Recalled · Class II · 2013Model Ssa-680A Xario XG Diagnostic Ultrasound System, Version 1.00
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Toshibamedical Systems Corporation · ITX Transducer, Ultrasonic, Diagnostic
Cleared on Oct 29, 2007 as substantially equivalent to K072000 (Toshiba America Medical Systems, In.C, 2007) after a 14-day review. Its family tree reaches back 9 generations to 1989. It has been recalled 3 times.
- Decided
- Oct 29, 2007
- Review
- 14 days
- Class
- Class II
- Product code
- ITX
- Recalls
- 3
- Predicates
- 1
Its longest line of descent runs back 9 generations to K894632 (1989). 50 ancestors in all. No later device cites it as a predicate yet.
- Root · 9 generations backK894632 Model SSH-140A Ultrasound Imaging SystemToshiba America Medical Systems, In.C · 1989
- 5 generations between
- 3 generations backK032281 Ssa-770A - Version 4, Aplio Diagnostic Ultrasound SystemToshiba America Medical Systems, In.C · 2003
- 2 generations backK063130 Aplio XG Diagnostic Ultrasound System, Model Ssa-790A, Version 1.0Toshibamedical Systems Corporation · 2006 · Recalled
- 1 generation backK072000 Aplio XG Diagnostic Ultrasound System, Model Ssa-790A Version 2.0Toshiba America Medical Systems, In.C · 2007
- This deviceK072918 Model Ssa-680A Xario XG Diagnostic Ultrasound System, Version 1.00Toshibamedical Systems Corporation · 2007
Show all 9 generations
- Root · 9 generations backK894632 Model SSH-140A Ultrasound Imaging SystemToshiba America Medical Systems, In.C · 1989
- 8 generations backK941593 Diagnostic Ultrasonic TransducerToshiba America Medical Systems, In.C · 1996
- 7 generations backK970047 Phased Array Transducer - - PSK-20CTToshiba America Medical Systems, In.C · 1997
- 6 generations backK991858 Ssa-390A/Power Vision 8000Toshiba America Medical Systems, In.C · 1999
- 5 generations backK013633 Ssa-770A, Aplio Ultrasound SystemToshiba America Medical Systems, In.C · 2001
- 4 generations backK022400 Ssa-700A, Aplio Diagnostic Ultrasound SystemToshiba America Medical Systems, In.C · 2002 · Recalled
- 3 generations backK032281 Ssa-770A - Version 4, Aplio Diagnostic Ultrasound SystemToshiba America Medical Systems, In.C · 2003
- 2 generations backK063130 Aplio XG Diagnostic Ultrasound System, Model Ssa-790A, Version 1.0Toshibamedical Systems Corporation · 2006 · Recalled
- 1 generation backK072000 Aplio XG Diagnostic Ultrasound System, Model Ssa-790A Version 2.0Toshiba America Medical Systems, In.C · 2007
- This deviceK072918 Model Ssa-680A Xario XG Diagnostic Ultrasound System, Version 1.00Toshibamedical Systems Corporation · 2007
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −9 | K894632 | Model SSH-140A Ultrasound Imaging System | Toshiba America Medical Systems, In.C | 1989 | — |
| −8 | K941593 | Diagnostic Ultrasonic Transducer | Toshiba America Medical Systems, In.C | 1996 | — |
| −7 | K970047 | Phased Array Transducer - - PSK-20CT | Toshiba America Medical Systems, In.C | 1997 | — |
| −6 | K991858 | Ssa-390A/Power Vision 8000 | Toshiba America Medical Systems, In.C | 1999 | — |
| −5 | K013633 | Ssa-770A, Aplio Ultrasound System | Toshiba America Medical Systems, In.C | 2001 | — |
| −4 | K022400 | Ssa-700A, Aplio Diagnostic Ultrasound System | Toshiba America Medical Systems, In.C | 2002 | 1 |
| −3 | K032281 | Ssa-770A - Version 4, Aplio Diagnostic Ultrasound System | Toshiba America Medical Systems, In.C | 2003 | — |
| −2 | K063130 | Aplio XG Diagnostic Ultrasound System, Model Ssa-790A, Version 1.0 | Toshibamedical Systems Corporation | 2006 | 1 |
| −1 | K072000 | Aplio XG Diagnostic Ultrasound System, Model Ssa-790A Version 2.0 | Toshiba America Medical Systems, In.C | 2007 | — |
| 0 (this device) | K072918 | Model Ssa-680A Xario XG Diagnostic Ultrasound System, Version 1.00 | Toshibamedical Systems Corporation | 2007 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1992 | 1 | 0 |
