Stryker Navigation System - Cranial Module, Model 6000-650-000

K062640510(k) TraditionalCleared Dec 14, 200699-day reviewClass II21 CFR 882.4560Recalls 19

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Stryker Instruments · HAW Neurological Stereotaxic Instrument

Cleared on Dec 14, 2006 as substantially equivalent to K993239 (Stryker Corp., 2000) and 2 other predicates after a 99-day review. Its family tree reaches back 1 generation to 2000. 1 of the 54 devices in its family, counting itself, has a Class I recall. 50 later devices descend from it. It has been recalled 19 times.

Decided
Dec 14, 2006
Review
99 days
Class
Class II
Product code
HAW
Recalls
19
Predicates
3

Its longest line of descent runs back 1 generation to K993239 (2000). 3 ancestors in all. 50 later devices descend from it, 5 citing it directly. 1 of the 54 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2005: 1 other ancestor2000: 1 other ancestor2014: 2 descendants cleared2013: 1 descendant cleared2010: 1 descendant cleared2015: 1 descendant cleared2009: 1 descendant cleared2017: 3 descendants cleared2016: 5 descendants cleared, 1 with a Class I recall12018: 2 descendants cleared2019: 3 descendants cleared2022: 3 descendants cleared2023: 6 descendants cleared2020: 1 descendant cleared2021: 1 descendant cleared2024: 9 descendants cleared2025: 3 descendants cleared2026: 8 descendants clearedK993239 · Stryker Navigation System - Neuro Module, Model 6000-XXX-XXX · Stryker Corp. · 2000 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 5 descendants2 generations forward: 8 descendants, 1 with a Class I recall13 generations forward: 11 descendants4 generations forward: 8 descendants5 generations forward: 8 descendants6 generations forward: 8 descendants7 generations forward: 2 descendants K993239 · Stryker Navigation System - Neuro Module, Model 6000-XXX-XXX · Stryker Corp. · 2000 · recalled0+5
  1. Root · 1 generation backK993239 Stryker Navigation System - Neuro Module, Model 6000-XXX-XXXStryker Corp. · 2000 · Recalled
  2. This deviceK062640 Stryker Navigation System - Cranial Module, Model 6000-650-000Stryker Instruments · 2006
  3. Descendant · cites it directly · 26 of its ownK131214 Cranial Map NeuroStryker Corporate · 2013
  4. Descendant · 2 generations on · 22 of its ownK141941 Spinemask Tracker, Spinemap 3D 3.0-Software,Spinemap 3D…Stryker Corporate · 2014
  5. Descendant · cites it directly · 20 of its ownK141551 Profess Software, Profess Curved 70 (Degrees) Suction, Profess…Stryker Corporate · 2014 · Recalled
  6. Descendant · 2 generations on · 17 of its ownK153240 Styrker OrthoMap Express Knee SystemStryker Corporate · 2016
  7. Descendant · 3 generations on · 16 of its ownK162341 Stryker OrthoMap Precision Knee systemStryker Leibinger GmbH & Co KG · 2016

Showing 5 of 50 descendants. See all descendants

Descendants cleared per year

2009: 12010: 12013: 12014: 22015: 12016: 52017: 32018: 22019: 32020: 12021: 12022: 32023: 62024: 92025: 32026: 8
2009high 9 (2024) · low 0 (2011)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 50 · See all
  • K131214Cranial Map NeuroStryker Corporate · 2013 · cites it directly26descendants
  • K141941Spinemask Tracker, Spinemap 3D 3.0-Software,Spinemap 3D 3.0-Upgreade, Spinemap 3D…Stryker Corporate · 2014 · 2 generations on22descendants
  • K141551Profess Software, Profess Curved 70 (Degrees) Suction, Profess Curved 70 (Degrees)…Stryker Corporate · 2014 · cites it directly20descendants
  • K153240Styrker OrthoMap Express Knee SystemStryker Corporate · 2016 · 2 generations on17descendants
  • K162341Stryker OrthoMap Precision Knee systemStryker Leibinger GmbH & Co KG · 2016 · 3 generations on16descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K993239Stryker Navigation System - Neuro Module, Model 6000-XXX-XXXStryker Corp.20006
0 (this device)K062640Stryker Navigation System - Cranial Module, Model 6000-650-000Stryker Instruments2006—
+1K131214Cranial Map NeuroStryker Corporate2013—
+2K141941Spinemask Tracker, Spinemap 3D 3.0-Software,Spinemap 3D 3.0-Upgreade, Spinemap…Stryker Corporate2014—
+1K141551Profess Software, Profess Curved 70 (Degrees) Suction, Profess Curved 70…Stryker Corporate20141
+2K153240Styrker OrthoMap Express Knee SystemStryker Corporate2016—
+3K162341Stryker OrthoMap Precision Knee systemStryker Leibinger GmbH & Co KG2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201010
201310
201420
201510
201651
201730
201820
201930
202010
202110
202230
202360
202490
202530
202680
FIG. 1 — Lineage of K062640, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HAW device

