ABL8X7 Flex

K051968510(k) TraditionalCleared Oct 16, 2006453-day reviewClass II21 CFR 862.1225Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Radiometer Medical Aps · CGL Electrode, Ion Based, Enzymatic, Creatinine

Cleared on Oct 16, 2006 as substantially equivalent to K041874 (Radiometer Medical Aps, 2004) and 3 other predicates after a 453-day review. Its family tree reaches back 6 generations to 1986. It has been recalled 4 times.

Decided
Oct 16, 2006
Review
453 days
Class
Class II
Product code
CGL
Recalls
4
Predicates
4

Its longest line of descent runs back 6 generations to K861087 (1986). 18 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 6 generations backK861087 Avl Model 983 Electrolyte AnalyzerAvl Scientific Corp. · 1986
  2. 5 generations backK961458 Avl 9180 Electrolyte AnalyzerAvl Scientific Corp. · 1996 · Recalled
  3. 4 generations backK972673 Avl 9181 Electrolyte AnalyzerAvl Scientific Corp. · 1997 · Recalled
  4. 3 generations backK992859 ABL700 Series Analyzer with Autocheck ModuleRadiometer America, Inc. · 2000 · Recalled
  5. 2 generations backK041874 ABL800 FlexRadiometer Medical Aps · 2004 · Recalled
  6. 1 generation backK043218 ABL800 Flex with Flexq Module, Model ABL8XXRadiometer Medical Aps · 2005 · Recalled
  7. This deviceK051968 ABL8X7 FlexRadiometer Medical Aps · 2006
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K861087Avl Model 983 Electrolyte AnalyzerAvl Scientific Corp.1986—
−5K961458Avl 9180 Electrolyte AnalyzerAvl Scientific Corp.19961
−4K972673Avl 9181 Electrolyte AnalyzerAvl Scientific Corp.19977
−3K992859ABL700 Series Analyzer with Autocheck ModuleRadiometer America, Inc.20004
−2K041874ABL800 FlexRadiometer Medical Aps200422
−1K043218ABL800 Flex with Flexq Module, Model ABL8XXRadiometer Medical Aps20053
0 (this device)K051968ABL8X7 FlexRadiometer Medical Aps2006—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198710
199210
199510
199630
199720
199820
200010
FIG. 1 — Lineage of K051968, 1986–2006Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CGL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CGL device

  1. #1 K261039 i-STAT Crea cartridge with the i-STAT 1 Systemscore 11.9 · 2 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041874ABL800 Flex
Summary page 1
…Analyzer Classification Name: Creatinine Test System (21 CFR Section 862.1225) Predicate Devices ABL800 FLEX (K041874), ABL800 FLEX with FLEXQ Module (K043218), ABL700 Series Upgrade (K002290) all by Radiometer Medical ApS and i-STAT System Creati…
Radiometer Medical Aps200422
*K043218ABL800 Flex with Flexq Module, Model ABL8XX
Summary page 1
…e: Creatinine Test System (21 CFR Section 862.1225) Predicate Devices ABL800 FLEX (K041874), ABL800 FLEX with FLEXQ Module (K043218), ABL700 Series Upgrade (K002290) all by Radiometer Medical ApS and i-STAT System Creatinine Test (K973292) by i-STAT. Device…
Radiometer Medical Aps20053
*K002290Abl-700 Series, Model Abl 7XX (Various Models)
Summary page 1
…tem (21 CFR Section 862.1225) Predicate Devices ABL800 FLEX (K041874), ABL800 FLEX with FLEXQ Module (K043218), ABL700 Series Upgrade (K002290) all by Radiometer Medical ApS and i-STAT System Creatinine Test (K973292) by i-STAT. Device Description The ABL837 FLEX is an…
Radiometer Medical A/S20009
*K973292I-Stat System
Summary page 1
…Devices ABL800 FLEX (K041874), ABL800 FLEX with FLEXQ Module (K043218), ABL700 Series Upgrade (K002290) all by Radiometer Medical ApS and i-STAT System Creatinine Test (K973292) by i-STAT. Device Description The ABL837 FLEX is an analyzer with the capability of measuring creatinine. The wet section of…
I-Stat Corp.19972

Also cleared against K041874: 5 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 2, 2018Class IIZ-0801-2019Under Investigation by firmFrom three incidents in Denmark and Sweden, comparison studies on patient blood samples carried out for quality assurance demonstrated that creatinine measurements as determined on analyzers could be underestimated with up to 28% as…
Oct 29, 2014Class IIZ-1293-2015Under Investigation by firmNegative drift on QC during in-use lifetime of CREA membranes can be observed.
Dec 16, 2011Not classifiedZ-1008-2012Device DesignWhen the user activates the dosicap (containing the crea additive) and mounts it onto the bottle, the threads will not seal correctly causing approximately 10% of the solution to leak out when the bottle is placed horizontally for…
Mar 26, 2009Not classifiedZ-1015-2010OtherAnalysis of customer data, (for ABL 8X7 customers) has shown that the reference membranes are out of analytical specifications when replaced as specified in the device labeling. These errors in analytical specifications could lead to…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 20 years and passes through 3 companies; a new generation was cleared every 2.4 years on median.
  • 11 of the 18 ancestors have a recall of any class.
  • 5 other devices also cite its predicate K041874.

How does its category compare?

All CGL devices

CGL Electrode, Ion Based, Enzymatic, Creatinine: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CGL

1in 2017

2007: 12012: 12017: 12026 (partial year): 1
2007high 1 (2007) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CGL

Too few CGL 510(k)s in any year since 2007 to give a median.

Recalls per year · CGL

1in 2011

2007: 32008: 12009: 12010: 12011: 1
2007high 3 (2007) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CGL

8in 2025

2007: 172008: 32009: 62010: 122011: 42012: 82013: 62014: 12015: 42016: 22017: 142018: 112019: 252020: 192021: 202022: 112023: 52024: 72025: 8
2007high 25 (2019) · low 1 (2014)2025

MAUDE reports naming product code CGL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—317
20080—13
20090—16
20100—112
20110—14
20121—08
20130—06
20140—01
20150—04
20160—02
20171—014
20180—011
20190—025
20200—019
20210—020
20220—011
20230—05
20240—07
20250—08
20261—0—

What is it indicated for?

The ABL837 FLEX analyzer is intended for in vitro testing of samples of heparinized whole blood, plasma and serum for the parameter Creatinine. Creatinine measurements measure the concentration of creatinine in blood. Creatinine measurements are used in the diagnosis and treatment of renal diseases and in monitoring renal dialysis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K051968: ABL8X7 Flex.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K051968

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026