510(k) Report

510(k) · Cleared

Recalled · Class I · 2018

Coaguchek XS Plus System

K071041510(k) SpecialCleared May 11, 200729-day reviewClass II21 CFR 864.7750Recalls 3 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Diagnostics · GJS Test, Time, Prothrombin

Cleared on May 11, 2007 as substantially equivalent to K060978 (Roche Diagnostics, 2006) after a 29-day review. Its family tree reaches back 3 generations to 2001. 3 of the 10 devices in its family, counting itself, have a Class I recall. 6 later devices descend from it. It has been recalled 3 times, including a Class I recall.

Decided
May 11, 2007
Review
29 days
Class
Class II
Product code
GJS
Recalls
3 (1 Class I)
Predicates
1

Its longest line of descent runs back 3 generations to K010599 (2001). 6 later devices descend from it, 4 citing it directly. 3 of the 10 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2018: 2 descendants cleared, 1 with a Class I recall12023: 1 descendant cleared2010: 2 descendants cleared, 1 with a Class I recall12025: 1 descendant clearedK010599 · Protime Microcoagulation System / Protime 3 Cuvette · International Technidyne Corp. · 2001 · recalledK020831 · PT S Test Strips and Controls for the Coaguchek S System, Models 2032422 and 3271404 · Roche Diagnostics Corp. · 2002K060978 · Coaguchek XS System · Roche Diagnostics · 2006 · recalled200020052010201520202025
  1. Root · 3 generations backK010599 Protime Microcoagulation System / Protime 3 CuvetteInternational Technidyne Corp. · 2001 · Recalled
  2. 2 generations backK020831 PT S Test Strips and Controls for the Coaguchek S System, Models…Roche Diagnostics Corp. · 2002
  3. 1 generation backK060978 Coaguchek XS SystemRoche Diagnostics · 2006 · Recalled
  4. This deviceK071041 Coaguchek XS Plus SystemRoche Diagnostics · 2007
  5. Descendant · cites it directly · 1 of its ownK180684 CoaguChek XS Plus SystemRoche Diagnostics · 2018 · Recalled (Class I)
  6. Descendant · cites it directly · 1 of its ownK093460 Coaguchek XS Pro SystemRoche Diagnostics · 2010 · Recalled (Class I)
  7. Descendant · 2 generations on · 0 of its ownK253188 CoaguChek XS Plus SystemRoche Diagnostics · 2025
  8. Descendant · cites it directly · 0 of its ownK220282 i-STAT PTplus Cartridge with the i-STAT 1 SystemAbbott Laboratories · 2023
  9. Descendant · 2 generations on · 0 of its ownK180693 CoaguChek XS Pro SystemRoche Diagnostics · 2018

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2010: 22018: 22023: 12025: 1
2010high 2 (2010) · low 0 (2011)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 6 · See all
  • K180684CoaguChek XS Plus SystemRoche Diagnostics · 2018 · cites it directly1descendant
  • K093460Coaguchek XS Pro SystemRoche Diagnostics · 2010 · cites it directly1descendant
  • K253188CoaguChek XS Plus SystemRoche Diagnostics · 2025 · 2 generations on0descendants
  • K220282i-STAT PTplus Cartridge with the i-STAT 1 SystemAbbott Laboratories · 2023 · cites it directly0descendants
  • K180693CoaguChek XS Pro SystemRoche Diagnostics · 2018 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K010599Protime Microcoagulation System / Protime 3 CuvetteInternational Technidyne Corp.20013
−2K020831PT S Test Strips and Controls for the Coaguchek S System, Models 2032422 and…Roche Diagnostics Corp.2002—
−1K060978Coaguchek XS SystemRoche Diagnostics20063
0 (this device)K071041Coaguchek XS Plus SystemRoche Diagnostics2007—
+1K180684CoaguChek XS Plus SystemRoche Diagnostics2018Class I
+1K093460Coaguchek XS Pro SystemRoche Diagnostics2010Class I
+2K253188CoaguChek XS Plus SystemRoche Diagnostics2025—
+1K220282i-STAT PTplus Cartridge with the i-STAT 1 SystemAbbott Laboratories2023—
+2K180693CoaguChek XS Pro SystemRoche Diagnostics2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201021
201821
202310
202510
FIG. 1 — Lineage of K071041, 2001–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GJS device

