510(k) Report

510(k) · Cleared

Recalled · Class I · 2019

CrossCath Support Catheter

K161801510(k) SpecialCleared Jul 29, 201629-day reviewClass II21 CFR 870.1210Recalls 4 (4 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cook Incorporated · KRA Catheter, Continuous Flush

Cleared on Jul 29, 2016 as substantially equivalent to K093052 (Cook, Inc., 2009) after a 29-day review. Its family tree reaches back 3 generations to 2005. 1 of the 5 devices in its family, counting itself, has a Class I recall. It has been recalled 4 times, including 4 Class I recalls.

Decided
Jul 29, 2016
Review
29 days
Class
Class II
Product code
KRA
Recalls
4 (4 Class I)
Predicates
1

Its longest line of descent runs back 3 generations to K052841 (2005). 4 ancestors in all. No later device cites it as a predicate yet. 1 of the 5 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year cleared 2006: 1 other ancestorK052841 · Mira-Flex 18 Microcatheter · Cook, Inc. · 2005K072724 · Cxi Support Catheter · Cook, Inc. · 2007 · recalledK093052 · Crosscath Support Cathter · Cook, Inc. · 20092006200820102012201420162018
  1. Root · 3 generations backK052841 Mira-Flex 18 MicrocatheterCook, Inc. · 2005
  2. 2 generations backK072724 Cxi Support CatheterCook, Inc. · 2007 · Recalled
  3. 1 generation backK093052 Crosscath Support CathterCook, Inc. · 2009
  4. This deviceK161801 CrossCath Support CatheterCook Incorporated · 2016
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K052841Mira-Flex 18 MicrocatheterCook, Inc.2005—
−2K072724Cxi Support CatheterCook, Inc.20071
−1K093052Crosscath Support CathterCook, Inc.2009—
0 (this device)K161801CrossCath Support CatheterCook Incorporated2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200610
FIG. 1 — Lineage of K161801, 2005–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KRA device

  1. #1 K253509 TriSalus TriNav Advance Infusion Systemscore 14.5 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K093052Crosscath Support Cathter
Summary page 4
…edicate Device: The device, subject of this submission, is substantially equivalent to the predicate device, the CrossCath Support Catheters cleared under 510(k) number K093052. Comparison to Predicate Device: It has been demonstrated that the CrossCath Support Catheters are comparable to the predicate d…
Cook, Inc.2009—

Also cleared against K093052: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 8, 2019Class IZ-0588-2020Process controlA manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with…
Nov 8, 2019Class IZ-0589-2020Process controlA manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with…
Nov 8, 2019Class IZ-0587-2020Process controlA manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with…
Nov 8, 2019Class IZ-0590-2020Process controlA manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 2 companies; a new generation was cleared every 2.1 years on median.
  • 1 of the 4 ancestors has a recall of any class.
  • 2 other devices also cite its predicate K093052.

How does its category compare?

All KRA devices

KRA Catheter, Continuous Flush: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · KRA

1in 2025

2007: 52008: 72009: 52010: 22011: 22012: 32013: 32014: 42016: 82017: 42018: 42019: 42020: 62021: 62022: 62023: 72025: 12026 (partial year): 5
2007high 8 (2016) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KRA

63in 2023

2007: 44 days median2008: 144 days median2009: 69 days median2012: 29 days median2013: 126 days median2014: 61 days median2016: 122 days median2017: 166 days median2018: 178 days median2019: 133 days median2020: 29 days median2021: 108 days median2022: 228 days median2023: 63 days median2026 (partial year): 268 days median
2007high 268 (2026) · low 29 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KRA

41so far in 2026, the most since 2007

2007: 22008: 62009: 82010: 262011: 32012: 32013: 22014: 3, 2 Class I2015: 12016: 3, 1 Class I2019: 4, 4 Class I2020: 12023: 22026 (partial year): 41
2007high 41 (2026) · low 0 (2017)2026

Class I (7 since 2007) Class II/III not classified

Adverse event reports per year · KRA

980in 2025

2007: 1182008: 1172009: 1872010: 3552011: 2392012: 2172013: 2092014: 2762015: 3392016: 4182017: 5382018: 4342019: 4182020: 3812021: 4402022: 5432023: 8292024: 9422025: 980
2007high 980 (2025) · low 117 (2008)2025

MAUDE reports naming product code KRA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075442118
200871446117
20095698187
20102—26355
20112—3239
20123293217
201331262209
20144613 (2)276
20150—1339
201681223 (1)418
201741660538
201841780434
201941334 (4)418
20206291381
202161080440
202262280543
20237632829
20240—0942
20251—0980
2026526841—

What is it indicated for?

The CrossCath Support Catheter is intended for use in small vessel or superselective anatomy for diagnostic and interventional procedures, including peripheral use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K161801: CrossCath Support Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K161801

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026