| 1993 | 1 | 0 |
| 1994 | 5 | 0 |
| 1995 | 5 | 0 |
| 1996 | 2 | 0 |
| 1997 | 6 | 0 |
| 1998 | 2 | 0 |
| 1999 | 6 | 0 |
| 2000 | 3 | 0 |
| 2001 | 1 | 0 |
| 2002 | 2 | 0 |
| 2003 | 3 | 0 |
| 2004 | 2 | 0 |
| 2005 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in ITX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new ITX device
- #1 K260114 Align® Cover and Pad Guidance Toolscore 13.6 · 6 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K072000 | Aplio XG Diagnostic Ultrasound System, Model Ssa-790A Version 2.0Summary page 1…Devices: Toshiba America Medical Systems believes that this device is substantially equivalent to: Toshiba SSA-790A, Aplio XG Version 2.00 Diagnostic Ultrasound; 510(k) K072000 Device Description: The Xario XG Ultrasound System is a mobile system. This system is a Track 3 device that employs a wide array… | 2007 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 5, 2013 | Class II | Z-0120-2014 | Device Design | The firm initiated this recall of certain Toshiba Diagnostic Ultrasound Systems because it has been found that under certain conditions, the temperature of the transducer head may increase in 2D+PW mode or 2D+CDI (POWER, ADF) +PW mode.… |
| Jan 5, 2010 | Not classified | Z-1997-2010 | Other | Toshiba America Medical Systems is initiating a field correction on the PVT-681MV Endocavitary Transducer used with the Aplio XG and Xario XG Ultrasound Systems due to an incorrect sweep direction setting. Images acquired using the… |
| May 11, 2009 | Not classified | Z-1100-2010 | Device Design | Toshiba America Medical Systems Inc is removing the PET-511BTM Multi-plane Transesophageal Transducer from the market after receiving several reports of injury regarding the transducer. Bending or buckling of the flexible portion of the… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 18 years and passes through 2 companies; a new generation was cleared every 12 months on median.
- 5 of the 50 ancestors have a recall of any class.
- 3 other devices also cite its predicate K072000.
How does its category compare?
All ITX devicesITX Transducer, Ultrasonic, Diagnostic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.
FDA decisions per year · ITX
4in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · ITX
126in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · ITX
2in 2025
Class I (10 since 2007) Class II/III not classified
Adverse event reports per year · ITX
571in 2025
MAUDE reports naming product code ITX, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 10 | 26 | 0 | 16 |
| 2008 | 12 | 73 | 4 | 39 |
| 2009 | 17 | 28 | 4 | 23 |
| 2010 | 10 | 42 | 4 | 30 |
| 2011 | 10 | 16 | 1 | 34 |
| 2012 | 7 | 25 | 1 | 29 |
| 2013 | 3 | 105 | 0 | 38 |
| 2014 | 4 | 41 | 1 | 43 |
| 2015 | 1 | — | 5 | 183 |
| 2016 | 4 | 151 | 0 | 269 |
| 2017 | 6 | 110 | 1 | 412 |
| 2018 | 2 | — | 15 | 130 |
| 2019 | 1 | — | 0 | 119 |
| 2020 | 3 | 120 | 3 | 116 |
| 2021 | 3 | 139 | 10 (10) | 146 |
| 2022 | 2 | — | 6 | 87 |
| 2023 | 6 | 213 | 1 | 138 |
| 2024 | 3 | 81 | 1 | 988 |
| 2025 | 4 | 126 | 2 | 571 |
| 2026 | 6 | 158 | 0 | — |
What is it indicated for?
The Xario XG is intended to be used for the following type of studies; fetal, abdominal, intraoperative, pediatric, small organs, neonatal cephalic, adult cephalic, cardiac, transrectal, transvaginal, transesophageal, peripheral vascular and musculo-skeletal (both conventional and superficial).
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K072918 · printed Nov 4, 2026
Cite this page
510(k) Report. “K072918: Model Ssa-680A Xario XG Diagnostic Ultrasound System, Version 1.00.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K072918
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026