510(k) Report

510(k) · Cleared

Cleared

Accurex

K061126510(k) TraditionalCleared May 9, 200615-day reviewClass II21 CFR 892.2050

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cybermed, Inc. · LLZ System, Image Processing, Radiological

Cleared on May 9, 2006 as substantially equivalent to K013878 (Cybermed, Inc., 2001) and 1 other predicate after a 15-day review. Its family tree reaches back 5 generations to 1996. 3 later devices descend from it.

Decided
May 9, 2006
Review
15 days
Class
Class II
Product code
LLZ
Recalls
None
Predicates
2

Its longest line of descent runs back 5 generations to K954678 (1996). 9 ancestors in all, the oldest K924810 (1993). 3 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor1997: 1 other ancestor1993: 2 other ancestors2009: 1 descendant cleared2019: 1 descendant cleared2023: 1 descendant clearedK954678 · Pro Vision · Algotec Systems, Ltd. · 1996K980648 · Pro Vision · Algotec Systems, Ltd. · 1998K992654 · Plug 'N View 3D, Version 1.0 · Voxar Limited · 1999K013878 · Cybermed, Inc., V-Works · Cybermed, Inc. · 2001K053155 · Vimplant · Cybermed, Inc. · 20051990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 2 other ancestors4 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant K053155 · Vimplant · Cybermed, Inc. · 2005K013878 · Cybermed, Inc., V-Works · Cybermed, Inc. · 2001K992654 · Plug 'N View 3D, Version 1.0 · Voxar Limited · 1999K980648 · Pro Vision · Algotec Systems, Ltd. · 1998K954678 · Pro Vision · Algotec Systems, Ltd. · 1996−50
  1. Root · 5 generations backK954678 Pro VisionAlgotec Systems, Ltd. · 1996
  2. 4 generations backK980648 Pro VisionAlgotec Systems, Ltd. · 1998
  3. 3 generations backK992654 Plug 'N View 3D, Version 1.0Voxar Limited · 1999
  4. 2 generations backK013878 Cybermed, Inc., V-WorksCybermed, Inc. · 2001
  5. 1 generation backK053155 VimplantCybermed, Inc. · 2005
  6. This deviceK061126 AccurexCybermed, Inc. · 2006
  7. Descendant · cites it directly · 2 of its ownK090607 Blue Sky Bio Plan, Version .38Blue Sky Bio, LLC · 2009
  8. Descendant · 2 generations on · 1 of its ownK190096 R2GATEMegagen Implant Co., Ltd. · 2019
  9. Descendant · 3 generations on · 0 of its ownK223909 R2GATE Lite TMMegagen Implant Co., Ltd. · 2023

Most influential descendants

Showing 3 of 3 · See all
  • K090607Blue Sky Bio Plan, Version .38Blue Sky Bio, LLC · 2009 · cites it directly2descendants
  • K190096R2GATEMegagen Implant Co., Ltd. · 2019 · 2 generations on1descendant
  • K223909R2GATE Lite TMMegagen Implant Co., Ltd. · 2023 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K954678Pro VisionAlgotec Systems, Ltd.1996—
−4K980648Pro VisionAlgotec Systems, Ltd.1998—
−3K992654Plug 'N View 3D, Version 1.0Voxar Limited1999—
−2K013878Cybermed, Inc., V-WorksCybermed, Inc.2001—
−1K053155VimplantCybermed, Inc.2005—
0 (this device)K061126AccurexCybermed, Inc.2006—
+1K090607Blue Sky Bio Plan, Version .38Blue Sky Bio, LLC2009—
+2K190096R2GATEMegagen Implant Co., Ltd.2019—
+3K223909R2GATE Lite TMMegagen Implant Co., Ltd.2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199320
199710
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201910
202310
FIG. 1 — Lineage of K061126, 1996–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K013878Cybermed, Inc., V-Works
Summary page 1
…communications system 3. Predicate Devices: Manufacturer CyberMed, Inc. Device: V-worksTm 510(k) Number: K01 3878 (Decision Date - 12107/2001) Manufacturer: CyberMed, Inc. Device: Vimplant 5…
Cybermed, Inc.2001—
*K053155Vimplant
Summary page 1
…K01 3878 (Decision Date - 12107/2001) Manufacturer: CyberMed, Inc. Device: Vimplant 510(k) Number: K053155 (Decision Date- 11/23/2005) 4. Classifications Names & Citations: 21CFR 8B1.2050, LLZ, Picture archiving and communicat…
Cybermed, Inc.2005—

Also cleared against K013878: 2 other devices →

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K090607Blue Sky Bio Plan, Version .38Blue Sky Bio, LLC20092

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 3 companies; a new generation was cleared every 2.1 years on median.
  • None of the 9 ancestors has a recall.
  • Of its descendants, K090607 (Blue Sky Bio, LLC, 2009) has gone furthest, with 2 descendants of its own.
  • 2 other devices also cite its predicate K013878.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

Use as a software for the display and 3D visualization of dental image files from scanning devices such as CT, MRI for radiologists, dental clinicians and practitioners to acquire, process, render, review, store, print and distribute DICOM 3.0 complaint image studies, utilizing standard PC hardware.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K061126: Accurex.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K061126

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026