510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Zimmer Segmental System

K070978510(k) TraditionalCleared Jul 3, 200788-day reviewClass II21 CFR 888.3510Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zimmer, Inc. · KRO Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer

Cleared on Jul 3, 2007 as substantially equivalent to K013385 (Zimmer, Inc., 2002) and 1 other predicate after an 88-day review. Its family tree reaches back 2 generations to 1996. 13 later devices descend from it. It has been recalled 6 times.

Decided
Jul 3, 2007
Review
88 days
Class
Class II
Product code
KRO
Recalls
6
Predicates
2

Its longest line of descent runs back 2 generations to K945028 (1996). 3 ancestors in all. 13 later devices descend from it, 5 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor2019: 2 descendants cleared2015: 3 descendants cleared2011: 1 descendant cleared2020: 1 descendant cleared2008: 1 descendant cleared2014: 1 descendant cleared2025: 3 descendants cleared2026: 1 descendant clearedK945028 · Finn Knee System · Biomet, Inc. · 1996K013385 · Nexgen Complete Knee Solution Rotatng Hinge Knee · Zimmer, Inc. · 2002 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 5 descendants2 generations forward: 5 descendants3 generations forward: 3 descendants K013385 · Nexgen Complete Knee Solution Rotatng Hinge Knee · Zimmer, Inc. · 2002 · recalledK945028 · Finn Knee System · Biomet, Inc. · 19960
  1. Root · 2 generations backK945028 Finn Knee SystemBiomet, Inc. · 1996
  2. 1 generation backK013385 Nexgen Complete Knee Solution Rotatng Hinge KneeZimmer, Inc. · 2002 · Recalled
  3. This deviceK070978 Zimmer Segmental SystemZimmer, Inc. · 2007
  4. Descendant · cites it directly · 10 of its ownK110940 Zimmer Segment System Trabecular Metal Proximal Tibial Component…Zimmer, Inc. · 2011
  5. Descendant · 2 generations on · 3 of its ownK141331 Biomet Orthopaedic Salvage System (Oss)Biomet, Inc. · 2014
  6. Descendant · cites it directly · 2 of its ownK190100 USTAR II SystemUnited Orthopedic Corporation · 2019 · Recalled
  7. Descendant · 2 generations on · 1 of its ownK183136 Zimmer Segmental System Proximal Femoral ComponentZimmer, Inc. · 2019
  8. Descendant · 3 generations on · 1 of its ownK142814 Biomet Orthopaedic Salvage System (OSSTM) - Proximal Femoral &…Biomet Manufacturing, LLC · 2015 · Recalled

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2008: 12011: 12014: 12015: 32019: 22020: 12025: 32026: 1
2008high 3 (2015) · low 0 (2009)2026

Most influential descendants

Showing 5 of 13 · See all
  • K110940Zimmer Segment System Trabecular Metal Proximal Tibial Component, Trabecular Metal…Zimmer, Inc. · 2011 · cites it directly10descendants
  • K141331Biomet Orthopaedic Salvage System (Oss)Biomet, Inc. · 2014 · 2 generations on3descendants
  • K190100USTAR II SystemUnited Orthopedic Corporation · 2019 · cites it directly2descendants
  • K183136Zimmer Segmental System Proximal Femoral ComponentZimmer, Inc. · 2019 · 2 generations on1descendant
  • K142814Biomet Orthopaedic Salvage System (OSSTM) - Proximal Femoral & Hybrid TibialBiomet Manufacturing, LLC · 2015 · 3 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K945028Finn Knee SystemBiomet, Inc.1996—
−1K013385Nexgen Complete Knee Solution Rotatng Hinge KneeZimmer, Inc.20022
0 (this device)K070978Zimmer Segmental SystemZimmer, Inc.2007—
+1K110940Zimmer Segment System Trabecular Metal Proximal Tibial Component, Trabecular…Zimmer, Inc.2011—
+2K141331Biomet Orthopaedic Salvage System (Oss)Biomet, Inc.2014—
+1K190100USTAR II SystemUnited Orthopedic Corporation20195
+2K183136Zimmer Segmental System Proximal Femoral ComponentZimmer, Inc.2019—
+3K142814Biomet Orthopaedic Salvage System (OSSTM) - Proximal Femoral & Hybrid TibialBiomet Manufacturing, LLC20156

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
201110
201410
201530
201920
202010
202530
202610
FIG. 1 — Lineage of K070978, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KRO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KRO device