  1. #1 K253856 Geniant Cranial (Navigated Neurosurgical…score 15.0 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K993239Stryker Navigation System - Neuro Module, Model 6000-XXX-XXX
Summary page 1
…gation System - Cranial Module Common Name: Navigation System Classification Name: Stereotaxic Instruments Equivalent to: K993239 Stryker Navigation - Neuro Module K002732 Stryker Navigation - ENT Module…
Stryker Corp.20006
*K002732Stryker Navigation System - Ent Module
Summary page 1
…System Classification Name: Stereotaxic Instruments Equivalent to: K993239 Stryker Navigation - Neuro Module K002732 Stryker Navigation - ENT Module K050438 Medtronic StealthStation System Device Description:…
Stryker Corp.2000—
*K050438Stealthstation System Update
Summary page 1
…alent to: K993239 Stryker Navigation - Neuro Module K002732 Stryker Navigation - ENT Module K050438 Medtronic StealthStation System Device Description: Stryker Navigation System - Cranial Module is part of the product…
Medtronic Surgical Navigation Technologies200511

Also cleared against K993239: 2 other devices →

What descends from it? (50)

See all 50

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K131214Cranial Map NeuroStryker Corporate201326
K141551Profess Software, Profess Curved 70 (Degrees) Suction, Profess Curved 70 (Degrees)…Stryker Corporate201420
K142024Brightmatter Navigation SystemSynaptive Medical, Inc.201512
K083009Ortho-Map-Oncology Navigation Software, Models 6005-650-000, 6005-651-000Stryker Leibinger GmbH & Co KG20091
K093206Orthomap 3D 1.1 ModuleStryker Corp.20100

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 10, 2010Not classifiedZ-1419-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1410-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1418-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1405-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1404-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1407-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1417-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1409-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1408-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1406-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1412-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1411-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1415-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1414-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1416-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1413-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1402-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1403-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…
Nov 10, 2010Not classifiedZ-1420-2011Software designWhen users imports more than one image set and changes the viewing direction for both images, the orientation of every other image series might be stored incorrectly, as a mirror image. This may lead to insufficient or inaccurate…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K153281BrightMatter Guide with Surface Trace RegistrationSynaptive Medical, Inc. · 2016 · 2 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • 2 of the 3 ancestors have a recall of any class.
  • Of its descendants, K131214 (Stryker Corporate, 2013) has gone furthest, with 26 descendants of its own.
  • 2 other devices also cite its predicate K993239.

How does its category compare?

All HAW devices

HAW Neurological Stereotaxic Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · HAW

9in 2025

2007: 152008: 122009: 82010: 62011: 42012: 82013: 32014: 52015: 102016: 92017: 72018: 62019: 102020: 52021: 92022: 72023: 142024: 72025: 92026 (partial year): 5
2007high 15 (2007) · low 3 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HAW

163in 2025

2007: 107 days median2008: 112 days median2009: 221 days median2010: 138 days median2011: 66 days median2012: 178 days median2013: 172 days median2014: 238 days median2015: 232 days median2016: 175 days median2017: 226 days median2018: 112 days median2019: 147 days median2020: 88 days median2021: 168 days median2022: 128 days median2023: 142 days median2024: 107 days median2025: 163 days median2026 (partial year): 162 days median
2007high 238 (2014) · low 66 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HAW

2in 2025

2007: 22008: 102009: 62010: 282011: 52012: 42013: 6, 1 Class I2014: 52015: 32016: 52017: 22018: 2, 2 Class I2019: 4, 1 Class I2020: 12021: 4, 2 Class I2022: 12023: 4, 1 Class I2024: 2, 2 Class I2025: 22026 (partial year): 6
2007high 28 (2010) · low 1 (2020)2026

Class I (9 since 2007) Class II/III not classified

Adverse event reports per year · HAW

3,425in 2025

2007: 492008: 532009: 962010: 2992011: 1,7252012: 9862013: 1,2902014: 1,6062015: 1,7672016: 1,8282017: 3,5052018: 4,4392019: 6,0622020: 3,4802021: 2,8272022: 1,9292023: 2,5102024: 3,7552025: 3,425
2007high 6,062 (2019) · low 49 (2007)2025

MAUDE reports naming product code HAW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715107249
2008121121053
20098221696
2010613828299
201146651,725
201281784986
201331726 (1)1,290
2014523851,606
20151023231,767
2016917551,828
2017722623,505
201861122 (2)4,439
2019101474 (1)6,062
202058813,480
202191684 (2)2,827
2022712811,929
2023141424 (1)2,510
202471072 (2)3,755
2025916323,425
202651626—

What is it indicated for?

The system should be operated only by trained personnel such as surgeons and clinic staff. The Cranial Navigation system supports, but is not limited to, the following surgical procedures: ENT Procedures * Endoscopic Sinus Surgery (ESS) * Intranasal procedures * Ear implant procedures Neuro Procedures * Cranial biopsies * Puncture of abscesses * Craniotomies * Craniectomies * Resection of tumors and other lesions * Removal of foreign objects * Skull base…

Read the full statement (123 words)

The system should be operated only by trained personnel such as surgeons and clinic staff. The Cranial Navigation system supports, but is not limited to, the following surgical procedures:

ENT Procedures

* Endoscopic Sinus Surgery (ESS)

* Intranasal procedures

* Ear implant procedures

Neuro Procedures

* Cranial biopsies

* Puncture of abscesses

* Craniotomies

* Craniectomies

* Resection of tumors and other lesions

* Removal of foreign objects

* Skull base procedures

* Transnasal neurosurgical procedures

* Transphenoidal pituitary surgery

* Shunt placement, including pediatric shunt placement

* Placement of electrodes for recording, stimulation and lesion generation

* Craniofacial procedures

* Skull reconstruction procedures

* Orbital cavity reconstruction procedures

Contraindications

* Surgical situation where increasing surgical time may be detrimental to the patient.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K062640: Stryker Navigation System - Cranial Module, Model 6000-650-000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K062640

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026