510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Accutorr 7/VS-900 Vital Signs Monitor

K170712510(k) TraditionalCleared Aug 2, 2017146-day reviewClass II21 CFR 870.2300Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Shenzhen Mindray Bio-Meidcal Electronics Co., Ltd. · MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

Cleared on Aug 2, 2017 as substantially equivalent to K132038 (Mindray North America, 2014) after a 146-day review. Its family tree reaches back 10 generations to 1994. 1 later device descends from it. It has been recalled once.

Decided
Aug 2, 2017
Review
146 days
Class
Class II
Product code
MWI
Recalls
1
Predicates
1

Its longest line of descent runs back 10 generations to K933143 (1994). 46 ancestors in all, the oldest K842616 (1984). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2006: 2 other ancestors2000: 2 other ancestors1997: 4 other ancestors1996: 4 other ancestors1995: 5 other ancestors1994: 4 other ancestors1992: 1 other ancestor1991: 5 other ancestors1990: 3 other ancestors1989: 2 other ancestors1999: 1 other ancestor1998: 1 other ancestor1993: 1 other ancestor1984: 1 other ancestor2019: 1 descendant clearedReference device: K152902 · Passport Series Patient Monitors (including Passport 17m, Passport 12m and T1) · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2016 · recalledK933143 · Model Eli 300 Electrocardiograph · Mortara Instrument, Inc. · 1994K962970 · Spacelabs Medical 12-Lead Diagnostic Ecg Analysis Option · Spacelabs Medical, Inc. · 1997K972502 · Spacelabs Medical Integrated Multiparameter Module 90496 · Spacelabs Medical, Inc. · 1998 · recalledK993531 · Passport 2 Vital Signs Monitor, Model 0998-00-0170-XX · Datascope Corp. · 2000 · recalledK032733 · PM-8000 Patient Monitor, Model 8000 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2004K043348 · Modification to PM-8000 Patient Monitor · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2005K060281 · VS-800 Vital Signs Monitor · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2006K063055 · VS-800 Vital Signs Monitor · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2006K072235 · PM-50 Pulse Oximeter and VS-800 Vital Signs Monitor · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2007 · recalledK132038 · Accutorr 7 Vital Signs Monitor · Mindray North America · 201419801990200020102020
  1. Root · 10 generations backK933143 Model Eli 300 ElectrocardiographMortara Instrument, Inc. · 1994
  2. 6 generations between
  3. 3 generations backK063055 VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2006
  4. 2 generations backK072235 PM-50 Pulse Oximeter and VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2007 · Recalled
  5. 1 generation backK132038 Accutorr 7 Vital Signs MonitorMindray North America · 2014
  6. This deviceK170712 Accutorr 7/VS-900 Vital Signs MonitorShenzhen Mindray Bio-Meidcal Electronics Co., Ltd. · 2017
  7. Descendant · cites it directly · 0 of its ownK182821 Accutorr 7/VS-900/VS-900c Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2019 · Recalled
Show all 10 generations
  1. Root · 10 generations backK933143 Model Eli 300 ElectrocardiographMortara Instrument, Inc. · 1994
  2. 9 generations backK962970 Spacelabs Medical 12-Lead Diagnostic Ecg Analysis OptionSpacelabs Medical, Inc. · 1997
  3. 8 generations backK972502 Spacelabs Medical Integrated Multiparameter Module 90496Spacelabs Medical, Inc. · 1998 · Recalled
  4. 7 generations backK993531 Passport 2 Vital Signs Monitor, Model 0998-00-0170-XXDatascope Corp. · 2000 · Recalled
  5. 6 generations backK032733 PM-8000 Patient Monitor, Model 8000Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2004
  6. 5 generations backK043348 Modification to PM-8000 Patient MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2005
  7. 4 generations backK060281 VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2006
  8. 3 generations backK063055 VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2006
  9. 2 generations backK072235 PM-50 Pulse Oximeter and VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2007 · Recalled
  10. 1 generation backK132038 Accutorr 7 Vital Signs MonitorMindray North America · 2014
  11. This deviceK170712 Accutorr 7/VS-900 Vital Signs MonitorShenzhen Mindray Bio-Meidcal Electronics Co., Ltd. · 2017

