510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Compass

K072374510(k) TraditionalCleared Dec 7, 2007106-day reviewClass II21 CFR 892.5050Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Scanditronix Wellhofer GmbH · IYE Accelerator, Linear, Medical

Cleared on Dec 7, 2007 as substantially equivalent to K062817 (Ptw-New York Corp., 2006) and 2 other predicates after a 106-day review. Its family tree reaches back 3 generations to 1992. 31 later devices descend from it. It has been recalled 3 times.

Decided
Dec 7, 2007
Review
106 days
Class
Class II
Product code
IYE
Recalls
3
Predicates
3

Its longest line of descent runs back 3 generations to K914698 (1992). 11 ancestors in all, the oldest K874893 (1988). 31 later devices descend from it, 4 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2003: 1 other ancestor2001: 1 other ancestor1997: 1 other ancestor1994: 1 other ancestor1995: 1 other ancestor1990: 1 other ancestor1988: 1 other ancestor2014: 2 descendants cleared2016: 5 descendants cleared2013: 1 descendant cleared2015: 1 descendant cleared2018: 2 descendants cleared2017: 3 descendants cleared2026: 3 descendants cleared2020: 1 descendant cleared2019: 2 descendants cleared2022: 3 descendants cleared2024: 2 descendants cleared2021: 1 descendant cleared2023: 2 descendants cleared2025: 3 descendants clearedK914698 · Setup Computer with Software V-1261,62 and 63 · K.S., Inc. · 1992K980904 · Mucheck Monitor Unit Validation Program · Medical Dosimetry Services, Inc. · 1998K012227 · Mucheck - Monitor Unit Validation Program, Model V4.1 · Oncology Data Systems, Inc. · 20011990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 3 other ancestors3 generations back: 3 other ancestors 1 generation forward: 4 descendants2 generations forward: 10 descendants3 generations forward: 7 descendants4 generations forward: 3 descendants5 generations forward: 3 descendants6 generations forward: 3 descendants7 generations forward: 1 descendant K012227 · Mucheck - Monitor Unit Validation Program, Model V4.1 · Oncology Data Systems, Inc. · 2001K980904 · Mucheck Monitor Unit Validation Program · Medical Dosimetry Services, Inc. · 1998K914698 · Setup Computer with Software V-1261,62 and 63 · K.S., Inc. · 19920+5
  1. Root · 3 generations backK914698 Setup Computer with Software V-1261,62 and 63K.S., Inc. · 1992
  2. 2 generations backK980904 Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc. · 1998
  3. 1 generation backK012227 Mucheck - Monitor Unit Validation Program, Model V4.1Oncology Data Systems, Inc. · 2001
  4. This deviceK072374 CompassScanditronix Wellhofer GmbH · 2007
  5. Descendant · cites it directly · 23 of its ownK140660 MOBIUS3DMobius Medical Systems, LP · 2014 · Recalled
  6. Descendant · 2 generations on · 13 of its ownK153014 Mobius3DMobius Medical Systems, LP · 2016
  7. Descendant · 3 generations on · 5 of its ownK191761 Mobius3DVarian Medical Systems, Inc. · 2019
  8. Descendant · cites it directly · 5 of its ownK141800 PerfractionSun Nuclear Corporation · 2014
  9. Descendant · 4 generations on · 4 of its ownK192424 Mobius3DVarian Medical Systems, Inc. · 2019

Showing 5 of 31 descendants. See all descendants

Descendants cleared per year

2013: 12014: 22015: 12016: 52017: 32018: 22019: 22020: 12021: 12022: 32023: 22024: 22025: 32026: 3
2013high 5 (2016) · low 1 (2013)2026

