510(k) Report

510(k) · Cleared

Cleared

Infinitas Mini-Implant System/Guidance System

K072086510(k) TraditionalCleared Oct 26, 200788-day reviewClass II21 CFR 872.3640

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

D B Orthodontics, Ltd. · OAT Implant, Endosseous, Orthodontic

Cleared on Oct 26, 2007 after an 88-day review. FDA’s file for it is a 510(k) statement, which names no predicate.

Decided
Oct 26, 2007
Review
88 days
Class
Class II
Product code
OAT
Recalls
None
Predicates
0

FDA’s file for it is a 510(k) statement, which names no predicate.

FIG. 1 — Lineage of K072086Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OAT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OAT device

  1. #1 K260268 OrthoEasy Pal; OrthoEasy Pal Long Neckscore 13.2 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

How does its category compare?

All OAT devices

OAT Implant, Endosseous, Orthodontic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · OAT

1in 2025

2007: 62008: 32009: 22010: 22011: 52012: 12013: 22014: 12015: 12016: 22017: 12018: 12019: 22020: 32021: 42022: 22023: 12024: 22025: 12026 (partial year): 1
2007high 6 (2007) · low 1 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OAT

224in 2021

2007: 104 days median2008: 94 days median2011: 152 days median2020: 289 days median2021: 224 days median
2007high 289 (2020) · low 94 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OAT

2in 2007

2007: 2
2007high 2 (2007) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OAT

5in 2025

2007: 22010: 12012: 32013: 22014: 42015: 282016: 122018: 12020: 382021: 682022: 842023: 782024: 92025: 5
2007high 84 (2022) · low 0 (2008)2025

MAUDE reports naming product code OAT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007610422
200839400
20092—00
20102—01
2011515200
20121—03
20132—02
20141—04
20151—028
20162—012
20171—00
20181—01
20192—00
20203289038
20214224068
20222—084
20231—078
20242—09
20251—05
20261—0—

What is it indicated for?

Temporary anchorage point for the attachment of orthodontic appliances

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K072086: Infinitas Mini-Implant System/Guidance System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K072086

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026