510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Permobil Powered Wheelchair 1280

K991658510(k) AbbreviatedCleared Oct 8, 1999147-day reviewClass II21 CFR 890.3860Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Permobil AB · ITI Wheelchair, Powered

Cleared on Oct 8, 1999 as substantially equivalent to K960951 (Permobil AB, 1997) and 1 other predicate after a 147-day review. Its family tree reaches back 1 generation to 1997. 2 of the 35 devices in its family, counting itself, have a Class I recall. 32 later devices descend from it. It has been recalled 5 times.

Decided
Oct 8, 1999
Review
147 days
Class
Class II
Product code
ITI
Recalls
5
Predicates
2

Its longest line of descent runs back 1 generation to K960951 (1997). 2 ancestors in all. 32 later devices descend from it, 2 citing it directly. 2 of the 35 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1998: 1 other ancestor2002: 1 descendant cleared2004: 2 descendants cleared2015: 4 descendants cleared2012: 1 descendant cleared2007: 1 descendant cleared2008: 2 descendants cleared2011: 1 descendant cleared2013: 1 descendant cleared2019: 3 descendants cleared2017: 1 descendant cleared2018: 2 descendants cleared2021: 3 descendants cleared, 1 with a Class I recall12020: 1 descendant cleared2016: 1 descendant cleared2022: 2 descendants cleared, 1 with a Class I recall12024: 2 descendants cleared2023: 2 descendants cleared2026: 2 descendants clearedK960951 · Permobil Chairman L/CS 8 · Permobil AB · 19971995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 4 descendants3 generations forward: 6 descendants4 generations forward: 8 descendants5 generations forward: 5 descendants6 generations forward: 1 descendant, 1 with a Class I recall17 generations forward: 3 descendants, 1 with a Class I recall18 generations forward: 1 descendant9 generations forward: 2 descendants K960951 · Permobil Chairman L/CS 8 · Permobil AB · 19970+5
  1. Root · 1 generation backK960951 Permobil Chairman L/CS 8Permobil AB · 1997
  2. This deviceK991658 Permobil Powered Wheelchair 1280Permobil AB · 1999
  3. Descendant · cites it directly · 22 of its ownK041219 ElectroPermobil AB · 2004 · Recalled
  4. Descendant · 2 generations on · 21 of its ownK071650 C350 Powered WheelchairPermobil AB · 2007 · Recalled
  5. Descendant · 3 generations on · 12 of its ownK103477 Permobil M300 / M400Permobil AB · 2011
  6. Descendant · 4 generations on · 9 of its ownK143180 F3Permobil AB · 2015
  7. Descendant · cites it directly · 8 of its ownK021264 TRZ-CG Power Positioning System with Center-of-Gravity Shifting…Motion Concepts · 2002

Showing 5 of 32 descendants. See all descendants

Descendants cleared per year

2002: 12004: 22007: 12008: 22011: 12012: 12013: 12015: 42016: 12017: 12018: 22019: 32020: 12021: 32022: 22023: 22024: 22026: 2
2002high 4 (2015) · low 0 (2003)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 32 · See all
  • K041219ElectroPermobil AB · 2004 · cites it directly22descendants
  • K071650C350 Powered WheelchairPermobil AB · 2007 · 2 generations on21descendants
  • K103477Permobil M300 / M400Permobil AB · 2011 · 3 generations on12descendants
  • K143180F3Permobil AB · 2015 · 4 generations on9descendants
  • K021264TRZ-CG Power Positioning System with Center-of-Gravity Shifting Power Tilt, Recline, and…Motion Concepts · 2002 · cites it directly8descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K960951Permobil Chairman L/CS 8Permobil AB1997—
0 (this device)K991658Permobil Powered Wheelchair 1280Permobil AB1999—
+1K041219ElectroPermobil AB20044
+2K071650C350 Powered WheelchairPermobil AB20073
+3K103477Permobil M300 / M400Permobil AB2011—
+4K143180F3Permobil AB2015—
+1K021264TRZ-CG Power Positioning System with Center-of-Gravity Shifting Power Tilt…Motion Concepts2002—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200210
200420
200710
200820
201110
201210
201310
201540
201610
201710
201820
201930
202010
202131
202221
202320
202420
202620
FIG. 1 — Lineage of K991658, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new ITI device