  1. #1 K243543 microINR Systemscore 13.4 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K060978Coaguchek XS System
Summary page 1
…e blood. Predicate The Roche Diagnostics CoaguChek XS Plus System is substantially Device equivalent to to the previously cleared CoaguChek XS System (K060978). Similarities The table below indicates the similarities between the CoaguChek XS Plus System and the Coa…
Roche Diagnostics20063

Also cleared against K060978: 4 other devices →

What descends from it? (6)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K180684CoaguChek XS Plus System Class IRoche Diagnostics20181
K093460Coaguchek XS Pro System Class IRoche Diagnostics20101
K220282i-STAT PTplus Cartridge with the i-STAT 1 SystemAbbott Laboratories20230
K092940Coaguchek XS Plus PST SystemRoche Diagnostics Corp.20100

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 13, 2018Class IZ-0360-2019Nonconforming Material/ComponentAbnormally high INR test results with the affected CoaguChek test strips
May 16, 2013Class IIZ-1583-2013Software designRoche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.
Jul 22, 2009Not classifiedZ-2004-2009Device DesignThe recommended cleaning procedures for the meter may cause a malfunction and falsely elevated test result reports.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K180684CoaguChek XS Plus SystemRoche Diagnostics · 2018 · Cites it directly · ◆ in the 2018 descendants bar of FIG. 11 Class I recall
  • K093460Coaguchek XS Pro SystemRoche Diagnostics · 2010 · Cites it directly · ◆ in the 2010 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 6 years and passes through 3 companies; a new generation was cleared every 1.3 years on median.
  • 2 of the 3 ancestors have a recall of any class.
  • Of its descendants, K180684 (Roche Diagnostics, 2018) has gone furthest, with 1 descendant of its own.
  • 4 other devices also cite its predicate K060978.

How does its category compare?

All GJS devices

GJS Test, Time, Prothrombin: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · GJS

3in 2025

2007: 42008: 12010: 42012: 12013: 12014: 12016: 12017: 12018: 22019: 22020: 12022: 32023: 22024: 12025: 3
2007high 4 (2007) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GJS

60in 2025

2007: 77 days median2010: 168 days median2022: 399 days median2025: 60 days median
2007high 399 (2022) · low 60 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GJS

1in 2022

2007: 52008: 32009: 42010: 22011: 12012: 12013: 12014: 4, 3 Class I2017: 52018: 3, 2 Class I2020: 22021: 12022: 1
2007high 5 (2007) · low 0 (2015)2026

Class I (5 since 2007) Class II/III not classified

Adverse event reports per year · GJS

105in 2025

2007: 4392008: 4202009: 8762010: 2,1772011: 2,9652012: 2,1982013: 2,5842014: 2,3972015: 1,2602016: 8642017: 5322018: 1,9582019: 2,3852020: 1,3902021: 1,3592022: 1,2102023: 8092024: 1772025: 105
2007high 2,965 (2011) · low 105 (2025)2025

MAUDE reports naming product code GJS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074775439
20081—3420
20090—4876
2010416822,177
20110—12,965
20121—12,198
20131—12,584
20141—4 (3)2,397
20150—01,260
20161—0864
20171—5532
20182—3 (2)1,958
20192—02,385
20201—21,390
20210—11,359
2022339911,210
20232—0809
20241—0177
20253600105
20260—0—

What is it indicated for?

The CoaguChek XS Plus System is intended for use by professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy. The system uses fresh capillary or non-anticoagulated venous whole blood.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K071041: Coaguchek XS Plus System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K071041

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026