510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Titanium Powerport Isp Implanted Port with 6 FR Chronoflex Polyurethane Catheter

K072549510(k) SpecialCleared Nov 14, 200765-day reviewClass II21 CFR 880.5965Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

C.R. Bard, Inc. · LJT Port & Catheter, Implanted, Subcutaneous, Intravascular

Cleared on Nov 14, 2007 as substantially equivalent to K060812 (C.R. Bard, Inc., 2006) and 1 other predicate after a 65-day review. Its family tree reaches back 8 generations to 2000. 4 later devices descend from it. It has been recalled 2 times.

Decided
Nov 14, 2007
Review
65 days
Class
Class II
Product code
LJT
Recalls
2
Predicates
2

Its longest line of descent runs back 8 generations to K001901 (2000). 21 ancestors in all, the oldest K863606 (1986). 4 later devices descend from it, 3 citing it directly.

  1. Root · 8 generations backK001901 Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath…C.R. Bard, Inc. · 2000 · Recalled
  2. 7 generations backK012902 Poly Per-Q-Cath Single Lumen Picc Catheter, Model 3143100; Poly…C.R. Bard, Inc. · 2001
  3. 6 generations backK031129 Poly Per-Q-Cath Picc Catheters; Poly RadpiccC.R. Bard, Inc. · 2003
  4. 5 generations backK034019 Poly Per-Q-Cath Picc CathetersC.R. Bard, Inc. · 2004 · Recalled
  5. 4 generations backK050931 6 FR DL Powerpicc CatheterC.R. Bard, Inc. · 2005
  6. 3 generations backK051672 PowerpiccBard Access Systems, Inc. · 2005 · Recalled
  7. 2 generations backK060812 Powerport (TM)Implanted Titanium Port W/8 FR. Chronoflex…C.R. Bard, Inc. · 2006 · Recalled
  8. 1 generation backK063377 Powerport Polymetric Port with 8 FR S/L Chronoflex CatheterC.R. Bard, Inc. · 2007 · Recalled
  9. This deviceK072549 Titanium Powerport Isp Implanted Port with 6 FR Chronoflex…C.R. Bard, Inc. · 2007
  10. Descendant · cites it directly · 1 of its ownK251253 PowerPort™ isp Implantable Port; PowerPort™ Slim Implantable Port…Bard Access Systems, Inc. · 2025
  11. Descendant · 2 generations on · 0 of its ownK252478 Vaccess™ CT Low-Profile Power-Injectable Implantable Port; Vaccess™…Bard Access Systems, Inc. · 2025
  12. Descendant · cites it directly · 0 of its ownK232737 PowerPort™ ClearVUE™ Slim ECG Enabled Implantable Port; PowerPort™…Bard Peripheral Vascular, Inc. · 2023
  13. Descendant · cites it directly · 0 of its ownK181446 Bard Power-Injectable Implantable Ports (PowerPorts®)C.R. Bard, Inc. · 2019

Most influential descendants

Showing 4 of 4 · See all
  • K251253PowerPort™ isp Implantable Port; PowerPort™ Slim Implantable Port; PowerPort™ M.R.I. ™…Bard Access Systems, Inc. · 2025 · cites it directly1descendant
  • K252478Vaccess™ CT Low-Profile Power-Injectable Implantable Port; Vaccess™ CT Power-Injectable…Bard Access Systems, Inc. · 2025 · 2 generations on0descendants
  • K232737PowerPort™ ClearVUE™ Slim ECG Enabled Implantable Port; PowerPort™ ClearVUE™ isp ECG…Bard Peripheral Vascular, Inc. · 2023 · cites it directly0descendants
  • K181446Bard Power-Injectable Implantable Ports (PowerPorts®)C.R. Bard, Inc. · 2019 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K001901Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath Midline…C.R. Bard, Inc.20002
−7K012902Poly Per-Q-Cath Single Lumen Picc Catheter, Model 3143100; Poly Per-Q-Cath…C.R. Bard, Inc.2001—
−6K031129Poly Per-Q-Cath Picc Catheters; Poly RadpiccC.R. Bard, Inc.2003—
−5K034019Poly Per-Q-Cath Picc CathetersC.R. Bard, Inc.20041
−4K0509316 FR DL Powerpicc CatheterC.R. Bard, Inc.2005—
−3K051672PowerpiccBard Access Systems, Inc.20051
−2K060812Powerport (TM)Implanted Titanium Port W/8 FR. Chronoflex Catheter-MultipleC.R. Bard, Inc.20061
−1K063377Powerport Polymetric Port with 8 FR S/L Chronoflex CatheterC.R. Bard, Inc.20077
0 (this device)K072549Titanium Powerport Isp Implanted Port with 6 FR Chronoflex Polyurethane CatheterC.R. Bard, Inc.2007—
+1K251253PowerPort™ isp Implantable Port; PowerPort™ Slim Implantable Port; PowerPort™…Bard Access Systems, Inc.2025—
+2K252478Vaccess™ CT Low-Profile Power-Injectable Implantable Port; Vaccess™ CT…Bard Access Systems, Inc.2025—
+1K232737PowerPort™ ClearVUE™ Slim ECG Enabled Implantable Port; PowerPort™ ClearVUE™…Bard Peripheral Vascular, Inc.2023—
+1K181446Bard Power-Injectable Implantable Ports (PowerPorts®)C.R. Bard, Inc.2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198710
198940
199110
200010
200110
200420
200520

