510(k) Report

510(k) · Cleared

Recalled · Class I · 2022

Filter Carestar 30; Filter Safestar 55; Filter/Hme Twinstar 55

K072002510(k) TraditionalCleared Jan 31, 2008192-day reviewClass II21 CFR 868.5260Recalls 1 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Draeger Medical AG & Co. KG · CAH Filter, Bacterial, Breathing-Circuit

Cleared on Jan 31, 2008 as substantially equivalent to K031653 (Datex-Ohmeda, 2003) and 4 other predicates after a 192-day review. Its family tree reaches back 5 generations to 1994. 2 of the 22 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it. It has been recalled once, including a Class I recall.

Decided
Jan 31, 2008
Review
192 days
Class
Class II
Product code
CAH
Recalls
1 (1 Class I)
Predicates
5

Its longest line of descent runs back 5 generations to K941381 (1994). 18 ancestors in all, the oldest K881657 (1988). 3 later devices descend from it, 2 citing it directly. 2 of the 22 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor2003: 1 other ancestor2002: 2 other ancestors2001: 3 other ancestors1995: 1 other ancestor1994: 3 other ancestors1988: 1 other ancestor1996: 1 other ancestor2014: 1 descendant cleared2009: 1 descendant cleared2022: 1 descendant cleared, 1 with a Class I recall1K941381 · Hygrobac,S Hygroboy and Hygrobaby · Mallinckrodt Medical · 1994 · recalledK013089 · Ems Hepa Filter and Hme Combinations · Engineered Medical Systems · 2002×2: 2 generations cleared close together, K021265 (2002), K023641 (2002) K030990 · Datex-Ohmeda Mini-Filter/S and Uni-Filter Junior · Datex-Ohmeda · 20031990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors2 generations back: 9 other ancestors3 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 1 descendant, 1 with a Class I recall1 K030990 · Datex-Ohmeda Mini-Filter/S and Uni-Filter Junior · Datex-Ohmeda · 2003K023641 · Datex-Ohmeda Hmef Mini · Datex-Ohmeda · 2002K021265 · Datex-Ohmeda Hmef 500 · Datex-Ohmeda · 2002K013089 · Ems Hepa Filter and Hme Combinations · Engineered Medical Systems · 2002K941381 · Hygrobac,S Hygroboy and Hygrobaby · Mallinckrodt Medical · 1994 · recalled−50
  1. Root · 5 generations backK941381 Hygrobac,S Hygroboy and HygrobabyMallinckrodt Medical · 1994 · Recalled
  2. 4 generations backK013089 Ems Hepa Filter and Hme CombinationsEngineered Medical Systems · 2002
  3. 3 generations backK021265 Datex-Ohmeda Hmef 500Datex-Ohmeda · 2002
  4. 2 generations backK023641 Datex-Ohmeda Hmef MiniDatex-Ohmeda · 2002
  5. 1 generation backK030990 Datex-Ohmeda Mini-Filter/S and Uni-Filter JuniorDatex-Ohmeda · 2003
  6. This deviceK072002 Filter Carestar 30; Filter Safestar 55; Filter/Hme Twinstar 55Draeger Medical AG & Co. KG · 2008
  7. Descendant · cites it directly · 1 of its ownK132709 VR Medical Heat and Moisture Exchanger and FilterVr Medical · 2014
  8. Descendant · 2 generations on · 0 of its ownK221836 Filter CareStar Plus, Filter SafeStar Plus, Filter/HME TwinStar PlusDrägerwerk AG & Co. KGaA · 2022 · Recalled (Class I)
  9. Descendant · cites it directly · 0 of its ownK090738 Thermoflo HCH, Thermoflo FilterArc Medical, Inc. · 2009

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K132709VR Medical Heat and Moisture Exchanger and FilterVr Medical · 2014 · cites it directly1descendant
  • K221836Filter CareStar Plus, Filter SafeStar Plus, Filter/HME TwinStar PlusDrägerwerk AG & Co. KGaA · 2022 · 2 generations on0descendants
  • K090738Thermoflo HCH, Thermoflo FilterArc Medical, Inc. · 2009 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K941381Hygrobac,S Hygroboy and HygrobabyMallinckrodt Medical19941
−4K013089Ems Hepa Filter and Hme CombinationsEngineered Medical Systems2002—
−3K021265Datex-Ohmeda Hmef 500Datex-Ohmeda2002—
−2K023641Datex-Ohmeda Hmef MiniDatex-Ohmeda2002—
−1K030990Datex-Ohmeda Mini-Filter/S and Uni-Filter JuniorDatex-Ohmeda2003—
0 (this device)K072002Filter Carestar 30; Filter Safestar 55; Filter/Hme Twinstar 55Draeger Medical AG & Co. KG2008—
+1K132709VR Medical Heat and Moisture Exchanger and FilterVr Medical2014—
+2K221836Filter CareStar Plus, Filter SafeStar Plus, Filter/HME TwinStar PlusDrägerwerk AG & Co. KGaA2022Class I
+1K090738Thermoflo HCH, Thermoflo FilterArc Medical, Inc.2009—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
199430
199510
199610
200130
200220
200310
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201410
202211
FIG. 1 — Lineage of K072002, 1994–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CAH device

