510(k) · Cleared
Recalled · 2010Easyra Cholesterol Reagent; Easyra Calcium Reagent Reagent
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Medica Corp. · CHH Enzymatic Esterase--Oxidase, Cholesterol
Cleared on Feb 20, 2008 after a 191-day review. The FDA summary names its predicate without a number. It has been recalled once.
- Decided
- Feb 20, 2008
- Review
- 191 days
- Class
- Class I
- Product code
- CHH
- Recalls
- 1
- Predicates
- 0
The FDA summary names its predicate without a number.
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in CHH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new CHH device
- #1 K251091 Lipidsscore 10.9 · 2 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | — | Cobas Ready Cholesterol Reagent Named without a submission number | |
| * | — | Cobas Ready Calcium Reagent Named without a submission number |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Mar 29, 2010 | Not classified | Z-0078-2011 | Other | Chemistry analyzers may report Calcium results below the sample's actual concentration. Reported results may be ten to fifteen percent lower than expected sample results. The defect is a shortened path length at specific cuvette positions… |
No ancestor or descendant has a Class I recall.
How does its category compare?
All CHH devicesCHH Enzymatic Esterase--Oxidase, Cholesterol: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.
FDA decisions per year · CHH
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · CHH
198in 2012
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · CHH
2in 2021
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · CHH
14in 2025
MAUDE reports naming product code CHH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 0 | 1 |
| 2008 | 2 | — | 0 | 0 |
| 2009 | 1 | — | 3 | 0 |
| 2010 | 0 | — | 2 | 10 |
| 2011 | 1 | — | 0 | 18 |
| 2012 | 4 | 198 | 0 | 11 |
| 2013 | 1 | — | 1 | 0 |
| 2014 | 1 | — | 1 | 1 |
| 2015 | 1 | — | 0 | 0 |
| 2016 | 1 | — | 2 | 5 |
| 2017 | 1 | — | 0 | 2 |
| 2018 | 2 | — | 0 | 1 |
| 2019 | 1 | — | 0 | 5 |
| 2020 | 0 | — | 0 | 1 |
| 2021 | 0 | — | 2 | 2 |
| 2022 | 1 | — | 0 | 4 |
| 2023 | 0 | — | 0 | 4 |
| 2024 | 1 | — | 0 | 8 |
| 2025 | 1 | — | 0 | 14 |
| 2026 | 0 | — | 1 | — |
What is it indicated for?
EasyRA Calcium EasyRA Calcium Reagent is intended for the quantitative determination of total calcium concentration in serum using the Medica EasyRA Chemistry Analyzer. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). For in-vitro diagnostic use only. For Professional use only. EasyRA Cholesterol The EasyRA…
Read the full statement (104 words)
EasyRA Calcium
EasyRA Calcium Reagent is intended for the quantitative determination of total calcium concentration in serum using the Medica EasyRA Chemistry Analyzer. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).
For in-vitro diagnostic use only. For Professional use only.
EasyRA Cholesterol
The EasyRA Cholesterol Reagent is intended for the quantitative determination of cholesterol in human serum on the Medica EasyRA Chemistry analyzer to screen for elevated cholesterol as a risk factor in coronary artery disease.
For in-vitro diagnostic use only. For Professional Use Only.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K072249 · printed Nov 6, 2026
Cite this page
510(k) Report. “K072249: Easyra Cholesterol Reagent; Easyra Calcium Reagent Reagent.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K072249
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026