510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Nova Statsensor Creatinine Hospital Meter, Control Solutions,Linearity Solutions, and Test Strips

K070068510(k) TraditionalCleared May 7, 2007119-day reviewClass II21 CFR 862.1225Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Nova Biomedical Corp. · CGL Electrode, Ion Based, Enzymatic, Creatinine

Cleared on May 7, 2007 after a 119-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 3 later devices descend from it. It has been recalled 5 times.

Decided
May 7, 2007
Review
119 days
Class
Class II
Product code
CGL
Recalls
5
Predicates
0

3 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2017: 1 descendant cleared2026: 2 descendants cleared20052010201520202025
  1. This deviceK070068 Nova Statsensor Creatinine Hospital Meter, Control…Nova Biomedical Corp. · 2007
  2. Descendant · cites it directly · 2 of its ownK171059 Nova StatSensor Creatinine Hospital Meter SystemNova Biomedical Corporation · 2017
  3. Descendant · 2 generations on · 0 of its ownK261039 i-STAT Crea cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2026
  4. Descendant · 2 generations on · 0 of its ownK251281 Nova Max Creat eGFR Monitoring SystemNova Biomedical Corporation · 2026

Most influential descendants

Showing 3 of 3 · See all
  • K171059Nova StatSensor Creatinine Hospital Meter SystemNova Biomedical Corporation · 2017 · cites it directly2descendants
  • K261039i-STAT Crea cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2026 · 2 generations on0descendants
  • K251281Nova Max Creat eGFR Monitoring SystemNova Biomedical Corporation · 2026 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K070068Nova Statsensor Creatinine Hospital Meter, Control Solutions,Linearity…Nova Biomedical Corp.2007—
+1K171059Nova StatSensor Creatinine Hospital Meter SystemNova Biomedical Corporation2017—
+2K261039i-STAT Crea cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2026—
+2K251281Nova Max Creat eGFR Monitoring SystemNova Biomedical Corporation2026—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
202620
FIG. 1 — Lineage of K070068, 2007–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CGL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CGL device

  1. #1 K261039 i-STAT Crea cartridge with the i-STAT 1 Systemscore 11.9 · 2 generations · no Class I recall
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K171059Nova StatSensor Creatinine Hospital Meter SystemNova Biomedical Corporation20172

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 10, 2013Class IIZ-0650-2014Component design/selectionRevised Battery Instructions for Removal and Expiration Date due to battery swelling and cell leakage
Feb 21, 2008Not classifiedZ-1472-2008Component design/selectionLithium batteries may fail if the glucose or creatinine meters have been dropped.
Nov 2, 2007Not classifiedZ-0725-2008Device DesignIncorrect Results - Creatinine results were lower than the laboratory reference method.
Nov 2, 2007Not classifiedZ-0726-2008Device DesignIncorrect Results - Creatinine results were lower than the laboratory reference method.
Nov 2, 2007Not classifiedZ-0724-2008Device DesignIncorrect Results - Creatinine results were lower than the laboratory reference method.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K171059 (Nova Biomedical Corporation, 2017) has gone furthest, with 2 descendants of its own.

How does its category compare?

All CGL devices

CGL Electrode, Ion Based, Enzymatic, Creatinine: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · CGL

1in 2017

2007: 12012: 12017: 12026 (partial year): 1
2007high 1 (2007) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CGL

Too few CGL 510(k)s in any year since 2007 to give a median.

Recalls per year · CGL

1in 2011

2007: 32008: 12009: 12010: 12011: 1
2007high 3 (2007) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CGL

8in 2025

2007: 172008: 32009: 62010: 122011: 42012: 82013: 62014: 12015: 42016: 22017: 142018: 112019: 252020: 192021: 202022: 112023: 52024: 72025: 8
2007high 25 (2019) · low 1 (2014)2025

MAUDE reports naming product code CGL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—317
20080—13
20090—16
20100—112
20110—14
20121—08
20130—06
20140—01
20150—04
20160—02
20171—014
20180—011
20190—025
20200—019
20210—020
20220—011
20230—05
20240—07
20250—08
20261—0—

What is it indicated for?

The Nova StatSensor Creatinine Hospital Meter is intended for in vitro diagnostic use by health care professionals and for Point-Of-Care usage for the quantitative measurement of creatinine in capillary, venous, and arterial whole blood. Creatinine measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis. Not for use in neonates. Nova StatSensor Creatinine Test Strips are intended for use only with the…

Read the full statement (217 words)

The Nova StatSensor Creatinine Hospital Meter is intended for in vitro diagnostic use by health care professionals and for Point-Of-Care usage for the quantitative measurement of creatinine in capillary, venous, and arterial whole blood. Creatinine measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis. Not for use in neonates.

Nova StatSensor Creatinine Test Strips are intended for use only with the StatSensor Creatinine Hospital Meter for quantitative tests. The Nova StatSensor Creatinine Hospital Meter is intended to quantitatively measure creatinine in whole blood. The Creatinine Meter is calibrated to provide plasma equivalent results to laboratory methods. Nova StatSensor Creatinine Test Strips are for testing outside the body (in vitro diagnostic use only).

Nova StatSensor Creatinine Control Solutions are intended for use with the Nova StatSensor Creatinine Hospital Meter and Nova StatSensor Creatinine Test Strips as a quality control check to verify the accuracy of blood creatinine test results. There are three levels of controls, (Level 1, Level 2, and Level 3). These solutions will be offered for sale separately from the meter.

Nova StatSensor Creatinine Linearity Kit solutions are used to check the linearity of the Nova StatSensor Creatinine Hospital Meter System. There are five levels of Linearity Solutions, (Level 1, Level 2, Level 3, Level 4, and Level 5).

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K070068: Nova Statsensor Creatinine Hospital Meter, Control Solutions,Linearity Solutions, and Test Strips.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K070068

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026