510(k) · Cleared
Recalled · Class II · 2014Foundation (FMP) Porous Coated Spiked Acetabular System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Encore Medical L.P. · LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Cleared on Dec 14, 2007 as substantially equivalent to K974093 (Encore Orthopedics, Inc., 1998) and 3 other predicates after a 133-day review. Its family tree reaches back 1 generation to 1997. 61 later devices descend from it. It has been recalled once.
- Decided
- Dec 14, 2007
- Review
- 133 days
- Class
- Class II
- Product code
- LPH
- Recalls
- 1
- Predicates
- 4
Its longest line of descent runs back 1 generation to K973302 (1997). 4 ancestors in all. 61 later devices descend from it, 4 citing it directly.
- Root · 1 generation backK973302 Foundation Porous Hip StemEncore Orthopedics, Inc. · 1997
- This deviceK072154 Foundation (FMP) Porous Coated Spiked Acetabular SystemEncore Medical L.P. · 2007
- Descendant · cites it directly · 38 of its ownK130365 X-Alt Highly Cross Linked VE Acetabular LinerEncore Medical L.P. · 2013
- Descendant · 2 generations on · 21 of its ownK141006 RSP Humeral Socket InsertDjo Global · 2014 · Recalled
- Descendant · 3 generations on · 18 of its ownK172351 AltiVate Reverse Humeral Stem, AltiVate Reverse Small Spacer…Encore Medical L.P. · 2017 · Recalled
- Descendant · 4 generations on · 15 of its ownK190290 AltiVate Reverse Humeral Stem, 48mmEncore Medical L.P. · 2019
- Descendant · cites it directly · 13 of its ownK112158 Modulus Stems, Modulus Necks, Modular Femoral Heads, Cemented Cups…Lima Corporate S.P.A. · 2012
Showing 5 of 61 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 61 · See all- K130365X-Alt Highly Cross Linked VE Acetabular LinerEncore Medical L.P. · 2013 · cites it directly38descendants
- K141006RSP Humeral Socket InsertDjo Global · 2014 · 2 generations on21descendants
- K172351AltiVate Reverse Humeral Stem, AltiVate Reverse Small Spacer, Altivate Reverse, Small…Encore Medical L.P. · 2017 · 3 generations on18descendants
- K190290AltiVate Reverse Humeral Stem, 48mmEncore Medical L.P. · 2019 · 4 generations on15descendants
- K112158Modulus Stems, Modulus Necks, Modular Femoral Heads, Cemented Cups, Lock Bipolar HeadsLima Corporate S.P.A. · 2012 · cites it directly13descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K973302 | Foundation Porous Hip Stem | Encore Orthopedics, Inc. | 1997 | — |
| 0 (this device) | K072154 | Foundation (FMP) Porous Coated Spiked Acetabular System | Encore Medical L.P. | 2007 | — |
| +1 | K130365 | X-Alt Highly Cross Linked VE Acetabular Liner | Encore Medical L.P. | 2013 | — |
| +2 | K141006 | RSP Humeral Socket Insert | Djo Global | 2014 | 2 |
| +3 | K172351 | AltiVate Reverse Humeral Stem, AltiVate Reverse Small Spacer, Altivate… | Encore Medical L.P. | 2017 | 3 |
| +4 | K190290 | AltiVate Reverse Humeral Stem, 48mm | Encore Medical L.P. | 2019 | — |
| +1 | K112158 | Modulus Stems, Modulus Necks, Modular Femoral Heads, Cemented Cups, Lock… | Lima Corporate S.P.A. | 2012 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1998 | 3 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2012 | 2 | 0 |
| 2013 | 1 | 0 |
| 2014 | 6 | 0 |
| 2015 | 2 | 0 |
| 2016 | 3 | 0 |
| 2017 | 3 | 0 |
| 2018 | 4 | 0 |
| 2019 | 6 | 0 |
| 2020 | 4 | 0 |
| 2021 | 5 | 0 |
| 2022 | 5 | 0 |
| 2023 | 6 | 0 |
| 2024 | 3 | 0 |
| 2025 | 8 | 0 |
| 2026 | 3 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in LPH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LPH device
