510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

C350 Powered Wheelchair

K071650510(k) TraditionalCleared Sep 12, 200786-day reviewClass II21 CFR 890.3860Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Permobil AB · ITI Wheelchair, Powered

Cleared on Sep 12, 2007 as substantially equivalent to K041219 (Permobil AB, 2004) after an 86-day review. Its family tree reaches back 3 generations to 1997. 2 of the 26 devices in its family, counting itself, have a Class I recall. 21 later devices descend from it. It has been recalled 3 times.

Decided
Sep 12, 2007
Review
86 days
Class
Class II
Product code
ITI
Recalls
3
Predicates
1

Its longest line of descent runs back 3 generations to K960951 (1997). 4 ancestors in all. 21 later devices descend from it, 4 citing it directly. 2 of the 26 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1998: 1 other ancestor2008: 2 descendants cleared2011: 1 descendant cleared2013: 1 descendant cleared2018: 1 descendant cleared2019: 2 descendants cleared2015: 3 descendants cleared2021: 2 descendants cleared, 1 with a Class I recall12016: 1 descendant cleared2022: 2 descendants cleared, 1 with a Class I recall12024: 2 descendants cleared2023: 2 descendants cleared2026: 2 descendants clearedK960951 · Permobil Chairman L/CS 8 · Permobil AB · 1997K991658 · Permobil Powered Wheelchair 1280 · Permobil AB · 1999 · recalledK041219 · Electro · Permobil AB · 2004 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 3 generations back: 1 other ancestor 1 generation forward: 4 descendants2 generations forward: 6 descendants3 generations forward: 4 descendants4 generations forward: 1 descendant, 1 with a Class I recall15 generations forward: 3 descendants, 1 with a Class I recall16 generations forward: 1 descendant7 generations forward: 2 descendants K041219 · Electro · Permobil AB · 2004 · recalledK991658 · Permobil Powered Wheelchair 1280 · Permobil AB · 1999 · recalledK960951 · Permobil Chairman L/CS 8 · Permobil AB · 19970+5
  1. Root · 3 generations backK960951 Permobil Chairman L/CS 8Permobil AB · 1997
  2. 2 generations backK991658 Permobil Powered Wheelchair 1280Permobil AB · 1999 · Recalled
  3. 1 generation backK041219 ElectroPermobil AB · 2004 · Recalled
  4. This deviceK071650 C350 Powered WheelchairPermobil AB · 2007
  5. Descendant · cites it directly · 12 of its ownK103477 Permobil M300 / M400Permobil AB · 2011
  6. Descendant · 2 generations on · 9 of its ownK143180 F3Permobil AB · 2015
  7. Descendant · 3 generations on · 7 of its ownK153543 WHILL Model MWhill, Inc. · 2016
  8. Descendant · 4 generations on · 6 of its ownK213383 WHILL Model C2Whill, Inc. · 2021 · Recalled (Class I)
  9. Descendant · cites it directly · 4 of its ownK083017 LEVO C3Levo AG · 2008

Showing 5 of 21 descendants. See all descendants

Descendants cleared per year

2008: 22011: 12013: 12015: 32016: 12018: 12019: 22021: 22022: 22023: 22024: 22026: 2
2008high 3 (2015) · low 0 (2009)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 21 · See all
  • K103477Permobil M300 / M400Permobil AB · 2011 · cites it directly12descendants
  • K143180F3Permobil AB · 2015 · 2 generations on9descendants
  • K153543WHILL Model MWhill, Inc. · 2016 · 3 generations on7descendants
  • K213383WHILL Model C2Whill, Inc. · 2021 · 4 generations on6descendants
  • K083017LEVO C3Levo AG · 2008 · cites it directly4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K960951Permobil Chairman L/CS 8Permobil AB1997—
−2K991658Permobil Powered Wheelchair 1280Permobil AB19995
−1K041219ElectroPermobil AB20044
0 (this device)K071650C350 Powered WheelchairPermobil AB2007—
+1K103477Permobil M300 / M400Permobil AB2011—
+2K143180F3Permobil AB2015—
+3K153543WHILL Model MWhill, Inc.2016—
+4K213383WHILL Model C2Whill, Inc.2021Class I
+1K083017LEVO C3Levo AG2008—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200820
201110
201310
201530
201610
201810
201920
202121
202221
202320
202420
202620
FIG. 1 — Lineage of K071650, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new ITI device

