510(k) · Cleared
Recalled · Class I · 2025Magnetom Verio
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Siemens Medical Solutions USA, Inc. · LNH System, Nuclear Magnetic Resonance Imaging
Cleared on Oct 10, 2007 as substantially equivalent to K050200 (Siemens Medical Solutions USA, Inc., 2005) and 1 other predicate after a 58-day review. Its family tree reaches back 5 generations to 1997. 22 of the 310 devices in its family, counting itself, have a Class I recall. 300 later devices descend from it. It has been recalled 4 times, including a Class I recall.
- Decided
- Oct 10, 2007
- Review
- 58 days
- Class
- Class II
- Product code
- LNH
- Recalls
- 4 (1 Class I)
- Predicates
- 2
Its longest line of descent runs back 5 generations to K971684 (1997). 9 ancestors in all. 300 later devices descend from it, 9 citing it directly. 22 of the 310 devices in this family, counting this one, have a Class I recall.
- Root · 5 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
- 4 generations backK032428 Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc. · 2003 · Recalled
- 3 generations backK041112 Magnetom Espree SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
- 2 generations backK052164 Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc. · 2005
- 1 generation backK062454 Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc. · 2006
- This deviceK072237 Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007
- Descendant · cites it directly · 229 of its ownK082427 Syngo MR B17Siemens Medical Solutions USA, Inc. · 2008
- Descendant · cites it directly · 224 of its ownK101347 Magnetom Aera; Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2010 · Recalled (Class I)
- Descendant · cites it directly · 221 of its ownK103275 Specialty Coils for Magnetom Aera 1.5T & Magnetim Skyra 3T MR SystemsSiemens Medical Solutions USA, Inc. · 2011
- Descendant · cites it directly · 220 of its ownK111242 Magnetom Aera and Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2011 · Recalled
- Descendant · 2 generations on · 215 of its ownK121434 Software Syngo MR D13A for the Magnetom Systems…Siemens Medi Cal Solutions, Inc. · 2012 · Recalled (Class I)
Showing 5 of 300 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 300 · See all- K082427Syngo MR B17Siemens Medical Solutions USA, Inc. · 2008 · cites it directly229descendants
- K101347Magnetom Aera; Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2010 · cites it directly224descendants
- K103275Specialty Coils for Magnetom Aera 1.5T & Magnetim Skyra 3T MR SystemsSiemens Medical Solutions USA, Inc. · 2011 · cites it directly221descendants
- K111242Magnetom Aera and Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2011 · cites it directly220descendants
- K121434Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/VerioSiemens Medi Cal Solutions, Inc. · 2012 · 2 generations on215descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −5 | K971684 | Magnetom Project 047 System (Numaris 3.5 Software) | Siemens Medical Solutions USA, Inc. | 1997 | 4 |
| −4 | K032428 | Magnetom Avanto MR System | Siemens Medical Solutions USA, Inc. | 2003 | 4 |
| −3 | K041112 | Magnetom Espree System | Siemens Medical Solutions USA, Inc. | 2004 | 4 |
| −2 | K052164 | Magnetom Systems with Software Syngo MR 2006A | Siemens Medical Solutions USA, Inc. | 2005 | — |
| −1 | K062454 | Magnetom Systems - I-Class and T-Class Releases | Siemens Medi Cal Solutions, Inc. | 2006 | — |
| 0 (this device) | K072237 | Magnetom Verio | Siemens Medical Solutions USA, Inc. | 2007 | — |
| +1 | K082427 | Syngo MR B17 | Siemens Medical Solutions USA, Inc. | 2008 | — |