  1. #1 K251665 Triathlon® Hinge Knee Systemscore 19.0 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K013385Nexgen Complete Knee Solution Rotatng Hinge Knee
Summary page 1
…NexGen® Complete Knee Solutions Rotating Hinge Knee Systems, manufactured by Zimmer, Inc., K013385, cleared January 9, 2002. Modular Options for Severe Bone Loss and Trauma…
Zimmer, Inc.20022
*K002324Modified to Modular FR Severe Bone Loss Trauma
Summary page 1
…Modular Options for Severe Bone Loss and Trauma (MOST®) System, manufactured by Zimmer Inc., K002324, cleared August 24, 2000. Device Description: The Zimmer®Segmental System is a fully…
Sulzermedica20001

Also cleared against K013385: 8 other devices →

What descends from it? (13)

See all 13

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K110940Zimmer Segment System Trabecular Metal Proximal Tibial Component, Trabecular Metal…Zimmer, Inc.201110
K190100USTAR II SystemUnited Orthopedic Corporation20192
K150028Zimmer Segmental System XT ComponentsZimmer, Inc.20151
K081860Zimmer Segmental System Variable Stiffness Stem Extensions and Intercalary SegmentsZimmer, Inc.20081
K192798Zimmer Segmental SystemZimmer, Inc.20200

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 1, 2017Class IIZ-0721-2018Package design/selectionThe products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.
Feb 28, 2017Class IIZ-2566-2017PackagingZimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
Feb 28, 2017Class IIZ-2574-2017PackagingZimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
Feb 28, 2017Class IIZ-2575-2017PackagingZimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
Dec 19, 2013Class IIZ-0783-2014Device DesignAfter evaluation of 13 complaints(0.4% rate) for hyper-extension of the Zimmer Segmental System, Zimmer is updating the related surgical technique and instructions for use to provide additional instructions and patient conditions that may…
Dec 19, 2013Class IIZ-0784-2014Device DesignAfter evaluation of 13 complaints(0.4% rate) for hyper-extension of the Zimmer Segmental System, Zimmer is updating the related surgical technique and instructions for use to provide additional instructions and patient conditions that may…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 2 companies; a new generation was cleared every 5.9 years on median.
  • 2 of the 3 ancestors have a recall of any class.
  • Of its descendants, K110940 (Zimmer, Inc., 2011) has gone furthest, with 10 descendants of its own.
  • 8 other devices also cite its predicate K013385.

How does its category compare?

All KRO devices

KRO Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · KRO

3in 2025

2007: 12008: 32009: 32010: 12011: 22012: 22013: 32014: 22015: 62016: 32017: 12019: 32020: 12021: 32022: 12023: 42024: 32025: 32026 (partial year): 1
2007high 6 (2015) · low 0 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KRO

56in 2025

2008: 126 days median2009: 86 days median2013: 189 days median2015: 113 days median2016: 148 days median2019: 244 days median2021: 57 days median2023: 148 days median2024: 87 days median2025: 56 days median
2007high 244 (2019) · low 56 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KRO

14so far in 2026, the most since 2007

2007: 12008: 32009: 82011: 12012: 12013: 22015: 12016: 52017: 42018: 12021: 12025: 12026 (partial year): 14
2007high 14 (2026) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KRO

915in 2025

2007: 372008: 442009: 662010: 1102011: 1192012: 1732013: 1942014: 4062015: 3332016: 3692017: 5432018: 6142019: 7592020: 7492021: 7702022: 8922023: 7462024: 8552025: 915
2007high 915 (2025) · low 37 (2007)2025

MAUDE reports naming product code KRO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—137
20083126344
2009386866
20101—0110
20112—1119
20122—1173
201331892194
20142—0406
201561131333
201631485369
20171—4543
20180—1614
201932440759
20201—0749
20213571770
20221—0892
202341480746
20243870855
20253561915
20261—14—

What is it indicated for?

This device is indicated for: - Moderate to severe knee instability - Significant bone loss and/or ligament deficiencies caused by neoplasms, trauma, rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, and/or avascular necrosis of the femoral condyle - Valgus, varus or flexion deformities - The salvage of previously failed surgical attempts A total femoral replacement construct consisting of MOST Options proximal…

Read the full statement (153 words)

This device is indicated for:

- Moderate to severe knee instability

- Significant bone loss and/or ligament deficiencies caused by neoplasms, trauma, rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, and/or avascular necrosis of the femoral condyle

- Valgus, varus or flexion deformities

- The salvage of previously failed surgical attempts

A total femoral replacement construct consisting of MOST Options proximal femoral, Segmental System segments and Segmental System distal femoral components may be used without cement.

* Fluted stem extensions require the use of either a smooth or TrabecularMetal stem collar, which must be cemented to the stem. Following cementing to the stem extension, the smooth collar and the remainder of the stem must also be cemented against the bone.

* The TrabecularMetal collar may be used cemented or uncemented against the bone as long as the remainder of the stem extension is cemented.

* All other constructs are for cemented use only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K070978: Zimmer Segmental System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K070978

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026