Filterwire EZ Embolic Protection System (3.5 MM to 5.5 MM)

K063313510(k) TraditionalCleared Dec 5, 200633-day reviewClass II21 CFR 870.1250Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Boston Scientific Corp · NTE Temporary Carotid Catheter For Embolic Capture

Cleared on Dec 5, 2006 as substantially equivalent to K032884 (Boston Scientific Epi, 2004) and 2 other predicates after a 33-day review. Its family tree reaches back 5 generations to 2001. 1 of the 13 devices in its family, counting itself, has a Class I recall. 4 later devices descend from it. It has been recalled 6 times.

Decided
Dec 5, 2006
Review
33 days
Class
Class II
Product code
NTE
Recalls
6
Predicates
3

Its longest line of descent runs back 5 generations to K003992 (2001). 8 ancestors in all. 4 later devices descend from it, 1 citing it directly. 1 of the 13 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor2002: 2 other ancestors2015: 1 descendant cleared2018: 1 descendant cleared2020: 1 descendant cleared2021: 1 descendant cleared, 1 with a Class I recall1K003992 · Percusurge Guardwire Temporary Occlusion and Aspiration System · Percu Surge, Inc. · 2001K023691 · Boston Scientific Filterwire EX Embolic · Boston Scientific Epi · 2003K032884 · Filter Wire EZ Embolic Protection System and EZ Bent Tip Retrieval Sheath · Boston Scientific Epi · 2004K051984 · Filterwire EZ (2.25 MM - 3.5 MM) Embolic Protection System, Models 20120-190, 20120-300 · Boston Scientific Corp · 2006K061332 · Filterwire EZ Embolic Protection System (2.25 MM-3.5 MM) · Boston Scientific Corp · 2006200020052010201520202025
  1. Root · 5 generations backK003992 Percusurge Guardwire Temporary Occlusion and Aspiration SystemPercu Surge, Inc. · 2001
  2. 4 generations backK023691 Boston Scientific Filterwire EX EmbolicBoston Scientific Epi · 2003
  3. 3 generations backK032884 Filter Wire EZ Embolic Protection System and EZ Bent Tip Retrieval…Boston Scientific Epi · 2004
  4. 2 generations backK051984 Filterwire EZ (2.25 MM - 3.5 MM) Embolic Protection System, Models…Boston Scientific Corp · 2006
  5. 1 generation backK061332 Filterwire EZ Embolic Protection System (2.25 MM-3.5 MM)Boston Scientific Corp · 2006
  6. This deviceK063313 Filterwire EZ Embolic Protection System (3.5 MM to 5.5 MM)Boston Scientific Corp · 2006
  7. Descendant · cites it directly · 3 of its ownK143570 WirionGardia Medical · 2015
  8. Descendant · 2 generations on · 2 of its ownK180023 WirionGardia Medical, Ltd. · 2018
  9. Descendant · 3 generations on · 1 of its ownK200198 WirionCardiovascular Systems, Inc. · 2020
  10. Descendant · 4 generations on · 0 of its ownK210282 WIRION Embolic Protection SystemCardiovascular Systems, Inc. · 2021 · Recalled (Class I)

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K143570WirionGardia Medical · 2015 · cites it directly3descendants
  • K180023WirionGardia Medical, Ltd. · 2018 · 2 generations on2descendants
  • K200198WirionCardiovascular Systems, Inc. · 2020 · 3 generations on1descendant
  • K210282WIRION Embolic Protection SystemCardiovascular Systems, Inc. · 2021 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K003992Percusurge Guardwire Temporary Occlusion and Aspiration SystemPercu Surge, Inc.2001—
−4K023691Boston Scientific Filterwire EX EmbolicBoston Scientific Epi2003—
−3K032884Filter Wire EZ Embolic Protection System and EZ Bent Tip Retrieval SheathBoston Scientific Epi2004—
−2K051984Filterwire EZ (2.25 MM - 3.5 MM) Embolic Protection System, Models 20120-190…Boston Scientific Corp2006—
−1K061332Filterwire EZ Embolic Protection System (2.25 MM-3.5 MM)Boston Scientific Corp2006—
0 (this device)K063313Filterwire EZ Embolic Protection System (3.5 MM to 5.5 MM)Boston Scientific Corp2006—
+1K143570WirionGardia Medical2015—
+2K180023WirionGardia Medical, Ltd.2018—
+3K200198WirionCardiovascular Systems, Inc.2020—
+4K210282WIRION Embolic Protection SystemCardiovascular Systems, Inc.2021Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200220
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201810
202010
202111
FIG. 1 — Lineage of K063313, 2001–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NTE device

  1. #1 K222694 EmPro EPS (EP4514C-190…score 6.5 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K032884Filter Wire EZ Embolic Protection System and EZ Bent Tip Retrieval Sheath
Summary page 1
…Ernbolic Protection System (FilterWire EZ System) is substantially equivalent in design and intended use (embolic protection) to the same devices legally marketed uinder K032884 (3.5 mm to 5.5 mm) and K06 1332 (2.25 mm to 3.5 mm) for use in saphenous vein bypass grafts. The FilterWire EZ System is also sub…
Boston Scientific Epi2004—
*K061332Filterwire EZ Embolic Protection System (2.25 MM-3.5 MM)
Summary page 1
…terWire EZ System) is substantially equivalent in design and intended use (embolic protection) to the same devices legally marketed uinder K032884 (3.5 mm to 5.5 mm) and K06 1332 (2.25 mm to 3.5 mm) for use in saphenous vein bypass grafts. The FilterWire EZ System is also substantially equivalent in design…
Boston Scientific Corp2006—
*K042218Accunet Embolic Protection System, RX Accunet Embolic Protection System
Summary page 1
…saphenous vein bypass grafts. The FilterWire EZ System is also substantially equivalent in design and indications for use to the RX ACCUNET"' Emnbolic Protection System (K0422 18) legally marketed as an embolic protection device for use in the carotid vasculature. Device Description: The Boston Scientific…
Guidant Corp.2004—

