510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Vitros Chemistry Products OP Reagent, Calibrator Kit 26 and Dat Performance Verifiers I, II and Iii

K062460510(k) TraditionalCleared Dec 15, 2006114-day reviewClass II21 CFR 862.3650Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ortho-Clinical Diagnostics, Inc. · DJG Enzyme Immunoassay, Opiates

Cleared on Dec 15, 2006 as substantially equivalent to K011289 (Syva Co., Dade Behring, Inc., 2001) and 1 other predicate after a 114-day review. Its family tree reaches back 1 generation to 2001. It has been recalled once.

Decided
Dec 15, 2006
Review
114 days
Class
Class II
Product code
DJG
Recalls
1
Predicates
2

Its longest line of descent runs back 1 generation to K011289 (2001). 2 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2002: 1 other ancestorK011289 · Emit II Plus Opiate Assay, Model OSR9B229 · Syva Co., Dade Behring, Inc. · 200120002002200420062008
Other ancestors, by generation 1 generation back: 1 other ancestor K011289 · Emit II Plus Opiate Assay, Model OSR9B229 · Syva Co., Dade Behring, Inc. · 20010
  1. Root · 1 generation backK011289 Emit II Plus Opiate Assay, Model OSR9B229Syva Co., Dade Behring, Inc. · 2001
  2. This deviceK062460 Vitros Chemistry Products OP Reagent, Calibrator Kit 26 and Dat…Ortho-Clinical Diagnostics, Inc. · 2006
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K011289Emit II Plus Opiate Assay, Model OSR9B229Syva Co., Dade Behring, Inc.2001—
0 (this device)K062460Vitros Chemistry Products OP Reagent, Calibrator Kit 26 and Dat Performance…Ortho-Clinical Diagnostics, Inc.2006—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200210
FIG. 1 — Lineage of K062460, 2001–2006Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DJG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DJG device

  1. #1 K232659 Dochek® Multi-Drug Urine Test Cup Rx, Dochek®…score 17.6 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K011289Emit II Plus Opiate Assay, Model OSR9B229Syva Co., Dade Behring, Inc.2001—
*K022707Liquichek Urine Toxicology Control Level S1E, Model 438; Level S2E, Model 439
Summary page 3
…rformance Verifiers are substantially equivalent to the EMIT® II Plus Opiate assay devices: (KO1 1289) and Bio-Rad Liquicheck Tm Urine Toxicology Controls (K022707) predicate devices which were cleared by the FDA for IVD use.…
Bio-Rad2002—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 12, 2025Class IIZ-2073-2025Under Investigation by firmLots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than expected quality control (QC) results. If the customer is unable to get passing quality control results, then the customer would be unable to…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 2 ancestors has a recall.

How does its category compare?

All DJG devices

DJG Enzyme Immunoassay, Opiates: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DJG

3in 2025

2007: 22008: 22009: 22010: 62011: 62012: 52013: 52014: 92015: 122016: 22017: 42018: 52019: 32020: 52021: 32022: 22023: 102024: 52025: 32026 (partial year): 3
2007high 12 (2015) · low 2 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DJG

31in 2025

2010: 187 days median2011: 233 days median2012: 95 days median2013: 60 days median2014: 62 days median2015: 141 days median2017: 101 days median2018: 134 days median2019: 235 days median2020: 30 days median2021: 422 days median2023: 140 days median2024: 150 days median2025: 31 days median2026 (partial year): 157 days median
2007high 422 (2021) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DJG

2in 2025

2011: 12019: 12025: 2
2007high 2 (2025) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DJG

1in 2025

2011: 42012: 12013: 42015: 32016: 22017: 12018: 12021: 22022: 12024: 32025: 1
2007high 4 (2011) · low 0 (2007)2025

MAUDE reports naming product code DJG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20082—00
20092—00
2010618700
2011623314
201259501
201356004
201496200
20151214103
20162—02
2017410101
2018513401
2019323510
202053000
2021342202
20222—01
20231014000
2024515003
202533121
202631570—

What is it indicated for?

For in vitro diagnostic use only. VITROS Chemistry Products OP Reagent is used on VITROS 5,1 FS Chemistry Systems for the semi-quantitative or qualitative determination of opiates (OP) in human urine using a cutoff of either 300 ng/mL or 2000 ng/mL. Measurements obtained with the VITROS OP method are used in the diagnosis and treatment of opiate use or overdose. The VITROS Chemistry Products OP assay is intended for use by professional laboratory…

Read the full statement (190 words)

For in vitro diagnostic use only. VITROS Chemistry Products OP Reagent is used on VITROS 5,1 FS Chemistry Systems for the semi-quantitative or qualitative determination of opiates (OP) in human urine using a cutoff of either 300 ng/mL or 2000 ng/mL. Measurements obtained with the VITROS OP method are used in the diagnosis and treatment of opiate use or overdose.

The VITROS Chemistry Products OP assay is intended for use by professional laboratory personnel. It provides only a preliminary test result. A more specific alternative chemical method must be used to confirm a result obtained with this assay. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when evaluating a preliminary positive result.

For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 26 is used to calibrate VITROS 5,1 FS Chemistry Systems for the qualitative or semi-quantitative measurement of drugs of abuse.

For in vitro diagnostic use only. VITROS Chemistry Products DAT Performance Verifiers are assayed controls used to monitor performance of urine drugs of abuse screening assays on VITROS 5,1 FS Chemistry Systems.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K062460: Vitros Chemistry Products OP Reagent, Calibrator Kit 26 and Dat Performance Verifiers I, II and Iii.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K062460

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026