510(k) Report

510(k) · Cleared

Recalled · 2009

PM-50 Pulse Oximeter and VS-800 Vital Signs Monitor

K072235510(k) SpecialCleared Oct 19, 200770-day reviewClass II21 CFR 870.2700Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · DQA Oximeter

Cleared on Oct 19, 2007 as substantially equivalent to K061442 (Shenzhen Mindray Bio-Medical Electronics Co., Ltd., 2006) and 1 other predicate after a 70-day review. Its family tree reaches back 8 generations to 1994. 7 later devices descend from it. It has been recalled 2 times.

Decided
Oct 19, 2007
Review
70 days
Class
Class II
Product code
DQA
Recalls
2
Predicates
2

Its longest line of descent runs back 8 generations to K933143 (1994). 44 ancestors in all, the oldest K842616 (1984). 7 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2006: 2 other ancestors2000: 2 other ancestors1997: 4 other ancestors1996: 4 other ancestors1995: 5 other ancestors1994: 4 other ancestors1992: 1 other ancestor1991: 5 other ancestors1990: 3 other ancestors1989: 2 other ancestors1999: 1 other ancestor1998: 1 other ancestor1993: 1 other ancestor1984: 1 other ancestor2014: 2 descendants cleared2022: 1 descendant cleared2021: 1 descendant cleared2017: 1 descendant cleared2019: 1 descendant cleared2024: 1 descendant clearedK933143 · Model Eli 300 Electrocardiograph · Mortara Instrument, Inc. · 1994K962970 · Spacelabs Medical 12-Lead Diagnostic Ecg Analysis Option · Spacelabs Medical, Inc. · 1997K972502 · Spacelabs Medical Integrated Multiparameter Module 90496 · Spacelabs Medical, Inc. · 1998 · recalledK993531 · Passport 2 Vital Signs Monitor, Model 0998-00-0170-XX · Datascope Corp. · 2000 · recalledK032733 · PM-8000 Patient Monitor, Model 8000 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2004K043348 · Modification to PM-8000 Patient Monitor · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2005K060281 · VS-800 Vital Signs Monitor · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2006K063055 · VS-800 Vital Signs Monitor · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 200619801990200020102020
  1. Root · 8 generations backK933143 Model Eli 300 ElectrocardiographMortara Instrument, Inc. · 1994
  2. 7 generations backK962970 Spacelabs Medical 12-Lead Diagnostic Ecg Analysis OptionSpacelabs Medical, Inc. · 1997
  3. 6 generations backK972502 Spacelabs Medical Integrated Multiparameter Module 90496Spacelabs Medical, Inc. · 1998 · Recalled
  4. 5 generations backK993531 Passport 2 Vital Signs Monitor, Model 0998-00-0170-XXDatascope Corp. · 2000 · Recalled
  5. 4 generations backK032733 PM-8000 Patient Monitor, Model 8000Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2004
  6. 3 generations backK043348 Modification to PM-8000 Patient MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2005
  7. 2 generations backK060281 VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2006
  8. 1 generation backK063055 VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2006
  9. This deviceK072235 PM-50 Pulse Oximeter and VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2007
  10. Descendant · cites it directly · 3 of its ownK132037 Vital Signs MonitorMindray North America · 2014 · Recalled
  11. Descendant · cites it directly · 2 of its ownK132038 Accutorr 7 Vital Signs MonitorMindray North America · 2014
  12. Descendant · 2 generations on · 1 of its ownK202892 Vital Signs Monitor:iM3s, iM3As, iM3Bs, iHM3sEdan Instruments, Inc. · 2021 · Recalled
  13. Descendant · 2 generations on · 1 of its ownK170712 Accutorr 7/VS-900 Vital Signs MonitorShenzhen Mindray Bio-Meidcal Electronics Co., Ltd. · 2017 · Recalled
  14. Descendant · 3 generations on · 0 of its ownK233038 Vital Signs monitor, Model: iM3s, iM3As, iM3Bs, iHM3sEdan Instruments, Inc. · 2024 · Recalled

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2014: 22017: 12019: 12021: 12022: 12024: 1
2014high 2 (2014) · low 0 (2015)2026

