510(k) · Cleared
Recalled · Class II · 2015Anterolateral and Medial Locking Plating System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
DePuy Orthopaedics, Inc. · HRS Plate, Fixation, Bone
Cleared on Dec 14, 2007 as substantially equivalent to K990120 (Depuy Ace Medical Co., 1999) and 3 other predicates after a 72-day review. Its family tree reaches back 1 generation to 1988. 58 later devices descend from it. It has been recalled 2 times.
- Decided
- Dec 14, 2007
- Review
- 72 days
- Class
- Class II
- Product code
- HRS
- Recalls
- 2
- Predicates
- 4
Its longest line of descent runs back 1 generation to K882831 (1988). 4 ancestors in all. 58 later devices descend from it, 2 citing it directly.
- Root · 1 generation backK882831 Ace Cancellous Bone ScrewBuckman Co., Inc. · 1988
- This deviceK072832 Anterolateral and Medial Locking Plating SystemDePuy Orthopaedics, Inc. · 2007
- Descendant · cites it directly · 55 of its ownK092812 Synthes (Usa) 2.7MM/ 3.5MM LCP Anterolateral Distal Tibia PlatesSynthes (Usa) · 2010
- Descendant · 2 generations on · 19 of its ownK162674 BioPro Foot Plating SystemsBiopro, Inc. · 2017
- Descendant · 2 generations on · 15 of its ownK153716 GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating…Gpc Medical, Ltd. · 2016
- Descendant · 3 generations on · 14 of its ownK163361 ANTHEM™ Fracture SystemGlobus Medical, Inc. · 2017
- Descendant · 2 generations on · 11 of its ownK193633 APTUS® Ankle Trauma System 2.8/3.5Medartis AG · 2020
Showing 5 of 58 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 58 · See all- K092812Synthes (Usa) 2.7MM/ 3.5MM LCP Anterolateral Distal Tibia PlatesSynthes (Usa) · 2010 · cites it directly55descendants
- K162674BioPro Foot Plating SystemsBiopro, Inc. · 2017 · 2 generations on19descendants
- K153716GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System.Gpc Medical, Ltd. · 2016 · 2 generations on15descendants
- K163361ANTHEM™ Fracture SystemGlobus Medical, Inc. · 2017 · 3 generations on14descendants
- K193633APTUS® Ankle Trauma System 2.8/3.5Medartis AG · 2020 · 2 generations on11descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K882831 | Ace Cancellous Bone Screw | Buckman Co., Inc. | 1988 | — |
| 0 (this device) | K072832 | Anterolateral and Medial Locking Plating System | DePuy Orthopaedics, Inc. | 2007 | — |
| +1 | K092812 | Synthes (Usa) 2.7MM/ 3.5MM LCP Anterolateral Distal Tibia Plates | Synthes (Usa) | 2010 | — |
| +2 | K162674 | BioPro Foot Plating Systems | Biopro, Inc. | 2017 | — |
| +2 | K153716 | GPC BRAND Locking Bone Plates and Screws Osteosynthesis Plating System. | Gpc Medical, Ltd. | 2016 | — |
| +3 | K163361 | ANTHEM™ Fracture System | Globus Medical, Inc. | 2017 | — |
| +2 | K193633 | APTUS® Ankle Trauma System 2.8/3.5 | Medartis AG | 2020 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1991 | 1 | 0 |
| 1998 | 1 | 0 |
| 1999 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2010 | 1 | 0 |
| 2012 | 1 | 0 |
| 2014 | 1 | 0 |
| 2015 | 1 | 0 |
| 2016 | 2 | 0 |
| 2017 | 4 | 0 |
| 2018 | 2 | 0 |
| 2019 | 2 | 0 |
| 2020 | 4 | 0 |
| 2021 | 3 | 0 |
| 2022 | 2 | 0 |
| 2023 | 10 | 0 |
| 2024 | 10 | 0 |
| 2025 | 7 | 0 |
| 2026 | 8 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in HRS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new HRS device
