How does its category compare?

All GJS devices

GJS Test, Time, Prothrombin: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GJS

3in 2025

2007: 42008: 12010: 42012: 12013: 12014: 12016: 12017: 12018: 22019: 22020: 12022: 32023: 22024: 12025: 3
2007high 4 (2007) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GJS

60in 2025

2007: 77 days median2010: 168 days median2022: 399 days median2025: 60 days median
2007high 399 (2022) · low 60 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GJS

1in 2022

2007: 52008: 32009: 42010: 22011: 12012: 12013: 12014: 4, 3 Class I2017: 52018: 3, 2 Class I2020: 22021: 12022: 1
2007high 5 (2007) · low 0 (2015)2026

Class I (5 since 2007) Class II/III not classified

Adverse event reports per year · GJS

105in 2025

2007: 4392008: 4202009: 8762010: 2,1772011: 2,9652012: 2,1982013: 2,5842014: 2,3972015: 1,2602016: 8642017: 5322018: 1,9582019: 2,3852020: 1,3902021: 1,3592022: 1,2102023: 8092024: 1772025: 105
2007high 2,965 (2011) · low 105 (2025)2025

MAUDE reports naming product code GJS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074775439
20081—3420
20090—4876
2010416822,177
20110—12,965
20121—12,198
20131—12,584
20141—4 (3)2,397
20150—01,260
20161—0864
20171—5532
20182—3 (2)1,958
20192—02,385
20201—21,390
20210—11,359
2022339911,210
20232—0809
20241—0177
20253600105
20260—0—

What is it indicated for?

The CoaguChek XS System is intended for use by professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy. The CoaguChek XS System uses fresh capillary or non-anticoagulated venous whole blood.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K060978: Coaguchek XS System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K060978

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026