Most influential descendants

Showing 5 of 7 · See all
  • K132037Vital Signs MonitorMindray North America · 2014 · cites it directly3descendants
  • K132038Accutorr 7 Vital Signs MonitorMindray North America · 2014 · cites it directly2descendants
  • K202892Vital Signs Monitor:iM3s, iM3As, iM3Bs, iHM3sEdan Instruments, Inc. · 2021 · 2 generations on1descendant
  • K170712Accutorr 7/VS-900 Vital Signs MonitorShenzhen Mindray Bio-Meidcal Electronics Co., Ltd. · 2017 · 2 generations on1descendant
  • K233038Vital Signs monitor, Model: iM3s, iM3As, iM3Bs, iHM3sEdan Instruments, Inc. · 2024 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K933143Model Eli 300 ElectrocardiographMortara Instrument, Inc.1994—
−7K962970Spacelabs Medical 12-Lead Diagnostic Ecg Analysis OptionSpacelabs Medical, Inc.1997—
−6K972502Spacelabs Medical Integrated Multiparameter Module 90496Spacelabs Medical, Inc.19983
−5K993531Passport 2 Vital Signs Monitor, Model 0998-00-0170-XXDatascope Corp.20001
−4K032733PM-8000 Patient Monitor, Model 8000Shenzhen Mindray Bio-Medical Electronics Co., Ltd.2004—
−3K043348Modification to PM-8000 Patient MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.2005—
−2K060281VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.2006—
−1K063055VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.2006—
0 (this device)K072235PM-50 Pulse Oximeter and VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.2007—
+1K132037Vital Signs MonitorMindray North America20142
+1K132038Accutorr 7 Vital Signs MonitorMindray North America2014—
+2K202892Vital Signs Monitor:iM3s, iM3As, iM3Bs, iHM3sEdan Instruments, Inc.20211
+2K170712Accutorr 7/VS-900 Vital Signs MonitorShenzhen Mindray Bio-Meidcal Electronics Co., Ltd.20171
+3K233038Vital Signs monitor, Model: iM3s, iM3As, iM3Bs, iHM3sEdan Instruments, Inc.20241

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198410
198920
199030
199150
199210
199310
199440
199550
199640
199740
199810
199910
200020
200620

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201420
201710
201910
202110
202210
202410
FIG. 1 — Lineage of K072235, 1994–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DQA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQA device

  1. #1 K221282 Empatica Health Monitoring Platformscore 20.1 · 38 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K061442Pulse Oximeter, Model PM-50
Summary page 1
…21 CFR 880.2910, Clinical Electronic Thermometer, 80 FLL Predicate Device K061442, PM-SO Pulse Oximeter, K063055, VS-800 Vital Signs Monitor,…
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.2006—
*K063055VS-800 Vital Signs Monitor
Summary page 1
…910, Clinical Electronic Thermometer, 80 FLL Predicate Device K061442, PM-SO Pulse Oximeter, K063055, VS-800 Vital Signs Monitor, cleared on Sep 5, 2006…
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.2006—

What descends from it? (7)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K132037Vital Signs MonitorMindray North America20143
K132038Accutorr 7 Vital Signs MonitorMindray North America20142

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 31, 2009Not classifiedZ-1397-2010Process controlSmarTemp Disposable Probe covers may contain microscopic holes in the molded seam that would allow liquids or body fluids inside the cover and potentially contaminate the temperature probe.
Dec 31, 2009Not classifiedZ-1398-2010Process controlSmarTemp Disposable Probe covers may contain microscopic holes in the molded seam that would allow liquids or body fluids inside the cover and potentially contaminate the temperature probe.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 1.5 years on median.
  • 6 of the 44 ancestors have a recall of any class.
  • Of its descendants, K132037 (Mindray North America, 2014) has gone furthest, with 3 descendants of its own.

How does its category compare?

All DQA devices

DQA Oximeter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · DQA

15in 2025

2007: 202008: 222009: 352010: 292011: 152012: 142013: 152014: 112015: 132016: 142017: 92018: 182019: 192020: 152021: 102022: 202023: 132024: 142025: 152026 (partial year): 8
2007high 35 (2009) · low 9 (2017)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQA

182in 2025

2007: 122 days median2008: 97 days median2009: 117 days median2010: 127 days median2011: 195 days median2012: 120 days median2013: 182 days median2014: 175 days median2015: 158 days median2016: 246 days median2017: 200 days median2018: 260 days median2019: 158 days median2020: 126 days median2021: 174 days median2022: 265 days median2023: 317 days median2024: 285 days median2025: 182 days median2026 (partial year): 187 days median
2007high 317 (2023) · low 97 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQA

3in 2024

2007: 62008: 52009: 292010: 42011: 12014: 22015: 12018: 322024: 3
2007high 32 (2018) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DQA

399in 2025

2007: 3252008: 3122009: 2812010: 1852011: 1642012: 3222013: 4682014: 1,1242015: 2,0822016: 1,7752017: 1,1902018: 6942019: 5482020: 4262021: 5532022: 4782023: 3922024: 5452025: 399
2007high 2,082 (2015) · low 164 (2011)2025

MAUDE reports naming product code DQA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007201226325
200822975312
20093511729281
2010291274185
2011151951164
2012141200322
2013151820468
20141117521,124
20151315812,082
20161424601,775
2017920001,190
20181826032694
2019191580548
2020151260426
2021101740553
2022202650478
2023133170392
2024142853545
2025151820399
202681870—

What is it indicated for?

The PM-50 Pulse Oximeter is a non-invasive, spot-check, oxygen saturation and pulse rate monitor. It operates only on battery power using existing PM-50 disposable and reusable sensors labeled for patients ranging from neonates to adults.

The VS-800 Vital Signs Monitor is used to monitor physiologic parameters including SpO2, PR and NIBP, and to measure Temperature parameter on adult, pediatric, and neonatal patients in healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. It is not intended for transport or home use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K072235: PM-50 Pulse Oximeter and VS-800 Vital Signs Monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K072235

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026