510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Exactech Novation Crown Cup Constrained Liners and Rings

K071676510(k) SpecialCleared Jul 19, 200730-day reviewClass II21 CFR 888.3310Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Exactech, Inc. · KWZ Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer

Cleared on Jul 19, 2007 as substantially equivalent to K040601 (Exactech, Inc., 2004) after a 30-day review. Its family tree reaches back 4 generations to 1997. It has been recalled 2 times.

Decided
Jul 19, 2007
Review
30 days
Class
Class II
Product code
KWZ
Recalls
2
Predicates
1

Its longest line of descent runs back 4 generations to P960047 (1997). 11 ancestors in all, the oldest K921114 (1993). No later device cites it as a predicate yet.

  1. Root · 4 generations backP960047 Osteonics Constrained Acetabular InsertHowmedica Osteonics Corp. · 1997
  2. 3 generations backK021661 Ringloc Constrained LinersBiomet Orthopedics, Inc. · 2002
  3. 2 generations backK030923 Epsilon Durasul Constrained Acetabular LinerCenterpulse Orthopedics, Inc. · 2003
  4. 1 generation backK040601 Exactech Acumatch a-Series & MCS Constrained Acetabular LinersExactech, Inc. · 2004 · Recalled
  5. This deviceK071676 Exactech Novation Crown Cup Constrained Liners and RingsExactech, Inc. · 2007
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4P960047Osteonics Constrained Acetabular InsertHowmedica Osteonics Corp.1997—
−3K021661Ringloc Constrained LinersBiomet Orthopedics, Inc.2002—
−2K030923Epsilon Durasul Constrained Acetabular LinerCenterpulse Orthopedics, Inc.2003—
−1K040601Exactech Acumatch a-Series & MCS Constrained Acetabular LinersExactech, Inc.20041
0 (this device)K071676Exactech Novation Crown Cup Constrained Liners and RingsExactech, Inc.2007—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199420
199710
199920
200210
FIG. 1 — Lineage of K071676, 1997–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KWZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KWZ device

  1. #1 K241461 Mpact Constrained Linerscore 15.5 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K040601Exactech Acumatch a-Series & MCS Constrained Acetabular Liners
Summary page 1
…Information on devices to which substantial equivalence is claimed: 510(k) Trade or Proprietary or Model Name Manufacturer Number #K040601 AcuMatch A-Series & MCS Constrained Liners Exactech, Inc. Indications for Use: Novation Crown Cup TM…
Exactech, Inc.20041

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 21, 2023Class IIZ-0581-2024Packaging change controlExactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.
Aug 11, 2022Class IIZ-1730-2022Process controlSpecific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 4 companies; a new generation was cleared every 2.8 years on median.
  • 3 of the 11 ancestors have a recall of any class.

How does its category compare?

All KWZ devices

KWZ Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · KWZ

1in 2025

2007: 72008: 12009: 12010: 12011: 12012: 12016: 12017: 22024: 12025: 1
2007high 7 (2007) · low 0 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KWZ

104in 2007

2007: 104 days median
2007high 104 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KWZ

1in 2023

2010: 82012: 12023: 12026 (partial year): 1
2007high 8 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KWZ

193in 2025

2007: 412008: 442009: 812010: 982011: 1272012: 1682013: 2742014: 2782015: 6362016: 1,1802017: 4772018: 2832019: 5322020: 2882021: 2162022: 1582023: 1642024: 1452025: 193
2007high 1,180 (2016) · low 41 (2007)2025

MAUDE reports naming product code KWZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077104041
20081—044
20091—081
20101—898
20111—0127
20121—1168
20130—0274
20140—0278
20150—0636
20161—01,180
20172—0477
20180—0283
20190—0532
20200—0288
20210—0216
20220—0158
20230—1164
20241—0145
20251—0193
20260—1—

What is it indicated for?

Exactech Novation Crown Cup TM Constrained Liners and Rings are components of the Exactech Novation ® non-cemented acetabular cup system. The device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered. They are intended for press-fit fixation and are compatible with Exactech press-fit or cemented femoral stem components.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K071676: Exactech Novation Crown Cup Constrained Liners and Rings.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K071676

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026