  1. #1 K251583 Ambu® Virobac II® Exhalation Filterscore 11.6 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K031653Datex-Ohmeda Hmef 750
Summary page 2
…(73CAH) 868.5375 Heat and moisture condenser (artificial nose) (73BYD) Legally Marketed Device to which Substantial Equivalence is claimed: K031653 Datex-Ohmeda HMEF 750 Datex-Ohmeda K033008 Air Safety HEPA and non-HEPA Fil…
Datex-Ohmeda2003—
*K033008Air Safety Hepa and Non-Hepa Filters
Summary page 2
…se) (73BYD) Legally Marketed Device to which Substantial Equivalence is claimed: K031653 Datex-Ohmeda HMEF 750 Datex-Ohmeda K033008 Air Safety HEPA and non-HEPA Filters AirSafety K030990 Datex-Ohmeda Mini-Filter/S…
Air Safety, Ltd.2004—
*K030990Datex-Ohmeda Mini-Filter/S and Uni-Filter Junior
Summary page 2
…K031653 Datex-Ohmeda HMEF 750 Datex-Ohmeda K033008 Air Safety HEPA and non-HEPA Filters AirSafety K030990 Datex-Ohmeda Mini-Filter/S Datex-Ohmeda Datex-Ohmeda Uni-Filter Junior…
Datex-Ohmeda2003—
*K014282Bacstop Filter,Bacstop Filter/Hme
Summary page 2
…A Filters AirSafety K030990 Datex-Ohmeda Mini-Filter/S Datex-Ohmeda Datex-Ohmeda Uni-Filter Junior K014282 Bacstop, Bacstop mini, Bacstop Humini Munktell K013089 EMS HEPA Filter…
Munktell Filter AB2002—
*K013089Ems Hepa Filter and Hme Combinations
Summary page 2
…Datex-Ohmeda Datex-Ohmeda Uni-Filter Junior K014282 Bacstop, Bacstop mini, Bacstop Humini Munktell K013089 EMS HEPA Filter EMS Device Description: The filters CareStar 30 and SafeStar 55 a…
Engineered Medical Systems2002—

Also cleared against K031653: 1 other device →

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K132709VR Medical Heat and Moisture Exchanger and FilterVr Medical20141
K090738Thermoflo HCH, Thermoflo FilterArc Medical, Inc.20090

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 16, 2022Class IZ-1211-2022Process designPossible occlusion of filters due to manufacturing error.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K221836Filter CareStar Plus, Filter SafeStar Plus, Filter/HME TwinStar PlusDrägerwerk AG & Co. KGaA · 2022 · 2 generations on · ◆ in the 2022 descendants bar of FIG. 14 Class I recalls

What does this lineage say?

  • The line of descent spans 14 years and passes through 4 companies; a new generation was cleared every 7 months on median.
  • 3 of the 18 ancestors have a recall of any class.
  • Of its descendants, K132709 (Vr Medical, 2014) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K031653.

How does its category compare?

All CAH devices

CAH Filter, Bacterial, Breathing-Circuit: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · CAH

2in 2024

2007: 22008: 12009: 42010: 52011: 32012: 32013: 12014: 42015: 22017: 12019: 22020: 22021: 22022: 32023: 32024: 22026 (partial year): 2
2007high 5 (2010) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CAH

237in 2023

2009: 111 days median2010: 216 days median2011: 281 days median2012: 212 days median2014: 250 days median2022: 204 days median2023: 237 days median
2007high 281 (2011) · low 111 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CAH

1in 2025

2010: 22013: 12014: 12017: 12022: 6, 6 Class I2025: 1, 1 Class I
2007high 6 (2022) · low 0 (2007)2026

Class I (7 since 2007) Class II/III not classified

Adverse event reports per year · CAH

77in 2025

2007: 322008: 272009: 182010: 402011: 92012: 162013: 472014: 982015: 512016: 342017: 292018: 332019: 302020: 332021: 562022: 672023: 402024: 512025: 77
2007high 98 (2014) · low 9 (2011)2025

MAUDE reports naming product code CAH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—032
20081—027
20094111018
20105216240
2011328109
20123212016
20131—147
20144250198
20152—051
20160—034
20171—129
20180—033
20192—030
20202—033
20212—056
202232046 (6)67
20233237040
20242—051
20250—1 (1)77
20262—0—

What is it indicated for?

CareStar 30 and SafeStar 55 are Breathing System Filters which are designed to reduce possible airborne or liquid-borne cross contamination with micro-organisms and particulate matter via anaesthetic or ventilator breathing systems. The products may either be used on the patient side or on the device side of the ventilator/ anaesthetic device and are used as a hygienic measure alternatively to decontamination of breathing system and/or breathing gas…

Read the full statement (182 words)

CareStar 30 and SafeStar 55 are Breathing System Filters which are designed to reduce

possible airborne or liquid-borne cross contamination with micro-organisms and particulate

matter via anaesthetic or ventilator breathing systems.

The products may either be used on the patient side or on the device side of the ventilator/

anaesthetic device and are used as a hygienic measure alternatively to decontamination of

breathing system and/or breathing gas conveying parts of the ventilator.

TwinStar 55 is a Breathing System Filter and a Heat and Moisture Exchanger. The

combination of a filter and a Heat and Moisture Exchanger offer the benefit of both product

features. Heat and Moisture Exchanger are used as a conditioning system for mechanically

ventilated patients whose upper airways are bypassed. In almost all cases of mechanical

ventilation they are a fully valid alternative to heated humidifiers. The products are the only

conditioning opportunity of breathing gases in cases of emergency ventilation or during

transport since Heated Humidifiers are almost impossible to use.

The products mentioned above are designed for disposable use and should be changed at

least every 24 hours.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K072002: Filter Carestar 30; Filter Safestar 55; Filter/Hme Twinstar 55.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K072002

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026