Also cleared against K032884: 6 other devices →

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K143570WirionGardia Medical20153

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 20, 2008Not classifiedZ-0105-2009Employee errorBoston Scientific is initiating a field correction for 48 lots/batches of the Filter Wire EZ" Embolic Protection System. Boston Scientific has determined that the Directions for Use (DFU), which accompanies each packaged device, may be…
May 20, 2008Not classifiedZ-0108-2009Employee errorBoston Scientific is initiating a field correction for 48 lots/batches of the Filter Wire EZ" Embolic Protection System. Boston Scientific has determined that the Directions for Use (DFU), which accompanies each packaged device, may be…
May 20, 2008Not classifiedZ-0110-2009Employee errorBoston Scientific is initiating a field correction for 48 lots/batches of the Filter Wire EZ" Embolic Protection System. Boston Scientific has determined that the Directions for Use (DFU), which accompanies each packaged device, may be…
May 20, 2008Not classifiedZ-0106-2009Employee errorBoston Scientific is initiating a field correction for 48 lots/batches of the Filter Wire EZ" Embolic Protection System. Boston Scientific has determined that the Directions for Use (DFU), which accompanies each packaged device, may be…
May 20, 2008Not classifiedZ-0107-2009Employee errorBoston Scientific is initiating a field correction for 48 lots/batches of the Filter Wire EZ" Embolic Protection System. Boston Scientific has determined that the Directions for Use (DFU), which accompanies each packaged device, may be…
May 20, 2008Not classifiedZ-0109-2009Employee errorBoston Scientific is initiating a field correction for 48 lots/batches of the Filter Wire EZ" Embolic Protection System. Boston Scientific has determined that the Directions for Use (DFU), which accompanies each packaged device, may be…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K210282WIRION Embolic Protection SystemCardiovascular Systems, Inc. · 2021 · 4 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 5 years and passes through 3 companies; a new generation was cleared every 1.2 years on median.
  • None of the 8 ancestors has a recall.
  • Of its descendants, K143570 (Gardia Medical, 2015) has gone furthest, with 3 descendants of its own.
  • 6 other devices also cite its predicate K032884.

How does its category compare?

All NTE devices

NTE Temporary Carotid Catheter For Embolic Capture: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · NTE

2in 2023

2007: 12008: 32009: 32010: 12011: 42012: 32014: 12015: 32016: 12018: 12019: 12020: 12021: 12022: 12023: 2
2007high 4 (2011) · low 0 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NTE

105in 2015

2008: 95 days median2009: 86 days median2011: 105 days median2012: 30 days median2015: 105 days median
2007high 105 (2011) · low 30 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NTE

3in 2023

2008: 62011: 12021: 1, 1 Class I2023: 32026 (partial year): 2, 2 Class I
2007high 6 (2008) · low 0 (2007)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · NTE

405in 2025

2007: 1192008: 3682009: 3432010: 2942011: 3042012: 2612013: 2582014: 1992015: 1482016: 1482017: 1412018: 1582019: 2332020: 2382021: 3272022: 3262023: 3542024: 3982025: 405
2007high 405 (2025) · low 119 (2007)2025

MAUDE reports naming product code NTE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—0119
20083956368
20093860343
20101—0294
201141051304
20123300261
20130—0258
20141—0199
201531050148
20161—0148
20170—0141
20181—0158
20191—0233
20201—0238
20211—1 (1)327
20221—0326
20232—3354
20240—0398
20250—0405
20260—2 (2)—

What is it indicated for?

The FilterWire EZTM Embolic Protection System is indicated for use as a guide wire and embolic protection system to contain and remove embolic material (thrombus/debris) while performing angioplasty and stenting procedures in coronary saphenous vein bypass grafts and carotid arteries. The diameter of the vessel at the site of filter loop placement should be between 2.25 mm and 5.5 mm for coronary saphenous vein bypass graft procedures and between 3.5 mm…

Read the full statement (112 words)

The FilterWire EZTM Embolic Protection System is indicated for use as a guide wire and embolic protection system to contain and remove embolic material (thrombus/debris) while performing angioplasty and stenting procedures in coronary saphenous vein bypass grafts and carotid arteries. The diameter of the vessel at the site of filter loop placement should be between 2.25 mm and 5.5 mm for coronary saphenous vein bypass graft procedures and between 3.5 mm and 5.5 mm for carotid procedures.

The safety and effectiveness of this device as an embolic protection system has not been established in the cerebral vasculature, peripheral vessels other than carotid arteries, or in treating native coronaries, including acute myocardial infarction.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K063313: Filterwire EZ Embolic Protection System (3.5 MM to 5.5 MM).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K063313

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026