Deltec Cozmo Insulin Infusion Pump with Cozmonitor Blood Glucose Meter, Models 21-1815, 21-1816, 21-1817

K062323510(k) SpecialCleared Nov 1, 200684-day reviewClass II21 CFR 880.5725Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Smiths Medical MD, Inc. · LZG Pump, Infusion, Insulin

Cleared on Nov 1, 2006 after an 84-day review. The FDA summary names its predicate without a number. 1 of the 14 devices in its family, counting itself, has a Class I recall. 13 later devices descend from it. It has been recalled 5 times.

Decided
Nov 1, 2006
Review
84 days
Class
Class II
Product code
LZG
Recalls
5
Predicates
0

13 later devices descend from it, 2 citing it directly. 1 of the 14 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2007: 2 descendants cleared2008: 1 descendant cleared2011: 1 descendant cleared, 1 with a Class I recall12014: 2 descendants cleared2016: 3 descendants cleared2015: 1 descendant cleared2024: 1 descendant cleared2026: 2 descendants cleared20052010201520202025
  1. This deviceK062323 Deltec Cozmo Insulin Infusion Pump with Cozmonitor Blood Glucose…Smiths Medical MD, Inc. · 2006
  2. Descendant · cites it directly · 10 of its ownK071613 Amigo Insulin PumpNipro Diabetes Systems, Inc. · 2007
  3. Descendant · 2 generations on · 9 of its ownK111210 T:Slim (TM) Insulin Delivery SystemTandem Diabetes Care, Inc. · 2011 · Recalled (Class I)
  4. Descendant · 3 generations on · 8 of its ownK133593 T:Slim Insulin Delivery SystemTandem Diabetes Care, Inc. · 2014
  5. Descendant · 3 generations on · 7 of its ownK141758 T:Slim Insulin Delivery SystemTandem Diabetes Care, Inc. · 2014
  6. Descendant · 4 generations on · 4 of its ownK160056 t:slim Insulin Delivery SystemTandem Diabetes Care, Inc. · 2016

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2007: 22008: 12011: 12014: 22015: 12016: 32024: 12026: 2
2007high 3 (2016) · low 0 (2009)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 13 · See all
  • K071613Amigo Insulin PumpNipro Diabetes Systems, Inc. · 2007 · cites it directly10descendants
  • K111210T:Slim (TM) Insulin Delivery SystemTandem Diabetes Care, Inc. · 2011 · 2 generations on9descendants
  • K133593T:Slim Insulin Delivery SystemTandem Diabetes Care, Inc. · 2014 · 3 generations on8descendants
  • K141758T:Slim Insulin Delivery SystemTandem Diabetes Care, Inc. · 2014 · 3 generations on7descendants
  • K160056t:slim Insulin Delivery SystemTandem Diabetes Care, Inc. · 2016 · 4 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K062323Deltec Cozmo Insulin Infusion Pump with Cozmonitor Blood Glucose Meter, Models…Smiths Medical MD, Inc.2006—
+1K071613Amigo Insulin PumpNipro Diabetes Systems, Inc.2007—
+2K111210T:Slim (TM) Insulin Delivery SystemTandem Diabetes Care, Inc.2011Class I
+3K133593T:Slim Insulin Delivery SystemTandem Diabetes Care, Inc.2014—
+3K141758T:Slim Insulin Delivery SystemTandem Diabetes Care, Inc.2014—
+4K160056t:slim Insulin Delivery SystemTandem Diabetes Care, Inc.2016—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200720
200810
201111
201420
201510
201630
202410
202620
FIG. 1 — Lineage of K062323, 2006–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LZG device

  1. #1 K253534 Pivot Insulin Delivery Systemscore 16.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Deltec Cozmo® Insulin Pump Model 1700
Named without a submission number
—
*—Animas IR 1250
Named without a submission number
—

What descends from it? (13)

See all 13

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K071613Amigo Insulin PumpNipro Diabetes Systems, Inc.200710
K063674Ezmanager Max Diabetes Management SoftwareAnimas Corp.20071

