510(k) · Cleared
Recalled · Class II · 2016Medix Natal Care Intensive Care Incubator, Models-Natal Care ST-MX/LX/BX
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Medix I.C.S.A. · FMZ Incubator, Neonatal
Cleared on Sep 21, 2007 as substantially equivalent to K920049 (Medix I.C.S.A., 1993) and 1 other predicate after a 234-day review. Its family tree reaches back 1 generation to 1993. It has been recalled 2 times.
- Decided
- Sep 21, 2007
- Review
- 234 days
- Class
- Class II
- Product code
- FMZ
- Recalls
- 2
- Predicates
- 2
Its longest line of descent runs back 1 generation to K920049 (1993). 2 ancestors in all. No later device cites it as a predicate yet.
- Root · 1 generation backK920049 Medix PC-305 Neonatal IncubatorMedix I.C.S.A. · 1993
- This deviceK070291 Medix Natal Care Intensive Care Incubator, Models-Natal Care…Medix I.C.S.A. · 2007
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K920049 | Medix PC-305 Neonatal Incubator | Medix I.C.S.A. | 1993 | — |
| 0 (this device) | K070291 | Medix Natal Care Intensive Care Incubator, Models-Natal Care ST-MX/LX/BX | Medix I.C.S.A. | 2007 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2000 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FMZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FMZ device
- #1 K251663 Giraffe OmniBed Carestation (CS1); Giraffe…score 14.4 · 5 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K920049 | Medix PC-305 Neonatal IncubatorSummary page 1…redicate Devices: The NATAL CARE ST-MXILXIBX incubator issubstantially equivalent to other neonatal incubators on the market such as: Medix PC-305, K920049, Medix i.c.s.a. and Isolette Infant Incubator, Model C2HS, Hill-Rom Air Shields, K001242. 3. Intended Use State… | 1993 |
| * | K001242 | Isolette Infant Incubator, Model C2HSSummary page 1…r neonatal incubators on the market such as: Medix PC-305, K920049, Medix i.c.s.a. and Isolette Infant Incubator, Model C2HS, Hill-Rom Air Shields, K001242. 3. Intended Use Statement: Indications for Use: The Medix Natal Care ST-MXILXJBX Intensive Care… | 2000 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 25, 2016 | Class II | Z-0056-2017 | Unknown/Undetermined by firm | NatalCare LX incubators may have a defective base where the weld holding the wheel bushing to the tube is inadequate. There were several complaints of wheels bending out of alignment. |
| Jan 8, 2016 | Class II | Z-1363-2016 | Process control | NatalCare LX Incubators were manufactured using a defective oval port hinge assembly with the result that the screws holding the port in place can become loose and possibly result in the port detaching from the incubator. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- None of the 2 ancestors has a recall.
How does its category compare?
All FMZ devicesFMZ Incubator, Neonatal: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.
FDA decisions per year · FMZ
2in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FMZ
60in 2010
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FMZ
2in 2024
Class I (11 since 2007) Class II/III not classified
Adverse event reports per year · FMZ
250in 2025
MAUDE reports naming product code FMZ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 81 | 0 | 14 |
| 2008 | 1 | — | 4 | 10 |
| 2009 | 0 | — | 0 | 8 |
| 2010 | 4 | 60 | 0 | 11 |
| 2011 | 1 | — | 5 | 10 |
| 2012 | 0 | — | 0 | 12 |
| 2013 | 0 | — | 1 (1) | 11 |
| 2014 | 1 | — | 0 | 3 |
| 2015 | 1 | — | 0 | 9 |
| 2016 | 1 | — | 1 | 12 |
| 2017 | 1 | — | 0 | 4 |
| 2018 | 2 | — | 3 | 18 |
| 2019 | 2 | — | 1 | 24 |
| 2020 | 0 | — | 0 | 8 |
| 2021 | 0 | — | 0 | 15 |
| 2022 | 1 | — | 0 | 27 |
| 2023 | 0 | — | 8 (8) | 19 |
| 2024 | 0 | — | 2 (2) | 118 |
| 2025 | 2 | — | 0 | 250 |
| 2026 | 2 | — | 3 | — |
What is it indicated for?
The Medix Natal Care ST-MX/LX/BX Intensive Care Incubator provides
a controlled thermal environment for neonates who are unable to provide their own
thermoregulation. The Medix Natal Care ST-MX/LX/BX Intensive Care Incubator can be
used in two operating modes: Air Control Mode and Skin Control Mode with two probes that
can be used with twins or for a follow up of central/peripheral temperature variations in a
patient as a tool for the early detection of thermal stress-effect on the newborn.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K070291 · printed Nov 6, 2026
Cite this page
510(k) Report. “K070291: Medix Natal Care Intensive Care Incubator, Models-Natal Care ST-MX/LX/BX.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K070291
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026