510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Cxi Support Catheter

K072724510(k) SpecialCleared Nov 9, 200744-day reviewClass II21 CFR 870.1210Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cook, Inc. · KRA Catheter, Continuous Flush

Cleared on Nov 9, 2007 as substantially equivalent to K052841 (Cook, Inc., 2005) and 1 other predicate after a 44-day review. Its family tree reaches back 1 generation to 2005. 1 of the 16 devices in its family, counting itself, has a Class I recall. 13 later devices descend from it. It has been recalled once.

Decided
Nov 9, 2007
Review
44 days
Class
Class II
Product code
KRA
Recalls
1
Predicates
2

Its longest line of descent runs back 1 generation to K052841 (2005). 2 ancestors in all. 13 later devices descend from it, 4 citing it directly. 1 of the 16 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2009: 1 descendant cleared2012: 1 descendant cleared2011: 1 descendant cleared2016: 3 descendants cleared, 1 with a Class I recall12019: 1 descendant cleared2017: 1 descendant cleared2022: 2 descendants cleared2020: 1 descendant cleared2021: 1 descendant cleared2026: 1 descendant clearedK052841 · Mira-Flex 18 Microcatheter · Cook, Inc. · 200520052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 4 descendants2 generations forward: 6 descendants, 1 with a Class I recall13 generations forward: 1 descendant4 generations forward: 2 descendants K052841 · Mira-Flex 18 Microcatheter · Cook, Inc. · 20050
  1. Root · 1 generation backK052841 Mira-Flex 18 MicrocatheterCook, Inc. · 2005
  2. This deviceK072724 Cxi Support CatheterCook, Inc. · 2007
  3. Descendant · cites it directly · 7 of its ownK093052 Crosscath Support CathterCook, Inc. · 2009
  4. Descendant · 2 generations on · 3 of its ownK191731 Instylla MicrocatheterInstylla, Inc. · 2019
  5. Descendant · 3 generations on · 2 of its ownK200744 Instylla MicrocatheterInstylla, Inc. · 2020
  6. Descendant · cites it directly · 2 of its ownK160884 CXI Support CatheterCook Incorporated · 2016
  7. Descendant · 4 generations on · 1 of its ownK210808 Instylla Microcatheter 1.2Instylla, Inc. · 2021

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2009: 12011: 12012: 12016: 32017: 12019: 12020: 12021: 12022: 22026: 1
2009high 3 (2016) · low 0 (2010)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 13 · See all
  • K093052Crosscath Support CathterCook, Inc. · 2009 · cites it directly7descendants
  • K191731Instylla MicrocatheterInstylla, Inc. · 2019 · 2 generations on3descendants
  • K200744Instylla MicrocatheterInstylla, Inc. · 2020 · 3 generations on2descendants
  • K160884CXI Support CatheterCook Incorporated · 2016 · cites it directly2descendants
  • K210808Instylla Microcatheter 1.2Instylla, Inc. · 2021 · 4 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K052841Mira-Flex 18 MicrocatheterCook, Inc.2005—
0 (this device)K072724Cxi Support CatheterCook, Inc.2007—
+1K093052Crosscath Support CathterCook, Inc.2009—
+2K191731Instylla MicrocatheterInstylla, Inc.2019—
+3K200744Instylla MicrocatheterInstylla, Inc.2020—
+1K160884CXI Support CatheterCook Incorporated2016—
+4K210808Instylla Microcatheter 1.2Instylla, Inc.2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201110
201210
201631
201710
201910
202010
202110
202220
202610
FIG. 1 — Lineage of K072724, 2005–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KRA device

  1. #1 K253509 TriSalus TriNav Advance Infusion Systemscore 14.4 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K052841Mira-Flex 18 Microcatheter
Summary page 1
…lective anatomy for diagnostics and interventional procedures. The device most similar is the MiraFlex 18 Microcatheter cleared for marketing on December 8, 2005 (D.C.#K052841) and March 6, 2006 (D.C.#K060224). The similar indications for use, principles of operation, technological characteristics and…
Cook, Inc.2005—
*K060224Modification to Mira-Flex 18 Microcatheter
Summary page 1
…nd interventional procedures. The device most similar is the MiraFlex 18 Microcatheter cleared for marketing on December 8, 2005 (D.C.#K052841) and March 6, 2006 (D.C.#K060224). The similar indications for use, principles of operation, technological characteristics and performance testing results fo…
Cook Incorporated2006—

What descends from it? (13)

See all 13

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K093052Crosscath Support CathterCook, Inc.20097
K160884CXI Support CatheterCook Incorporated20162
K122796Cxi Support CatheterCook, Inc.20121
K111263Cxi TriforceCook Incorporated20111

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 21, 2015Class IIZ-0244-2016Labeling mix-upsCook Medical has received six product complaints associated with reports of the 2.6FR curved catheter being mixed with the 4.0FR straight catheter and vice versa. Investigation revealed the products were mixed during processing.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K161801CrossCath Support CatheterCook Incorporated · 2016 · 2 generations on · ◆ in the 2016 descendants bar of FIG. 14 Class I recalls

What does this lineage say?

  • None of the 2 ancestors has a recall.
  • Of its descendants, K093052 (Cook, Inc., 2009) has gone furthest, with 7 descendants of its own.

How does its category compare?

All KRA devices

KRA Catheter, Continuous Flush: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · KRA

1in 2025

2007: 52008: 72009: 52010: 22011: 22012: 32013: 32014: 42016: 82017: 42018: 42019: 42020: 62021: 62022: 62023: 72025: 12026 (partial year): 5
2007high 8 (2016) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KRA

63in 2023

2007: 44 days median2008: 144 days median2009: 69 days median2012: 29 days median2013: 126 days median2014: 61 days median2016: 122 days median2017: 166 days median2018: 178 days median2019: 133 days median2020: 29 days median2021: 108 days median2022: 228 days median2023: 63 days median2026 (partial year): 268 days median
2007high 268 (2026) · low 29 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KRA

41so far in 2026, the most since 2007

2007: 22008: 62009: 82010: 262011: 32012: 32013: 22014: 3, 2 Class I2015: 12016: 3, 1 Class I2019: 4, 4 Class I2020: 12023: 22026 (partial year): 41
2007high 41 (2026) · low 0 (2017)2026

Class I (7 since 2007) Class II/III not classified

Adverse event reports per year · KRA

980in 2025

2007: 1182008: 1172009: 1872010: 3552011: 2392012: 2172013: 2092014: 2762015: 3392016: 4182017: 5382018: 4342019: 4182020: 3812021: 4402022: 5432023: 8292024: 9422025: 980
2007high 980 (2025) · low 117 (2008)2025

MAUDE reports naming product code KRA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075442118
200871446117
20095698187
20102—26355
20112—3239
20123293217
201331262209
20144613 (2)276
20150—1339
201681223 (1)418
201741660538
201841780434
201941334 (4)418
20206291381
202161080440
202262280543
20237632829
20240—0942
20251—0980
2026526841—

What is it indicated for?

Intended for use in small vessel or superselective anatomy for diagnostic and interventional procedures including peripheral use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K072724: Cxi Support Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K072724

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026