Magnetom Avanto-Fit, Magnetom Skyra-Fit

K130885510(k) TraditionalCleared May 17, 201349-day reviewClass II21 CFR 892.1000Recalls 2 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on May 17, 2013 as substantially equivalent to K101347 (Siemens Medical Solutions USA, Inc., 2010) and 3 other predicates after a 49-day review. Its family tree reaches back 10 generations to 1997. 4 of the 33 devices in its family, counting itself, have a Class I recall. It has been recalled 2 times, including a Class I recall.

Decided
May 17, 2013
Review
49 days
Class
Class II
Product code
LNH
Recalls
2 (1 Class I)
Predicates
4

Its longest line of descent runs back 10 generations to K971684 (1997). 32 ancestors in all, the oldest K932271 (1994). No later device cites it as a predicate yet. 4 of the 33 devices in this family, counting this one, have a Class I recall.

  1. Root · 10 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
  2. 6 generations between
  3. 3 generations backK101347 Magnetom Aera; Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2010 · Recalled (Class I)
  4. 2 generations backK111242 Magnetom Aera and Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2011 · Recalled
  5. 1 generation backK121434 Software Syngo MR D13A for the Magnetom Systems…Siemens Medi Cal Solutions, Inc. · 2012 · Recalled (Class I)
  6. This deviceK130885 Magnetom Avanto-Fit, Magnetom Skyra-FitSiemens Medical Solutions USA, Inc. · 2013
Show all 10 generations
  1. Root · 10 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
  2. 9 generations backK032428 Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc. · 2003 · Recalled
  3. 8 generations backK041112 Magnetom Espree SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
  4. 7 generations backK052164 Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc. · 2005
  5. 6 generations backK062454 Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc. · 2006
  6. 5 generations backK072237 Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · Recalled (Class I)
  7. 4 generations backK082427 Syngo MR B17Siemens Medical Solutions USA, Inc. · 2008
  8. 3 generations backK101347 Magnetom Aera; Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2010 · Recalled (Class I)
  9. 2 generations backK111242 Magnetom Aera and Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2011 · Recalled
  10. 1 generation backK121434 Software Syngo MR D13A for the Magnetom Systems…Siemens Medi Cal Solutions, Inc. · 2012 · Recalled (Class I)
  11. This deviceK130885 Magnetom Avanto-Fit, Magnetom Skyra-FitSiemens Medical Solutions USA, Inc. · 2013

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K971684Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc.19974
−9K032428Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc.20034
−8K041112Magnetom Espree SystemSiemens Medical Solutions USA, Inc.20044
−7K052164Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc.2005—
−6K062454Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc.2006—
−5K072237Magnetom VerioSiemens Medical Solutions USA, Inc.2007Class I
−4K082427Syngo MR B17Siemens Medical Solutions USA, Inc.2008—
−3K101347Magnetom Aera; Magnetom SkyraSiemens Medical Solutions USA, Inc.2010Class I
−2K111242Magnetom Aera and Magnetom SkyraSiemens Medical Solutions USA, Inc.20113
−1K121434Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/VerioSiemens Medi Cal Solutions, Inc.2012Class I
0 (this device)K130885Magnetom Avanto-Fit, Magnetom Skyra-FitSiemens Medical Solutions USA, Inc.2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199810
199920
200020
200130
200220
200410
200530
200610
200710
200920
201010
201120
FIG. 1 — Lineage of K130885, 1997–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K101347Magnetom Aera; Magnetom Skyra
Summary page 2
…Clearance Preicae evie NmeNumber Date Siemens MAGNETOM Area* (1.5T) K101347 October 1, 2010 Siemens MAGNETOM Skyra* (3 T) K101 347 October 1,2010 *With synqo®D M…
Siemens Medical Solutions USA, Inc.2010Class I
*K121434Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/Verio
Summary page 3
…ms with syngo® MR VDI 3B software is substantially equivalent to the commercially available MAGNETOM Aera (1.5T) and MAGNETOM Skyra (3T) Systems with syngo® MR VD13A SW (K1 21434) Additionally, as specified in the FDA guidance document "Guidance for the Submission Of Premarket Notifications for Magnetic Re…
Siemens Medi Cal Solutions, Inc.2012Class I
*K082427Syngo MR B17
Summary page 2
…ith synqo®D MR VAlS3A SW Update K<121434 November 5, 2012 Siemens MAGNETOM Avanto(1.5T) & MAGNETOM Verio System with K082427 November 7, 2008 syngo® MR B17 update _______ Siemens MAGNETOM Verio (3 T) MR…
Siemens Medical Solutions USA, Inc.2008—
*K072237Magnetom Verio
Summary page 2
…th K082427 November 7, 2008 syngo® MR B17 update _______ Siemens MAGNETOM Verio (3 T) MR K<072237 October 10, 2007 System I_________________ Siemens 510(k) Premarket Notification…
Siemens Medical Solutions USA, Inc.2007Class I

Also cleared against K101347: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 28, 2025Class IZ-2661-2025Under Investigation by firmThere is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within…
Dec 7, 2015Class IIZ-0467-2016OtherPossible installation error of the quench lines of superconducting magnets, used to vent helium gas in the rare event of a quench. Any restriction, obstruction, or improper installation of the quench line may cause the helium gas to be…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K121434Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/VerioSiemens Medi Cal Solutions, Inc. · 2012 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
  • K101347Magnetom Aera; Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2010 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
  • K072237Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 16 years and passes through 2 companies; a new generation was cleared every 1.1 years on median.
  • 14 of the 32 ancestors have a recall of any class; 3 have a Class I recall.
  • 3 other devices also cite its predicate K101347.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

The MAGNETOM systems described above are indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents…

Read the full statement (123 words)

The MAGNETOM systems described above are indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities.

Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

The MAGNETOM systems described above may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-safe biopsy needles.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K130885: Magnetom Avanto-Fit, Magnetom Skyra-Fit.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K130885

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.