510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Brightmatter Navigation System

K142024510(k) TraditionalCleared Apr 2, 2015251-day reviewClass II21 CFR 882.4560Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Synaptive Medical, Inc. · HAW Neurological Stereotaxic Instrument

Cleared on Apr 2, 2015 as substantially equivalent to K062640 (Stryker Instruments, 2006) after a 251-day review. Its family tree reaches back 2 generations to 2000. 1 of the 17 devices in its family, counting itself, has a Class I recall. 12 later devices descend from it. It has been recalled once.

Decided
Apr 2, 2015
Review
251 days
Class
Class II
Product code
HAW
Recalls
1
Predicates
1

Its longest line of descent runs back 2 generations to K993239 (2000). 4 ancestors in all. 12 later devices descend from it, 4 citing it directly. 1 of the 17 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2005: 1 other ancestor2000: 1 other ancestor2016: 3 descendants cleared, 1 with a Class I recall12018: 1 descendant cleared2017: 1 descendant cleared2019: 2 descendants cleared2024: 3 descendants cleared2023: 1 descendant cleared2026: 1 descendant clearedK993239 · Stryker Navigation System - Neuro Module, Model 6000-XXX-XXX · Stryker Corp. · 2000 · recalledK062640 · Stryker Navigation System - Cranial Module, Model 6000-650-000 · Stryker Instruments · 2006 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors 1 generation forward: 4 descendants, 1 with a Class I recall12 generations forward: 3 descendants3 generations forward: 2 descendants4 generations forward: 1 descendant5 generations forward: 1 descendant6 generations forward: 1 descendant K062640 · Stryker Navigation System - Cranial Module, Model 6000-650-000 · Stryker Instruments · 2006 · recalledK993239 · Stryker Navigation System - Neuro Module, Model 6000-XXX-XXX · Stryker Corp. · 2000 · recalled0+5
  1. Root · 2 generations backK993239 Stryker Navigation System - Neuro Module, Model 6000-XXX-XXXStryker Corp. · 2000 · Recalled
  2. 1 generation backK062640 Stryker Navigation System - Cranial Module, Model 6000-650-000Stryker Instruments · 2006 · Recalled
  3. This deviceK142024 Brightmatter Navigation SystemSynaptive Medical, Inc. · 2015
  4. Descendant · cites it directly · 6 of its ownK181041 7D Surgical System Cranial Application7D Surgical, Inc. · 2018
  5. Descendant · 2 generations on · 5 of its ownK192945 7D Surgical System Cranial Biopsy and Ventricular Catheter Placement…7D Surgical, Inc. · 2019
  6. Descendant · 3 generations on · 3 of its ownK230661 Zeta Cranial Navigation SystemZeta Surgical · 2023
  7. Descendant · 4 generations on · 2 of its ownK233903 Zeta Cranial Navigation System (ZNS131-US)Zeta Surgical, Inc. · 2024
  8. Descendant · cites it directly · 2 of its ownK153281 BrightMatter Guide with Surface Trace RegistrationSynaptive Medical, Inc. · 2016 · Recalled (Class I)

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2016: 32017: 12018: 12019: 22023: 12024: 32026: 1
2016high 3 (2016) · low 0 (2020)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 12 · See all
  • K1810417D Surgical System Cranial Application7D Surgical, Inc. · 2018 · cites it directly6descendants
  • K1929457D Surgical System Cranial Biopsy and Ventricular Catheter Placement Application7D Surgical, Inc. · 2019 · 2 generations on5descendants
  • K230661Zeta Cranial Navigation SystemZeta Surgical · 2023 · 3 generations on3descendants
  • K233903Zeta Cranial Navigation System (ZNS131-US)Zeta Surgical, Inc. · 2024 · 4 generations on2descendants
  • K153281BrightMatter Guide with Surface Trace RegistrationSynaptive Medical, Inc. · 2016 · cites it directly2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K993239Stryker Navigation System - Neuro Module, Model 6000-XXX-XXXStryker Corp.20006
−1K062640Stryker Navigation System - Cranial Module, Model 6000-650-000Stryker Instruments200619
0 (this device)K142024Brightmatter Navigation SystemSynaptive Medical, Inc.2015—
+1K1810417D Surgical System Cranial Application7D Surgical, Inc.2018—
+2K1929457D Surgical System Cranial Biopsy and Ventricular Catheter Placement Application7D Surgical, Inc.2019—
+3K230661Zeta Cranial Navigation SystemZeta Surgical2023—
+4K233903Zeta Cranial Navigation System (ZNS131-US)Zeta Surgical, Inc.2024—
+1K153281BrightMatter Guide with Surface Trace RegistrationSynaptive Medical, Inc.2016Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201631
201710
201810
201920
202310
202430
202610
FIG. 1 — Lineage of K142024, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HAW device