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 20, 2008Not classifiedZ-0989-2009Software designSmiths Medical has become aware of a display irregularity with the Deltec Cozmo Insulin Pump, Model 1800 and Model 21-1700 containing Model 1800 software. There have been adverse events reported that the amount of Extended Bolus delivered…
Nov 20, 2008Not classifiedZ-0988-2009Software designSmiths Medical has become aware of a display irregularity with the Deltec Cozmo Insulin Pump, Model 1800 and Model 21-1700 containing Model 1800 software. There have been adverse events reported that the amount of Extended Bolus delivered…
Mar 10, 2008Not classifiedZ-1623-2008Component change controlSmith Medical discovered an issue with a motor component in specific serial numbers of the Deltec Cozmo Insulin Pump, Model Numbers 1700 and 1800. This issue affects how the motor operates and may cause an over-delivery of insulin which…
Mar 10, 2008Not classifiedZ-1622-2008Component change controlSmith Medical discovered an issue with a motor component in specific serial numbers of the Deltec Cozmo Insulin Pump, Model Numbers 1700 and 1800. This issue affects how the motor operates and may cause an over-delivery of insulin which…
Feb 22, 2008Not classifiedZ-1591-2008Device DesignSmith Medical has received reports of Deltec Cozmo Insulin Pumps powering down without an alarm. Investigation concluded that if the battery cap is not fully tightened, the Pump may power down and a brief "chirp" will sound. This may…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K111210T:Slim (TM) Insulin Delivery SystemTandem Diabetes Care, Inc. · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K071613 (Nipro Diabetes Systems, Inc., 2007) has gone furthest, with 10 descendants of its own.

How does its category compare?

All LZG devices

LZG Pump, Infusion, Insulin: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LZG

2in 2024

2007: 32008: 62009: 22010: 22011: 42012: 32013: 22014: 42015: 12016: 42017: 12018: 12019: 22021: 12022: 12023: 12024: 22026 (partial year): 2
2007high 6 (2008) · low 0 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LZG

116in 2016

2007: 112 days median2008: 129 days median2011: 189 days median2012: 73 days median2014: 123 days median2016: 116 days median
2007high 189 (2011) · low 73 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LZG

1in 2022

2007: 12008: 132009: 22010: 42011: 22012: 42013: 3, 1 Class I2014: 2, 1 Class I2015: 2, 2 Class I2018: 3, 2 Class I2019: 52022: 1, 1 Class I2026 (partial year): 2, 2 Class I
2007high 13 (2008) · low 0 (2016)2026

Class I (9 since 2007) Class II/III not classified

Adverse event reports per year · LZG

12,004in 2025

2007: 5,1102008: 5,7112009: 9,4712010: 10,1752011: 12,8582012: 18,3672013: 30,0592014: 49,1882015: 69,1362016: 56,5812017: 44,4722018: 32,6522019: 29,0272020: 32,5882021: 31,5222022: 28,2252023: 21,9722024: 18,2282025: 12,004
2007high 69,136 (2015) · low 5,110 (2007)2025

MAUDE reports naming product code LZG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007311215,110
20086129135,711
20092—29,471
20102—410,175
20114189212,858
2012373418,367
20132—3 (1)30,059
201441232 (1)49,188
20151—2 (2)69,136
20164116056,581
20171—044,472
20181—3 (2)32,652
20192—529,027
20200—032,588
20211—031,522
20221—1 (1)28,225
20231—021,972
20242—018,228
20250—012,004
20262—2 (2)—

What is it indicated for?

The Deltec Cozmo® Insulin Pump is a syringe infusion pump designed for Continuous Subcutaneous Insulin Infusion (CSII) for the control of diabetes. The Deltec CoZmonitorTM Blood Glucose Monitor is available as an accessory to be used solely with the Deltec Cozmo® Insulin Pump. Because the monitor is powered by FreeStyleTM, blood sampling can be performed using the finger, forearm, upper arm, thigh, calf or hand. All testing is performed outside of the…

Read the full statement (198 words)

The Deltec Cozmo® Insulin Pump is a syringe infusion pump designed for Continuous Subcutaneous Insulin Infusion (CSII) for the control of diabetes.

The Deltec CoZmonitorTM Blood Glucose Monitor is available as an accessory to be used solely with the Deltec Cozmo® Insulin Pump. Because the monitor is powered by FreeStyleTM, blood sampling can be performed using the finger, forearm, upper arm, thigh, calf or hand. All testing is performed outside of the body (in other words, in-vitro diagnostic use only). Use only with FreeStyleTM Test Strip and FreeStyleTM Control Solution, as other products can give inaccurate results.

The Deltec CoZmonitorTM Blood Glucose Monitoring System is designed to be used only with the Deltec Cozmo® Insulin Pump to help you and your healthcare provider monitor your diabetes management program. When used properly, the monitor provides a quantitative measurement of glucose (or sugar) in whole blood. Because the monitor is powered by FreeStyleTM, blood sampling can be performed using the finger, forearm, upper arm, thigh, calf or hand. All testing is performed outside of the body (in other words, in-vitro diagnostic use only). Use only with FreeStyleTM Test Strip and FreeStyleTM Control Solution, as other products can give inaccurate results.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K062323: Deltec Cozmo Insulin Infusion Pump with Cozmonitor Blood Glucose Meter, Models 21-1815, 21-1816, 21-1817.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K062323

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026