Vitros 5,1 FS Chemistry System with Engen Laboratory Automation System

K063144510(k) SpecialCleared Nov 13, 200628-day reviewClass I21 CFR 862.2160Recalls 35

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ortho-Clinical Diagnostics, Inc. · JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Cleared on Nov 13, 2006 after a 28-day review. The FDA summary names its predicate without a number. 3 later devices descend from it. It has been recalled 35 times.

Decided
Nov 13, 2006
Review
28 days
Class
Class I
Product code
JJE
Recalls
35
Predicates
0

3 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2016: 1 descendant cleared2018: 1 descendant cleared2019: 1 descendant cleared20052010201520202025
  1. This deviceK063144 Vitros 5,1 FS Chemistry System with Engen Laboratory Automation SystemOrtho-Clinical Diagnostics, Inc. · 2006
  2. Descendant · cites it directly · 2 of its ownK160495 VITROS® Automation SolutionsOrtho-Clinical Diagnostics, Inc. · 2016 · Recalled
  3. Descendant · 2 generations on · 1 of its ownK182063 VITROS Chemistry Products CRBM Slides, VITROS Chemistry Products…Ortho-Clinical Diagnostics, Inc. · 2018 · Recalled
  4. Descendant · 3 generations on · 0 of its ownK190326 VITROS XT Chemistry Products UREA-CREA SlidesOrtho-Clinical Diagnostics, Inc. · 2019

Most influential descendants

Showing 3 of 3 · See all
  • K160495VITROS® Automation SolutionsOrtho-Clinical Diagnostics, Inc. · 2016 · cites it directly2descendants
  • K182063VITROS Chemistry Products CRBM Slides, VITROS Chemistry Products CREA Slides, VITROS…Ortho-Clinical Diagnostics, Inc. · 2018 · 2 generations on1descendant
  • K190326VITROS XT Chemistry Products UREA-CREA SlidesOrtho-Clinical Diagnostics, Inc. · 2019 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K063144Vitros 5,1 FS Chemistry System with Engen Laboratory Automation SystemOrtho-Clinical Diagnostics, Inc.2006—
+1K160495VITROS® Automation SolutionsOrtho-Clinical Diagnostics, Inc.20161
+2K182063VITROS Chemistry Products CRBM Slides, VITROS Chemistry Products CREA Slides…Ortho-Clinical Diagnostics, Inc.201820
+3K190326VITROS XT Chemistry Products UREA-CREA SlidesOrtho-Clinical Diagnostics, Inc.2019—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
201810
201910
FIG. 1 — Lineage of K063144, 2006–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JJE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—VITROS dHDL assay performed on the VITROS 5,1 FS Chemistry System
Named without a submission number
—

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K160495VITROS® Automation SolutionsOrtho-Clinical Diagnostics, Inc.20162