- #1 K222465 Tyber Medical Anatomical Plating Systemscore 27.5 · 16 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (4)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K990120 | Modification of Depuy Ace Timax Medial Pilon PlateSummary page 1…e fixation appliances Name: and accessories (21 CFR § 888.3030) Predicate Devices: DePuy K990120 TiMax Medial Pilon Plate DePuy K983853 TiMax Meta Plate… | 1999 |
| * | K983853 | Depuy Ace Timax Meta PlateSummary page 1…te Devices: DePuy K990120 TiMax Medial Pilon Plate DePuy K983853 TiMax Meta Plate De~~uy K905048 3.5mm Cortical Bone… | 1998 |
| * | K905048 | Ace 3.5MM Cortical Bone ScrewSummary page 1…0 TiMax Medial Pilon Plate DePuy K983853 TiMax Meta Plate De~~uy K905048 3.5mm Cortical Bone De~uy K905048 Screws… | 1991 |
| * | K882831 | Ace Cancellous Bone Screw Our reading of the summary gave K882381, FDA’s number for Pti Heart-Lift (TM) Cardiac Retraction Gloves-Steri (Pioneering Technologies, Inc.). Corrected to the number with two digits swapped whose FDA name carries the summary’s words for the device, “Cancellous Bone Screws”. | 1988 |
What descends from it? (58)
See all 582 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Nov 20, 2015 | Class II | Z-0450-2016 | Process change control | These screws have a single lead (instead of a triple lead) thread. The screw may sit proud in the plate, If the screw does not lock into the plate; the screw may back out. If the screw sits proud, soft tissue irritation may occur. If the… |
| Dec 19, 2013 | Class II | Z-0716-2014 | Labeling mix-ups | 3.5mmx38mm Locking Cortical Screw, implants are being recalled since an investigation identified that a 3.5mmx28mm Cortical Locking Screw is labeled as a 3.5mmx38mm Cortical Locking Screw. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All HRS devicesHRS Plate, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.
FDA decisions per year · HRS
54in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · HRS
90in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · HRS
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · HRS
2,306in 2025
MAUDE reports naming product code HRS, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 35 | 63 | 2 | 339 |
| 2008 | 35 | 87 | 11 | 311 |
| 2009 | 47 | 85 | 3 | 435 |
| 2010 | 49 | 100 | 16 | 491 |
| 2011 | 43 | 88 | 5 | 773 |
| 2012 | 59 | 106 | 4 | 1,663 |
| 2013 | 56 | 89 | 11 | 2,643 |
| 2014 | 76 | 57 | 0 | 2,413 |
| 2015 | 56 | 97 | 7 | 3,095 |
| 2016 | 44 | 109 | 5 | 2,276 |
| 2017 | 57 | 122 | 3 | 2,220 |
| 2018 | 40 | 116 | 2 | 2,663 |
| 2019 | 46 | 89 | 25 | 3,359 |
| 2020 | 48 | 88 | 1 | 2,790 |
| 2021 | 40 | 87 | 9 | 2,442 |
| 2022 | 47 | 144 | 3 | 1,700 |
| 2023 | 40 | 120 | 1 | 2,049 |
| 2024 | 54 | 96 | 0 | 1,849 |
| 2025 | 54 | 90 | 1 | 2,306 |
| 2026 | 51 | 82 | 3 | — |
What is it indicated for?
The Medial Locking Plate is indicated for:
-Pilon Fractures: distal tibial intra-articular fractures
-High medial malleolar fractures
-Low boot type rotational distal extra-articular shaft fractures
while the Anterolateral Locking Plate is indicated for:
-Distal intra-articular tibia fractures
-Proximal tibia fractures
-Proximal and distal humerus fractures
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K072832 · printed Nov 6, 2026
Cite this page
510(k) Report. “K072832: Anterolateral and Medial Locking Plating System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K072832
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026