Most influential descendants

Showing 5 of 31 · See all
  • K140660MOBIUS3DMobius Medical Systems, LP · 2014 · cites it directly23descendants
  • K153014Mobius3DMobius Medical Systems, LP · 2016 · 2 generations on13descendants
  • K191761Mobius3DVarian Medical Systems, Inc. · 2019 · 3 generations on5descendants
  • K141800PerfractionSun Nuclear Corporation · 2014 · cites it directly5descendants
  • K192424Mobius3DVarian Medical Systems, Inc. · 2019 · 4 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K914698Setup Computer with Software V-1261,62 and 63K.S., Inc.1992—
−2K980904Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc.1998—
−1K012227Mucheck - Monitor Unit Validation Program, Model V4.1Oncology Data Systems, Inc.2001—
0 (this device)K072374CompassScanditronix Wellhofer GmbH2007—
+1K140660MOBIUS3DMobius Medical Systems, LP20141
+2K153014Mobius3DMobius Medical Systems, LP2016—
+3K191761Mobius3DVarian Medical Systems, Inc.2019—
+1K141800PerfractionSun Nuclear Corporation2014—
+4K192424Mobius3DVarian Medical Systems, Inc.2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
199010
199410
199510
199710
200110
200310
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201420
201510
201650
201730
201820
201920
202010
202110
202230
202320
202420
202530
202630
FIG. 1 — Lineage of K072374, 1992–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K062817PTW David Imrt QC Verification System
Summary page 1
…e Accelerator, linear, medical 5.6 Classification name Medical charged-particle radiation therapy system (21 CFR 892.5050, Product Code lYE) 5.7 Predicate devices K062817 David (PTW), K031634 OmnniPro I'mnRT (Scanditronix Wellhdfer), K012227 muCheck (Oncology Data Systems, Inc.), Delta4 (Scandidos…
Ptw-New York Corp.2006—
*K031634Omnipro I'mrt
Summary page 1
…r, medical 5.6 Classification name Medical charged-particle radiation therapy system (21 CFR 892.5050, Product Code lYE) 5.7 Predicate devices K062817 David (PTW), K031634 OmnniPro I'mnRT (Scanditronix Wellhdfer), K012227 muCheck (Oncology Data Systems, Inc.), Delta4 (Scandidos AB) 5.8 Device desc…
Wellhofer Dosimetrie2003—
*K012227Mucheck - Monitor Unit Validation Program, Model V4.1
Summary page 1
…ed-particle radiation therapy system (21 CFR 892.5050, Product Code lYE) 5.7 Predicate devices K062817 David (PTW), K031634 OmnniPro I'mnRT (Scanditronix Wellhdfer), K012227 muCheck (Oncology Data Systems, Inc.), Delta4 (Scandidos AB) 5.8 Device description The system is intended to be used with int…
Oncology Data Systems, Inc.2001—
*—Delta4 (Scandidos AB)
Named without a submission number
—

Also cleared against K062817: 2 other devices →

What descends from it? (31)

See all 31

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K140660MOBIUS3DMobius Medical Systems, LP201423
K141800PerfractionSun Nuclear Corporation20145
K151426ScandiDos Delta4 Discover, Alternative Trade Name: ScandiDos Delta4 Discover+Scandidos AB20161
K1318623DVHSun Nuclear Corporation20130

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 2, 2015Class IIZ-1212-2015Software design (manufacturing process)Error in the software. During internal tests of the current development version of the Compass SW it was found that dose reconstruction for DMLC plans when the jaws move or mlc leaves reverse during beam on will not be scaled correctly…
Sep 25, 2014Class IIZ-0168-2015Software designError in the software. A deviation between reconstructed and planned dose distribution may not be detected prior to treatment and this can result in an over or under-estimation of the pretreatment delivered dose.
Aug 20, 2013Class IIZ-0456-2014Software designThe development partner of IBA informed that there is problem in a SW component used in COMPASS when converting Region of Interest (ROI) geometries to dose grid representation. Under certain conditions this problem generates too large ROI…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 4 companies; a new generation was cleared every 6.0 years on median.
  • None of the 11 ancestors has a recall.
  • Of its descendants, K140660 (Mobius Medical Systems, LP, 2014) has gone furthest, with 23 descendants of its own.
  • 2 other devices also cite its predicate K062817.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

COMPASS is used for quality assurance and plan verification in external beam radiation therapy for multileaf collimated fields and intensity modulated fields.

It computes dose (energy per volume deposited by ionizing radiation) three-dimensionally in a geometrical representation of a patient or a phantom. The calculation is based on read-in treatment plans, and additionally on online or offline measurements of radiation fields with radiation-transparent or non-transparent, 2 dimensional multi-element ionization chamber arrays.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K072374: Compass.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K072374

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026