| +1 | K101347 | Magnetom Aera; Magnetom Skyra | Siemens Medical Solutions USA, Inc. | 2010 | Class I |
| +1 | K103275 | Specialty Coils for Magnetom Aera 1.5T & Magnetim Skyra 3T MR Systems | Siemens Medical Solutions USA, Inc. | 2011 | — |
| +1 | K111242 | Magnetom Aera and Magnetom Skyra | Siemens Medical Solutions USA, Inc. | 2011 | 3 |
| +2 | K121434 | Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/Verio | Siemens Medi Cal Solutions, Inc. | 2012 | Class I |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1999 | 1 | 0 |
| 2001 | 1 | 0 |
| 2005 | 2 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2008 | 3 | 0 |
| 2009 | 1 | 0 |
| 2010 | 2 | 1 |
| 2011 | 6 | 0 |
| 2012 | 8 | 2 |
| 2013 | 18 | 2 |
| 2014 | 9 | 0 |
| 2015 | 12 | 1 |
| 2016 | 9 | 1 |
| 2017 | 20 | 1 |
| 2018 | 21 | 2 |
| 2019 | 23 | 1 |
| 2020 | 28 | 3 |
| 2021 | 12 | 1 |
| 2022 | 29 | 2 |
| 2023 | 32 | 3 |
| 2024 | 20 | 0 |
| 2025 | 24 | 1 |
| 2026 | 23 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
22 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LNH device
- #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K050200 | Magnetom Trio a Tim SystemSummary page 2…___ ___ ___ ______ ___ ___ ___ __N um ber Siemens 3T MAGNETOM Trio a K050200 Feb 28, 2005 Tim System software syngo MR 815 (VB15A) K062454… | 2005 |
| * | K062454 | Magnetom Systems - I-Class and T-Class ReleasesSummary page 2…Trio a K050200 Feb 28, 2005 Tim System software syngo MR 815 (VB15A) K062454 Nov 3, 2006 Siemens 510(k) Premarket Notification August 8, 2007 Page 12… | 2006 |
What descends from it? (300)
See all 3009 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K082427 | Syngo MR B17 | 2008 | 229 |
| K101347 | Magnetom Aera; Magnetom Skyra Class I | 2010 | 224 |
| K103275 | Specialty Coils for Magnetom Aera 1.5T & Magnetim Skyra 3T MR Systems | 2011 | 221 |
| K111242 | Magnetom Aera and Magnetom Skyra | 2011 | 220 |
| K081028 | Signa 3.0T MR750 System | 2008 | 70 |
| K103429 | Biograph MMR | 2011 | 21 |
| K121160 | Magnetom Spectra Class I | 2012 | 1 |
| K130885 | Magnetom Avanto-Fit, Magnetom Skyra-Fit Class I | 2013 | 0 |
| K092925 | 3D ASL | 2010 | 0 |
Has anything in its family been recalled? (22 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 28, 2025 | Class I | Z-2664-2025 | Under Investigation by firm | There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within… |
| Sep 12, 2012 | Class II | Z-2464-2012 | Device Design | When positioning the Flex Large 4 Coil (part No. 8625761) off-center in left-right direction as it is used for example for hip and shoulder imaging and, in addition at an off-center position with respect to the head-foot direction, the… |
| Mar 26, 2012 | Not classified | Z-1364-2012 | Nonconforming Material/Component | During product monitoring, Siemens has discovered that the gradient cable connections on a few systems did not meet their specifications after installation. The cable connections overheated, leading to the emission of smoke and possible… |
| Jan 10, 2011 | Not classified | Z-1401-2011 | Other | On some systems the gradient cable connections did not meet firm's specifications after installation |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K101347Magnetom Aera; Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2010 · Cites it directly · ◆ in the 2010 descendants bar of FIG. 12 Class I recalls
- K130885Magnetom Avanto-Fit, Magnetom Skyra-FitSiemens Medical Solutions USA, Inc. · 2013 · Cites it directly · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
- K121160Magnetom SpectraSiemens Medical Solutions USA, Inc. · 2012 · Cites it directly · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