  1. #1 K253856 Geniant Cranial (Navigated Neurosurgical…score 15.0 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K062640Stryker Navigation System - Cranial Module, Model 6000-650-000
Summary page 4
…eotaxic instrument (21 CFR 882.4560) Regulatory Class: II Product Code: HAW III. PREDICATE DEVICE Stryker Navigation System – Cranial Module by Stryker Corporation, K062640 Date cleared: December 14, 2006 IV. DEVICE DESCRIPTION The subject device, BrightMatter Navigation system, is a planning an…
Stryker Instruments200619

Also cleared against K062640: 4 other devices →

What descends from it? (12)

See all 12

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K1810417D Surgical System Cranial Application7D Surgical, Inc.20186
K153281BrightMatter Guide with Surface Trace Registration Class ISynaptive Medical, Inc.20162
K151156Fiagon Navigation SystemFiagon GmbH20161
K160523BrightMatter Guide with BrightMatter PointerSynaptive Medical, Inc.20160

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 4, 2015Class IIZ-0480-2016Component design/selectionOut of tolerance for radio frequency emissions. At the 150-1000MHz frequency, the testing indicated the BrightMatter Navigation system was up to 20dB uV/meter higher than the applicable IEC 60601-1-2:2007 (Ed3.0) standard specification.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K153281BrightMatter Guide with Surface Trace RegistrationSynaptive Medical, Inc. · 2016 · Cites it directly · ◆ in the 2016 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 15 years and passes through 3 companies; a new generation was cleared every 8.3 years on median.
  • 3 of the 4 ancestors have a recall of any class.
  • Of its descendants, K181041 (7D Surgical, Inc., 2018) has gone furthest, with 6 descendants of its own.
  • 4 other devices also cite its predicate K062640.

How does its category compare?

All HAW devices

HAW Neurological Stereotaxic Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · HAW

9in 2025

2007: 152008: 122009: 82010: 62011: 42012: 82013: 32014: 52015: 102016: 92017: 72018: 62019: 102020: 52021: 92022: 72023: 142024: 72025: 92026 (partial year): 5
2007high 15 (2007) · low 3 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HAW

163in 2025

2007: 107 days median2008: 112 days median2009: 221 days median2010: 138 days median2011: 66 days median2012: 178 days median2013: 172 days median2014: 238 days median2015: 232 days median2016: 175 days median2017: 226 days median2018: 112 days median2019: 147 days median2020: 88 days median2021: 168 days median2022: 128 days median2023: 142 days median2024: 107 days median2025: 163 days median2026 (partial year): 162 days median
2007high 238 (2014) · low 66 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HAW

2in 2025

2007: 22008: 102009: 62010: 282011: 52012: 42013: 6, 1 Class I2014: 52015: 32016: 52017: 22018: 2, 2 Class I2019: 4, 1 Class I2020: 12021: 4, 2 Class I2022: 12023: 4, 1 Class I2024: 2, 2 Class I2025: 22026 (partial year): 6
2007high 28 (2010) · low 1 (2020)2026

Class I (9 since 2007) Class II/III not classified

Adverse event reports per year · HAW

3,425in 2025

2007: 492008: 532009: 962010: 2992011: 1,7252012: 9862013: 1,2902014: 1,6062015: 1,7672016: 1,8282017: 3,5052018: 4,4392019: 6,0622020: 3,4802021: 2,8272022: 1,9292023: 2,5102024: 3,7552025: 3,425
2007high 6,062 (2019) · low 49 (2007)2025

MAUDE reports naming product code HAW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715107249
2008121121053
20098221696
2010613828299
201146651,725
201281784986
201331726 (1)1,290
2014523851,606
20151023231,767
2016917551,828
2017722623,505
201861122 (2)4,439
2019101474 (1)6,062
202058813,480
202191684 (2)2,827
2022712811,929
2023141424 (1)2,510
202471072 (2)3,755
2025916323,425
202651626—

What is it indicated for?

BrightMatter Navigation System is intended as a planning and intraoperative guidance system to enable open and percutaneous computer assisted surgery. The system is indicated for medical conditions requiring neurosurgical cranial procedures where the use of computer assisted planning and surgery may be appropriate. The system can be used for intra-operative guidance where a reference to a rigid anatomical structure can be identified.

The system should be operated only by trained personnel such as surgeons and other clinic staff.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K142024: Brightmatter Navigation System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K142024

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026