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202310
202520
FIG. 1 — Lineage of K072549, 2000–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LJT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LJT device

  1. #1 K251253 PowerPort™ isp Implantable Port; PowerPort™…score 14.9 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K060812Powerport (TM)Implanted Titanium Port W/8 FR. Chronoflex Catheter-Multiple
Summary page 1
…80 LJT - Subcutaneous, Implanted, Intravascular Infusion Port & Catheter Premarket Notification: K060812, clearance date July 14, 2006 Predicate Devices Trade Name: MRI PowerPortTM…
C.R. Bard, Inc.20061
*K063377Powerport Polymetric Port with 8 FR S/L Chronoflex Catheter
Summary page 1
…aneous, Implanted, Intravascular Infusion Port & Catheter Premarket Notification: K063377, clearance date Jan 25, 2007 Class II Classification 21 CFR §880.5965…
C.R. Bard, Inc.20077

Also cleared against K060812: 18 other devices →

What descends from it? (4)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K251253PowerPort™ isp Implantable Port; PowerPort™ Slim Implantable Port; PowerPort™ M.R.I. ™…Bard Access Systems, Inc.20251
K232737PowerPort™ ClearVUE™ Slim ECG Enabled Implantable Port; PowerPort™ ClearVUE™ isp ECG…Bard Peripheral Vascular, Inc.20230
K181446Bard Power-Injectable Implantable Ports (PowerPorts®)C.R. Bard, Inc.20190

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 1, 2019Class IIZ-1573-2020Component change controlFirm has identified that the product may be at risk of incorrectly containing a tunneler with a barb tip meant to attach to a 6Fr catheter instead of the correct barb tip for a 9.6Fr catheter. This can result in a prolongation of operation.
Mar 18, 2010Not classifiedZ-1525-2010Mixed-up of materials/components8F PowerPort stems were mixed with 6F PowerPorts and packaged.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 7 years and passes through 2 companies; a new generation was cleared every 10 months on median.
  • 12 of the 21 ancestors have a recall of any class.
  • Of its descendants, K251253 (Bard Access Systems, Inc., 2025) has gone furthest, with 1 descendant of its own.
  • 18 other devices also cite its predicate K060812.

How does its category compare?

All LJT devices

LJT Port & Catheter, Implanted, Subcutaneous, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · LJT

2in 2025

2007: 112008: 72009: 22010: 22011: 52012: 22013: 72014: 22015: 42016: 32019: 12020: 12023: 12024: 12025: 22026 (partial year): 1
2007high 11 (2007) · low 0 (2017)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LJT

182in 2016

2007: 85 days median2008: 119 days median2011: 144 days median2013: 150 days median2015: 164 days median2016: 182 days median
2007high 182 (2016) · low 85 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LJT

1in 2022

2007: 12008: 22009: 22010: 192011: 12012: 22015: 12017: 12018: 12019: 72022: 12026 (partial year): 1, 1 Class I
2007high 19 (2010) · low 0 (2013)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · LJT

4,308in 2025

2007: 2082008: 1612009: 1722010: 2072011: 2022012: 1532013: 1412014: 1432015: 1492016: 1352017: 2622018: 8782019: 7652020: 9592021: 1,2852022: 1,3342023: 1,2212024: 1,8962025: 4,308
2007high 4,308 (2025) · low 135 (2016)2025

MAUDE reports naming product code LJT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711851208
200871192161
20092—2172
20102—19207
201151441202
20122—2153
201371500141
20142—0143
201541641149
201631820135
20170—1262
20180—1878
20191—7765
20201—0959
20210—01,285
20220—11,334
20231—01,221
20241—01,896
20252—04,308
20261—1 (1)—

What is it indicated for?

The PowerPortTM implanted port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medication, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. When used with a PowerLoc TM safety infusion set, the PowerPortTM device is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K072549: Titanium Powerport Isp Implanted Port with 6 FR Chronoflex Polyurethane Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K072549

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026