- K121434Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/VerioSiemens Medi Cal Solutions, Inc. · 2012 · 2 generations on · ◆ in the 2012 descendants bar of FIG. 12 Class I recalls
- K181613MAGNETOM Avanto, MAGNETOM Verio with software syngo MR E11DSiemens Medical Solutions USA, Inc. · 2018 · 3 generations on · ◆ in the 2018 descendants bar of FIG. 12 Class I recalls
- K132119Magnetom Prisma, Magnetom Prisma FitSiemens Medical Solutions USA, Inc. · 2013 · 3 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
- K153343MAGNETOM Aera, MAGNETOM Skyra, MAGNETOM Prisma, MAGNETOM Prisma fitSiemens Medi Cal Solutions, Inc. · 2016 · 4 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall
- K151579MAGNETOM Aera (24-channel), MAGNETOM Avanto fit, MAGNETOM Skyra fit, MAGNETOM Prisma, MAGNETOM Prisma fitSiemens Medical Solutions USA, Inc. · 2015 · 4 generations on · ◆ in the 2015 descendants bar of FIG. 12 Class I recalls
- K200213Biograph mMR with syngo MR E11P-AP01 system softwareSiemens Medical Solutions USA, Inc. · 2020 · 4 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall
- K170396MAGNETOM VidaSiemens Medical Solutions USA, Inc. · 2017 · 5 generations on · ◆ in the 2017 descendants bar of FIG. 11 Class I recall
- K173592MAGNETOM Aera, MAGNETOM Skyra/Skyrafit, MAGNETOM Prisma/Prismafit, MAGNETOM AvantofitSiemens Medical Solutions USA, Inc. · 2018 · 5 generations on · ◆ in the 2018 descendants bar of FIG. 12 Class I recalls
- K250443MAGNETOM Avanto Fit; MAGNETOM Skyra Fit; MAGNETOM Sola Fit; MAGNETOM Viato.MobileSiemens Healthcare GmbH · 2025 · 6 generations on · ◆ in the 2025 descendants bar of FIG. 12 Class I recalls
Nearest 12 of 21 shown.
Nearest 5 of 21; all are on the web page.
What does this lineage say?
- The line of descent spans 10 years and passes through 2 companies; a new generation was cleared every 1.1 years on median.
- 6 of the 9 ancestors have a recall of any class.
- Of its descendants, K082427 (Siemens Medical Solutions USA, Inc., 2008) has gone furthest, with 229 descendants of its own.
- 6 other devices also cite its predicate K050200.
How does its category compare?
All LNH devicesLNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.
FDA decisions per year · LNH
26in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LNH
122in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LNH
8in 2025
Class I (8 since 2007) Class II/III not classified
Adverse event reports per year · LNH
1,263in 2025
MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 35 | 25 | 114 |
| 2008 | 19 | 53 | 12 | 167 |
| 2009 | 17 | 70 | 7 | 194 |
| 2010 | 12 | 94 | 17 | 165 |
| 2011 | 16 | 81 | 11 | 182 |
| 2012 | 22 | 89 | 5 | 170 |
| 2013 | 19 | 108 | 2 | 107 |
| 2014 | 17 | 108 | 13 | 119 |
| 2015 | 19 | 140 | 12 (2) | 113 |
| 2016 | 20 | 108 | 1 | 135 |
| 2017 | 18 | 98 | 0 | 121 |
| 2018 | 29 | 128 | 71 | 158 |
| 2019 | 26 | 62 | 5 | 158 |
| 2020 | 32 | 121 | 4 | 98 |
| 2021 | 27 | 87 | 41 | 84 |
| 2022 | 30 | 74 | 13 | 138 |
| 2023 | 30 | 124 | 10 (1) | 159 |
| 2024 | 26 | 104 | 4 | 281 |
| 2025 | 26 | 122 | 8 (5) | 1,263 |
| 2026 | 24 | 126 | 43 | — |
What is it indicated for?
The MAGNETOM Verio is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Depending on the region of interest, contrast agents may be used. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
The MAGNETOM Verio may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in room display and MR-safe biopsy needles.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K072237 · printed Nov 4, 2026
Cite this page
510(k) Report. “K072237: Magnetom Verio.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K072237
Report an extraction error