  1. #1 K220747 Power Wheelchair (N5515B)score 40.2 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041219Electro
Summary page 1
…ed: June, 2007 Device name: C350 Classification Name: Powered wheelchair Predicate Devices: C300(Electro)( 1(041219) manufactured by Permobil AR Intended use: The intended use of the C350 powered wheelchair is to provide…
Permobil AB20044

What descends from it? (21)

See all 21

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K103477Permobil M300 / M400Permobil AB201112
K083017LEVO C3Levo AG20084
K123290M300 & M400Permobil AB20132
K081964K450 Powered WheelchairPermobil AB20080

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 15, 2018Class IIZ-3178-2018Device DesignThis correction decision has been made because of a potential failure of one or both backrest bars providing back support. If one or both back support telescoping bars fail the user could fall back leading to a potential injury.
Sep 1, 2017Class IIZ-0005-2018Device DesignThere is a potential failure of the top plate assembly, which is the component that connects the seating system of the wheelchair to the base.
Sep 1, 2017Class IIZ-0006-2018Device DesignThere is a potential failure of the top plate assembly, which is the component that connects the seating system of the wheelchair to the base.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K213383WHILL Model C2Whill, Inc. · 2021 · 4 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall
  • K221438WHILL Model FWhill, Inc. · 2022 · 5 generations on · ◆ in the 2022 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 10 years and passes through 1 company; a new generation was cleared every 3.1 years on median.
  • 2 of the 4 ancestors have a recall of any class.
  • Of its descendants, K103477 (Permobil AB, 2011) has gone furthest, with 12 descendants of its own.

How does its category compare?

All ITI devices

ITI Wheelchair, Powered: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · ITI

29in 2025

2007: 182008: 92009: 72010: 62011: 42012: 62013: 52014: 52015: 102016: 62017: 42018: 42019: 32020: 62021: 132022: 162023: 82024: 382025: 292026 (partial year): 36
2007high 38 (2024) · low 3 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · ITI

90in 2025

2007: 42 days median2008: 49 days median2009: 63 days median2010: 46 days median2011: 101 days median2012: 155 days median2013: 177 days median2014: 213 days median2015: 174 days median2016: 173 days median2017: 159 days median2018: 117 days median2019: 244 days median2020: 278 days median2021: 209 days median2022: 96 days median2023: 128 days median2024: 125 days median2025: 90 days median2026 (partial year): 108 days median
2007high 278 (2020) · low 42 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · ITI

13so far in 2026, the most since 2007

2007: 32008: 32009: 12010: 22012: 12016: 12017: 22021: 32024: 12025: 3, 2 Class I2026 (partial year): 13
2007high 13 (2026) · low 0 (2011)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · ITI

647in 2025

2007: 1232008: 1332009: 1832010: 1572011: 5252012: 2,9192013: 8,7412014: 7,2822015: 4,6852016: 1,5752017: 3012018: 2732019: 2642020: 2052021: 1662022: 1352023: 2232024: 1462025: 647
2007high 8,741 (2013) · low 123 (2007)2025

MAUDE reports naming product code ITI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200718423123
20089493133
20097631183
20106462157
201141010525
2012615512,919
2013517708,741
2014521307,282
20151017404,685
2016617311,575
201741592301
201841170273
201932440264
202062780205
2021132093166
202216960135
202381280223
2024381251146
202529903 (2)647
20263610813—

What is it indicated for?

The intended use of the C350 series of the powered wheelchair is to provide

indoor and outdoor mobility to persons limited to a seating position that are

capable of operating a powered wheelchair.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K071650: C350 Powered Wheelchair.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K071650

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026