Most influential descendants

Showing 1 of 1 · See all
  • K182821Accutorr 7/VS-900/VS-900c Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2019 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K933143Model Eli 300 ElectrocardiographMortara Instrument, Inc.1994—
−9K962970Spacelabs Medical 12-Lead Diagnostic Ecg Analysis OptionSpacelabs Medical, Inc.1997—
−8K972502Spacelabs Medical Integrated Multiparameter Module 90496Spacelabs Medical, Inc.19983
−7K993531Passport 2 Vital Signs Monitor, Model 0998-00-0170-XXDatascope Corp.20001
−6K032733PM-8000 Patient Monitor, Model 8000Shenzhen Mindray Bio-Medical Electronics Co., Ltd.2004—
−5K043348Modification to PM-8000 Patient MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.2005—
−4K060281VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.2006—
−3K063055VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.2006—
−2K072235PM-50 Pulse Oximeter and VS-800 Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.20072
−1K132038Accutorr 7 Vital Signs MonitorMindray North America2014—
0 (this device)K170712Accutorr 7/VS-900 Vital Signs MonitorShenzhen Mindray Bio-Meidcal Electronics Co., Ltd.2017—
+1K182821Accutorr 7/VS-900/VS-900c Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.20191

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198410
198920
199030
199150
199210
199310
199440
199550
199640
199740
199810
199910
200020
200620

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
FIG. 1 — Lineage of K170712, 1994–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MWI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MWI device

  1. #1 K230457 Empatica Health Monitoring Platform…score 28.0 · 39 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K132038Accutorr 7 Vital Signs Monitor
Summary page 5
…Page 1 K170712 Primary Predicate Device: K132038 - Accutorr 7 Vital Signs Monitor; Shenzhen Mindray Bio-Medical Electronics Co., Ltd Reference Predicat…
Mindray North America2014—
†K152902Passport Series Patient Monitors (including Passport 17m, Passport 12m and T1)Shenzhen Mindray Bio-Medical Electronics Co., Ltd.20161

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K182821Accutorr 7/VS-900/VS-900c Vital Signs MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.20190

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 1, 2019Class IIZ-0237-2020No Marketing ApplicationNIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 6 companies; a new generation was cleared every 1.6 years on median.
  • 7 of the 46 ancestors have a recall of any class.
  • Of its descendants, K182821 (Shenzhen Mindray Bio-Medical Electronics Co., Ltd., 2019) has gone furthest, with 0 descendants of its own.

How does its category compare?

All MWI devices

MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · MWI

4in 2025

2007: 92008: 92009: 62010: 172011: 112012: 132013: 202014: 162015: 192016: 122017: 132018: 112019: 142020: 82021: 92022: 52023: 82024: 102025: 42026 (partial year): 6
2007high 20 (2013) · low 4 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MWI

247in 2025

2007: 42 days median2008: 49 days median2009: 112 days median2010: 58 days median2011: 78 days median2012: 104 days median2013: 85 days median2014: 148 days median2015: 127 days median2016: 165 days median2017: 205 days median2018: 217 days median2019: 255 days median2020: 138 days median2021: 161 days median2022: 231 days median2023: 250 days median2024: 202 days median2025: 247 days median2026 (partial year): 239 days median
2007high 255 (2019) · low 42 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MWI

9so far in 2026, the most since 2007

2007: 12008: 12009: 22010: 22011: 32012: 52013: 32014: 62015: 12017: 22019: 42022: 22024: 52026 (partial year): 9
2007high 9 (2026) · low 0 (2016)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MWI

182in 2025

2007: 32008: 32009: 202010: 172011: 162012: 312013: 422014: 302015: 312016: 512017: 852018: 1582019: 1692020: 1102021: 1882022: 1682023: 1792024: 1202025: 182
2007high 188 (2021) · low 3 (2007)2025

MAUDE reports naming product code MWI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200794213
200894913
20096112220
20101758217
20111178316
201213104531
20132085342
201416148630
201519127131
201612165051
201713205285
2018112170158
2019142554169
202081380110
202191610188
202252312168
202382500179
2024102025120
202542470182
202662399—

What is it indicated for?

The Accutorr 7/VS-900 Vital Signs Monitor is intended for monitoring physiologic parameters, including Pulse Oximetry (SpO2), Pulse Rate (PR), Non Invasive Blood Pressure (NIBP), Temperature (TEMP) and Carbon Dioxide (CO2), on adult, pediatric, and neonatal patients in professional healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians.

The monitor also provides a Modified Early Warning Score (MEWS) for clinical assessment in adult patients.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K170712: Accutorr 7/VS-900 Vital Signs Monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K170712

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026