Exactech Acumatch a-Series & MCS Constrained Acetabular Liners

K040601510(k) AbbreviatedCleared Oct 4, 2004210-day reviewClass II21 CFR 888.3310Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Exactech, Inc. · KWZ Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer

Cleared on Oct 4, 2004 as substantially equivalent to K993082 (Exactech, Inc., 1999) and 3 other predicates after a 210-day review. Its family tree reaches back 3 generations to 1997. 1 later device descends from it. It has been recalled once.

Decided
Oct 4, 2004
Review
210 days
Class
Class II
Product code
KWZ
Recalls
1
Predicates
4

Its longest line of descent runs back 3 generations to P960047 (1997). 10 ancestors in all, the oldest K921114 (1993). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor1999: 2 other ancestors1993: 1 other ancestor1997: 1 other ancestor1994: 2 other ancestors2007: 1 descendant clearedP960047 · Osteonics Constrained Acetabular Insert · Howmedica Osteonics Corp. · 1997K021661 · Ringloc Constrained Liners · Biomet Orthopedics, Inc. · 2002K030923 · Epsilon Durasul Constrained Acetabular Liner · Centerpulse Orthopedics, Inc. · 20031990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors2 generations back: 4 other ancestors 1 generation forward: 1 descendant K030923 · Epsilon Durasul Constrained Acetabular Liner · Centerpulse Orthopedics, Inc. · 2003K021661 · Ringloc Constrained Liners · Biomet Orthopedics, Inc. · 2002P960047 · Osteonics Constrained Acetabular Insert · Howmedica Osteonics Corp. · 19970
  1. Root · 3 generations backP960047 Osteonics Constrained Acetabular InsertHowmedica Osteonics Corp. · 1997
  2. 2 generations backK021661 Ringloc Constrained LinersBiomet Orthopedics, Inc. · 2002
  3. 1 generation backK030923 Epsilon Durasul Constrained Acetabular LinerCenterpulse Orthopedics, Inc. · 2003
  4. This deviceK040601 Exactech Acumatch a-Series & MCS Constrained Acetabular LinersExactech, Inc. · 2004
  5. Descendant · cites it directly · 0 of its ownK071676 Exactech Novation Crown Cup Constrained Liners and RingsExactech, Inc. · 2007 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K071676Exactech Novation Crown Cup Constrained Liners and RingsExactech, Inc. · 2007 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3P960047Osteonics Constrained Acetabular InsertHowmedica Osteonics Corp.1997—
−2K021661Ringloc Constrained LinersBiomet Orthopedics, Inc.2002—
−1K030923Epsilon Durasul Constrained Acetabular LinerCenterpulse Orthopedics, Inc.2003—
0 (this device)K040601Exactech Acumatch a-Series & MCS Constrained Acetabular LinersExactech, Inc.2004—
+1K071676Exactech Novation Crown Cup Constrained Liners and RingsExactech, Inc.20072

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199420
199710
199920
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
FIG. 1 — Lineage of K040601, 1997–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KWZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KWZ device

  1. #1 K241461 Mpact Constrained Linerscore 15.5 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K993082Exactech Acumatch Integrated Hip System a-Series Porous Coated Acetabular Component
Summary page 3
…K040601 - Page 3 of 3 Exactech AcuMatch A-Series ~ & MCS ® Constrained Acetabular Liners The devices are identical to previously cleared Exactech AcuMatch® (K993082) and MCS®(K921114) Liners except for the femoral head constraining features. The devices are similar to the Ringloc Constrained L…
Exactech, Inc.19992
*K921114Exactech MCS Porous Coated Total Hip Acetabular
Summary page 3
…Page 3 of 3 Exactech AcuMatch A-Series ~ & MCS ® Constrained Acetabular Liners The devices are identical to previously cleared Exactech AcuMatch® (K993082) and MCS®(K921114) Liners except for the femoral head constraining features. The devices are similar to the Ringloc Constrained Liner (K021728) fro…
Exactech, Inc.1993—
*K021728Ringloc Constrained Liners II
Summary page 3
…leared Exactech AcuMatch® (K993082) and MCS®(K921114) Liners except for the femoral head constraining features. The devices are similar to the Ringloc Constrained Liner (K021728) from Biomet Orthopedics and the Epsilon Durasul Constrained Acetabular Liner (K030923) from Centerpulse Orthopedics. Performan…
Biomet, Inc.2002—
*K030923Epsilon Durasul Constrained Acetabular Liner
Summary page 3
…d constraining features. The devices are similar to the Ringloc Constrained Liner (K021728) from Biomet Orthopedics and the Epsilon Durasul Constrained Acetabular Liner (K030923) from Centerpulse Orthopedics. Performance Testing: The Exactech AcuMatch A-Series® Constrained Acetabular Liner & MCS®…
Centerpulse Orthopedics, Inc.2003—

Also cleared against K993082: 8 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K071676Exactech Novation Crown Cup Constrained Liners and RingsExactech, Inc.20070

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 11, 2022Class IIZ-1725-2022Process controlSpecific GXL acetabular polyethylene liners, packaged in non-conforming bags, may adversely impact the device and contribute to accelerated wear.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 7 years and passes through 4 companies; a new generation was cleared every 1.0 years on median.
  • 2 of the 10 ancestors have a recall of any class.
  • Of its descendants, K071676 (Exactech, Inc., 2007) has gone furthest, with 0 descendants of its own.
  • 8 other devices also cite its predicate K993082.

How does its category compare?

All KWZ devices

KWZ Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KWZ

1in 2025

2007: 72008: 12009: 12010: 12011: 12012: 12016: 12017: 22024: 12025: 1
2007high 7 (2007) · low 0 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KWZ

104in 2007

2007: 104 days median
2007high 104 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KWZ

1in 2023

2010: 82012: 12023: 12026 (partial year): 1
2007high 8 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KWZ

193in 2025

2007: 412008: 442009: 812010: 982011: 1272012: 1682013: 2742014: 2782015: 6362016: 1,1802017: 4772018: 2832019: 5322020: 2882021: 2162022: 1582023: 1642024: 1452025: 193
2007high 1,180 (2016) · low 41 (2007)2025

MAUDE reports naming product code KWZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077104041
20081—044
20091—081
20101—898
20111—0127
20121—1168
20130—0274
20140—0278
20150—0636
20161—01,180
20172—0477
20180—0283
20190—0532
20200—0288
20210—0216
20220—0158
20230—1164
20241—0145
20251—0193
20260—1—

What is it indicated for?

Exactech AcuMatch A-Series® and MCS® Constrained Acetabular Liners are compoients of the Exactech AcuMatch A-Series or MCS non-cemented acetabular cup system. The device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered. They are intended for press-fit fixation and compatible with Exactech press-fit or cemented femoral stem components.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K040601: Exactech Acumatch a-Series & MCS Constrained Acetabular Liners.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K040601

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026