- #1 K233498 Stryker Orthopaedics Hip Systems Labeling Updatescore 24.2 · 20 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (4)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K974093 | Foundation Porous Coated Hemispherical ShellSummary page 1…Product Classification .Product Code Foundation® (FMPTM) Porous Coated Spiked Acetabular K974093, January 2$, 1998, Class II LPH System Foundation® (FMP T M ) Porous Coated Hemispherical K974095, Januar… | 1998 |
| * | K974095 | Foundation Porous Coated Flared Rim ShellSummary page 1…PTM) Porous Coated Spiked Acetabular K974093, January 2$, 1998, Class II LPH System Foundation® (FMP T M ) Porous Coated Hemispherical K974095, January 28, 1998, Class 11 LPH Acetabular System Foundation® (FMP T M ) Porous Coated Flared Rim K973… | 1998 |
| * | K973119 | Foundation Porous Acetabular SystemSummary page 1…ous Coated Hemispherical K974095, January 28, 1998, Class 11 LPH Acetabular System Foundation® (FMP T M ) Porous Coated Flared Rim K973119, January 28, 1998, Class 11 LPH Acetabular System Foundation® Press Fit (Porous Coated) Hip Stem K9733… | 1998 |
| * | K973302 | Foundation Porous Hip StemSummary page 1…rous Coated Flared Rim K973119, January 28, 1998, Class 11 LPH Acetabular System Foundation® Press Fit (Porous Coated) Hip Stem K973302 - December 2,1997, Class 11 LPH Product Code Regulation and Classification… | 1997 |
What descends from it? (61)
See all 614 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K130365 | X-Alt Highly Cross Linked VE Acetabular Liner | 2013 | 38 |
| K112158 | Modulus Stems, Modulus Necks, Modular Femoral Heads, Cemented Cups, Lock Bipolar Heads | 2012 | 13 |
| K112898 | Delta TT Cups, Delta TT Liners, Bone Screws | 2012 | 12 |
| K191708 | Origin Stem, Evolve Stem, Aria Hip Stem, Remedy Stem, CoCr Femoral Head | 2020 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 7, 2014 | Class II | Z-1130-2014 | Packaging process control | Packaging error -- two different types and sizes of acetabular liners used during hip replacement surgery were mistakenly packaged into each other's outer boxes after sterilization, resulting in the products being mislabeled. The units… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All LPH devicesLPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.
FDA decisions per year · LPH
12in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LPH
153in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LPH
4in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · LPH
6,163in 2025
MAUDE reports naming product code LPH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 10 | 116 | 16 | 509 |
| 2008 | 10 | 86 | 1 | 706 |
| 2009 | 3 | 112 | 2 | 1,156 |
| 2010 | 15 | 100 | 22 | 2,007 |
| 2011 | 16 | 110 | 26 | 3,320 |
| 2012 | 25 | 76 | 30 | 3,785 |
| 2013 | 11 | 135 | 9 | 5,002 |
| 2014 | 20 | 129 | 3 | 5,163 |
| 2015 | 18 | 111 | 1 | 5,830 |
| 2016 | 10 | 150 | 8 | 4,900 |
| 2017 | 12 | 121 | 14 | 5,312 |
| 2018 | 9 | 145 | 2 | 5,954 |
| 2019 | 22 | 164 | 8 | 7,130 |
| 2020 | 7 | 164 | 6 | 5,429 |
| 2021 | 8 | 410 | 4 | 5,978 |
| 2022 | 14 | 247 | 2 | 5,223 |
| 2023 | 10 | 111 | 2 | 5,345 |
| 2024 | 6 | 92 | 1 | 5,626 |
| 2025 | 12 | 153 | 4 | 6,163 |
| 2026 | 8 | 256 | 2 | — |
What is it indicated for?
Joint replacement is indicated for patients suffering from disability due to:
· noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the
natural femoral head;
· rheumatoid arthritis;
* correction of functional deformity;
* femoral fracture.
This device may also be indicated in the salvage of previously failed surgical attempts. It is to be used
for cementless applications only.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K072154 · printed Nov 6, 2026
Cite this page
510(k) Report. “K072154: Foundation (FMP) Porous Coated Spiked Acetabular System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K072154
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026