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 8, 2019Class IIZ-2181-2019Software designPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System
Jul 8, 2019Class IIZ-2184-2019Software designPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System
Jul 8, 2019Class IIZ-2183-2019Software designPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System
Jul 8, 2019Class IIZ-2182-2019Software designPotential Unsuccessful Calibration Due to Software Anomaly in VITROS System
Oct 19, 2018Class IIZ-0651-2019Software Manufacturing/Software DeploymentSoftware anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.
Oct 19, 2018Class IIZ-0653-2019Software Manufacturing/Software DeploymentSoftware anomaly may incorrectly extend the on-analyzer stability time for a loaded VITROS Reagent.
Apr 26, 2017Class IIZ-0254-2018Software design (manufacturing process)A software anomaly with TCA Software V4.2 may potentially cause a delay in reporting of results due to patient samples being routed to an unintended location when samples are reintroduced back onto the enGEN track
Mar 8, 2017Class IIZ-2077-2017Software designSoftware anomaly; Thermo-Fisher Scientific initially discovered and Ortho-Clinical Diagnostics, subsequently, confirmed a software anomaly that may potentially result in miss-associated sample IDs involving the communication between enGen…
Sep 7, 2016Class IIZ-0300-2017Software designOrtho confirmed that it is possible for the mis-association of sample identification and assay results due to an error with Thermo Fishers centrifuge module. It is possible for mis-associated assay results to be reported from the…
May 16, 2016Class IIIZ-2143-2016Software designIf the user selects the Comment or Interpretation option (in the enGEN .gsb file), and the numerical results fall between Negative and Reactive, the IM incorrectly displays the interpretive text as Borderline and sends the incorrect text…
Mar 17, 2016Class IIZ-1569-2016Component design/selectionThe electrical cable that supplies power to the cap vibratory feeder of the Recapper Module may loosen from its correct position above the vibration feeder support plate and migrate to under the support plate. If this occurs, the uncapped…
Dec 21, 2015Class IIZ-1075-2016Error in labelingIncreased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954. A trend of complaints regarding customers actions following U90-382 or 6LU condition…
Dec 21, 2015Class IIZ-1073-2016Error in labelingIncreased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954. A trend of complaints regarding customers actions following U90-382 or 6LU condition…
Nov 3, 2015Class IIZ-0643-2016Software designSoftware anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems…
Nov 3, 2015Class IIZ-0642-2016Software designSoftware anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems…
Sep 10, 2015Class IIZ-0231-2016Software designSoftware Anomaly during ADD Installation on VITROS 5,1 FS Chemistry Systems using Software Versions 2.2.1 through 2.8. The user modified parameters would not be saved as expected, but remained changed to the default values obtained from…
May 6, 2015Class IIZ-1724-2015Component change controlRecapper Caps may fall off of, or be knocked from sample tubes placed in the storage racks of entry/exit modules. The gripper may grab a tube at the wrong height and drop the tube in the Rack Entry/Exit or Rack Exit module. These…
Apr 6, 2015Class IIZ-1729-2015Software designSoftware anomaly regarding urine samples that require acidified pretreatment. The software has allowed urine samples for assays that require acidified pretreatment to be metered from the same sample container as assays that required no…
Mar 10, 2015Class IIZ-1521-2015Software in the Use EnvironmentSoftware Anomaly; It is possible for the device to process samples with a cartridge other than the intended cartridge, potentially leading to erroneous patient results. If this anomaly occurred previously, an indication would be a series…
Nov 19, 2014Class IIZ-0967-2015Software designSoftware Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.
Nov 19, 2014Class IIZ-0969-2015Software designSoftware Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.
Nov 19, 2014Class IIZ-0968-2015Software designSoftware Anomaly: the firm has identified an anomaly with VITROS System Software Version 3.1 and below, and determined that the software may not properly identify an expired calibration.
Sep 29, 2014Class IIIZ-1004-2015Software Manufacturing/Software DeploymentCalibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Products Calibrator Kit 2.
Sep 8, 2014Class IIZ-0077-2015Software designOrtho Clinical Diagnostics (OCD) identified an anomaly with Software Version 3.0 and below on the VITROS 4600 Chemistry Systems and VITROS Integrated Systems. Internal testing confirmed that when using calibrator barcode labels supplied…
Jul 7, 2014Class IIZ-2140-2014Software in the Use EnvironmentOrtho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and below. Internal testing confirmed that under very specific conditions, incorrect patient demographics were associated with a sample…
Apr 17, 2014Class IIZ-1741-2014Software Design ChangeSoftware anomaly: potential buffer timeout issues when using enGen (TM) Laboratory Automation Systems Configured with TCAutomation (TM) Software v3.5.
Apr 10, 2014Class IIZ-1736-2014Software designPotential for biased results when a Multiple Window (MW) code is associated with results generated by the VITROS 4600 Chemistry System, the VITROS 5600 Integrated System and the VITROS 5,1 Chemistry System. MW codes may be associated with…
Apr 10, 2014Class IIZ-1738-2014Software designSoftware anomaly: potential for delay in the reporting of patient sample test results when using enGen with TCAutomation Software v2.6 , v3.2 and v3.5 Configured with Bypass Modules for ADVIA Centaur, Abbott Architect, VITROS and/or Tosoh…
Mar 20, 2014Class IIZ-1677-2014Software designSoftware Anomaly: Ortho Clinical Diagnostics (OCD) has become aware that if a sample requires follow-up action, such as dilution or retesting due to a mechanical issue, it is possible for that sample to be routed to the sample storage…
Jun 6, 2013Class IIZ-1948-2013Nonconforming Material/ComponentOrtho Clinical Diagnostics is recalling CM/RT wear pads which are a component in the CM/RT incubator due to becoming worn before their replacement.
Feb 7, 2013Class IIZ-1106-2013Nonconforming Material/ComponentOrtho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROS¿ Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an internal…
Mar 15, 2012Not classifiedZ-1471-2012Nonconforming Material/ComponentOrtho-Clinical Diagnostics is recalling VITROS Chemistry Systems due to possible internal hard drive failure.
Mar 15, 2012Not classifiedZ-1472-2012Nonconforming Material/ComponentOrtho-Clinical Diagnostics is recalling VITROS Chemistry Systems due to possible internal hard drive failure.
Apr 28, 2011Not classifiedZ-0914-2012Component design/selectionOrtho Clinical Diagnostics received complaints of centrifuge buckets disconnecting from their supports during centrifugation when using the Hettich Rotanta 46 RSC Robotic Centrifuges, Models 4815, 4816 and 4817, that may be configured…
May 17, 2007Not classifiedZ-1167-2007OtherSoftware design. The firm received a customer complaint regarding a qualitative positive THC (Cannabinoids) result obtained from a THC quality control fluid that was known to be negative when using VITROS Chemistry Products THC Reagent. A…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K160495 (Ortho-Clinical Diagnostics, Inc., 2016) has gone furthest, with 2 descendants of its own.

How does its category compare?

All JJE devices

JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JJE

1in 2019

2008: 12011: 22012: 12013: 12014: 32019: 1
2007high 3 (2014) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JJE

224in 2014

2014: 224 days median
2007high 224 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JJE

6in 2025

2007: 202008: 202009: 142010: 322011: 252012: 162013: 212014: 142015: 152016: 92017: 182018: 172019: 72020: 72021: 62023: 122024: 62025: 62026 (partial year): 5
2007high 32 (2010) · low 0 (2022)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JJE

4,400in 2025

2007: 2,1812008: 2,1592009: 3,4402010: 3,2732011: 11,6062012: 3,7072013: 2,4892014: 2,8412015: 2,7932016: 2,5802017: 1,8692018: 4,5902019: 3,9482020: 3,9862021: 3,9202022: 3,9982023: 3,7832024: 4,2212025: 4,400
2007high 11,606 (2011) · low 1,869 (2017)2025

MAUDE reports naming product code JJE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—202,181
20081—202,159
20090—143,440
20100—323,273
20112—2511,606
20121—163,707
20131—212,489
20143224142,841
20150—152,793
20160—92,580
20170—181,869
20180—174,590
20191—73,948
20200—73,986
20210—63,920
20220—03,998
20230—123,783
20240—64,221
20250—64,400
20260—5—

What is it indicated for?

The VITROS 5,1 FS Chemistry System with enGenTM Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.)

The VITROS Chemistry Products dHDL Reagent Pack, VITROS Chemistry Products Calibrator Kit 19 and VITROS Chemistry Products FS Calibrator 1 are used for the quantitative measurement of HDL cholesterol in serum or plasma.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K063144: Vitros 5,1 FS Chemistry System with Engen Laboratory Automation System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K063144

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026