  1. #1 K220747 Power Wheelchair (N5515B)score 40.2 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K960951Permobil Chairman L/CS 8
Summary page 1
…@permobil.se May 10 1999 Powered wheelchair 1280 Classification name Powered wheelchair Identification of predicate devices Intended use Permobil Chairman L/CS 8 (K960951) G-424 power wheelchair (K983677) The intended use of Permobil powered wheelchair 1280 is to provide indoor and outdoor mobility…
Permobil AB1997—
*K983677Quickie G Series, Model G-424
Summary page 1
…d wheelchair 1280 Classification name Powered wheelchair Identification of predicate devices Intended use Permobil Chairman L/CS 8 (K960951) G-424 power wheelchair (K983677) The intended use of Permobil powered wheelchair 1280 is to provide indoor and outdoor mobility to persons restricted to a sitti…
Sunrise Medical, Inc.1998—

Also cleared against K960951: 7 other devices →

What descends from it? (32)

See all 32

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041219ElectroPermobil AB200422
K021264TRZ-CG Power Positioning System with Center-of-Gravity Shifting Power Tilt, Recline, and…Motion Concepts20028

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 15, 2018Class IIZ-3178-2018Device DesignThis correction decision has been made because of a potential failure of one or both backrest bars providing back support. If one or both back support telescoping bars fail the user could fall back leading to a potential injury.
Sep 1, 2017Class IIZ-0005-2018Device DesignThere is a potential failure of the top plate assembly, which is the component that connects the seating system of the wheelchair to the base.
Sep 1, 2017Class IIZ-0006-2018Device DesignThere is a potential failure of the top plate assembly, which is the component that connects the seating system of the wheelchair to the base.
Mar 7, 2003Not classifiedZ-0834-03OtherOne of the wheelchair motors needed a grounding strap to ensure that the motors are at the same voltage potential as the chassis of the wheelchair.
Jan 1, 2003Not classifiedZ-0671-03OtherIncidents were reported where a drive wheel has come off its motor axle, due either to loosening of the wheel bolt or a broken wheel bolt.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K213383WHILL Model C2Whill, Inc. · 2021 · 6 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall
  • K221438WHILL Model FWhill, Inc. · 2022 · 7 generations on · ◆ in the 2022 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • None of the 2 ancestors has a recall.
  • Of its descendants, K041219 (Permobil AB, 2004) has gone furthest, with 22 descendants of its own.
  • 7 other devices also cite its predicate K960951.

How does its category compare?

All ITI devices

ITI Wheelchair, Powered: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · ITI

29in 2025

2007: 182008: 92009: 72010: 62011: 42012: 62013: 52014: 52015: 102016: 62017: 42018: 42019: 32020: 62021: 132022: 162023: 82024: 382025: 292026 (partial year): 36
2007high 38 (2024) · low 3 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · ITI

90in 2025

2007: 42 days median2008: 49 days median2009: 63 days median2010: 46 days median2011: 101 days median2012: 155 days median2013: 177 days median2014: 213 days median2015: 174 days median2016: 173 days median2017: 159 days median2018: 117 days median2019: 244 days median2020: 278 days median2021: 209 days median2022: 96 days median2023: 128 days median2024: 125 days median2025: 90 days median2026 (partial year): 108 days median
2007high 278 (2020) · low 42 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · ITI

13so far in 2026, the most since 2007

2007: 32008: 32009: 12010: 22012: 12016: 12017: 22021: 32024: 12025: 3, 2 Class I2026 (partial year): 13
2007high 13 (2026) · low 0 (2011)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · ITI

647in 2025

2007: 1232008: 1332009: 1832010: 1572011: 5252012: 2,9192013: 8,7412014: 7,2822015: 4,6852016: 1,5752017: 3012018: 2732019: 2642020: 2052021: 1662022: 1352023: 2232024: 1462025: 647
2007high 8,741 (2013) · low 123 (2007)2025

MAUDE reports naming product code ITI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200718423123
20089493133
20097631183
20106462157
201141010525
2012615512,919
2013517708,741
2014521307,282
20151017404,685
2016617311,575
201741592301
201841170273
201932440264
202062780205
2021132093166
202216960135
202381280223
2024381251146
202529903 (2)647
20263610813—

What is it indicated for?

The intended use is to provide indoor and outdoor moblhty to persons restricted to a sitting position.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K991658: Permobil Powered Wheelchair 1280.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K991658

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026