Model Ssa-680A Xario XG Diagnostic Ultrasound System, Version 1.00

K072918510(k) TraditionalCleared Oct 29, 200714-day reviewClass II21 CFR 892.1570Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Toshibamedical Systems Corporation · ITX Transducer, Ultrasonic, Diagnostic

Cleared on Oct 29, 2007 as substantially equivalent to K072000 (Toshiba America Medical Systems, In.C, 2007) after a 14-day review. Its family tree reaches back 9 generations to 1989. It has been recalled 3 times.

Decided
Oct 29, 2007
Review
14 days
Class
Class II
Product code
ITX
Recalls
3
Predicates
1

Its longest line of descent runs back 9 generations to K894632 (1989). 50 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 9 generations backK894632 Model SSH-140A Ultrasound Imaging SystemToshiba America Medical Systems, In.C · 1989
  2. 5 generations between
  3. 3 generations backK032281 Ssa-770A - Version 4, Aplio Diagnostic Ultrasound SystemToshiba America Medical Systems, In.C · 2003
  4. 2 generations backK063130 Aplio XG Diagnostic Ultrasound System, Model Ssa-790A, Version 1.0Toshibamedical Systems Corporation · 2006 · Recalled
  5. 1 generation backK072000 Aplio XG Diagnostic Ultrasound System, Model Ssa-790A Version 2.0Toshiba America Medical Systems, In.C · 2007
  6. This deviceK072918 Model Ssa-680A Xario XG Diagnostic Ultrasound System, Version 1.00Toshibamedical Systems Corporation · 2007
Show all 9 generations
  1. Root · 9 generations backK894632 Model SSH-140A Ultrasound Imaging SystemToshiba America Medical Systems, In.C · 1989
  2. 8 generations backK941593 Diagnostic Ultrasonic TransducerToshiba America Medical Systems, In.C · 1996
  3. 7 generations backK970047 Phased Array Transducer - - PSK-20CTToshiba America Medical Systems, In.C · 1997
  4. 6 generations backK991858 Ssa-390A/Power Vision 8000Toshiba America Medical Systems, In.C · 1999
  5. 5 generations backK013633 Ssa-770A, Aplio Ultrasound SystemToshiba America Medical Systems, In.C · 2001
  6. 4 generations backK022400 Ssa-700A, Aplio Diagnostic Ultrasound SystemToshiba America Medical Systems, In.C · 2002 · Recalled
  7. 3 generations backK032281 Ssa-770A - Version 4, Aplio Diagnostic Ultrasound SystemToshiba America Medical Systems, In.C · 2003
  8. 2 generations backK063130 Aplio XG Diagnostic Ultrasound System, Model Ssa-790A, Version 1.0Toshibamedical Systems Corporation · 2006 · Recalled
  9. 1 generation backK072000 Aplio XG Diagnostic Ultrasound System, Model Ssa-790A Version 2.0Toshiba America Medical Systems, In.C · 2007
  10. This deviceK072918 Model Ssa-680A Xario XG Diagnostic Ultrasound System, Version 1.00Toshibamedical Systems Corporation · 2007
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K894632Model SSH-140A Ultrasound Imaging SystemToshiba America Medical Systems, In.C1989—
−8K941593Diagnostic Ultrasonic TransducerToshiba America Medical Systems, In.C1996—
−7K970047Phased Array Transducer - - PSK-20CTToshiba America Medical Systems, In.C1997—
−6K991858Ssa-390A/Power Vision 8000Toshiba America Medical Systems, In.C1999—
−5K013633Ssa-770A, Aplio Ultrasound SystemToshiba America Medical Systems, In.C2001—
−4K022400Ssa-700A, Aplio Diagnostic Ultrasound SystemToshiba America Medical Systems, In.C20021
−3K032281Ssa-770A - Version 4, Aplio Diagnostic Ultrasound SystemToshiba America Medical Systems, In.C2003—
−2K063130Aplio XG Diagnostic Ultrasound System, Model Ssa-790A, Version 1.0Toshibamedical Systems Corporation20061
−1K072000Aplio XG Diagnostic Ultrasound System, Model Ssa-790A Version 2.0Toshiba America Medical Systems, In.C2007—
0 (this device)K072918Model Ssa-680A Xario XG Diagnostic Ultrasound System, Version 1.00Toshibamedical Systems Corporation2007—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199210
199310
199450
199550
199620
199760
199820
199960
200030
200110
200220
200330
200420
200520
FIG. 1 — Lineage of K072918, 1989–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in ITX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new ITX device

  1. #1 K260114 Align® Cover and Pad Guidance Toolscore 13.6 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K072000Aplio XG Diagnostic Ultrasound System, Model Ssa-790A Version 2.0
Summary page 1
…Devices: Toshiba America Medical Systems believes that this device is substantially equivalent to: Toshiba SSA-790A, Aplio XG Version 2.00 Diagnostic Ultrasound; 510(k) K072000 Device Description: The Xario XG Ultrasound System is a mobile system. This system is a Track 3 device that employs a wide array…
Toshiba America Medical Systems, In.C2007—

Also cleared against K072000: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 5, 2013Class IIZ-0120-2014Device DesignThe firm initiated this recall of certain Toshiba Diagnostic Ultrasound Systems because it has been found that under certain conditions, the temperature of the transducer head may increase in 2D+PW mode or 2D+CDI (POWER, ADF) +PW mode.…
Jan 5, 2010Not classifiedZ-1997-2010OtherToshiba America Medical Systems is initiating a field correction on the PVT-681MV Endocavitary Transducer used with the Aplio XG and Xario XG Ultrasound Systems due to an incorrect sweep direction setting. Images acquired using the…
May 11, 2009Not classifiedZ-1100-2010Device DesignToshiba America Medical Systems Inc is removing the PET-511BTM Multi-plane Transesophageal Transducer from the market after receiving several reports of injury regarding the transducer. Bending or buckling of the flexible portion of the…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 2 companies; a new generation was cleared every 12 months on median.
  • 5 of the 50 ancestors have a recall of any class.
  • 3 other devices also cite its predicate K072000.

How does its category compare?

All ITX devices

ITX Transducer, Ultrasonic, Diagnostic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · ITX

4in 2025

2007: 102008: 122009: 172010: 102011: 102012: 72013: 32014: 42015: 12016: 42017: 62018: 22019: 12020: 32021: 32022: 22023: 62024: 32025: 42026 (partial year): 6
2007high 17 (2009) · low 1 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · ITX

126in 2025

2007: 26 days median2008: 73 days median2009: 28 days median2010: 42 days median2011: 16 days median2012: 25 days median2013: 105 days median2014: 41 days median2016: 151 days median2017: 110 days median2020: 120 days median2021: 139 days median2023: 213 days median2024: 81 days median2025: 126 days median2026 (partial year): 158 days median
2007high 213 (2023) · low 16 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · ITX

2in 2025

2008: 42009: 42010: 42011: 12012: 12014: 12015: 52017: 12018: 152020: 32021: 10, 10 Class I2022: 62023: 12024: 12025: 2
2007high 15 (2018) · low 0 (2007)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · ITX

571in 2025

2007: 162008: 392009: 232010: 302011: 342012: 292013: 382014: 432015: 1832016: 2692017: 4122018: 1302019: 1192020: 1162021: 1462022: 872023: 1382024: 9882025: 571
2007high 988 (2024) · low 16 (2007)2025

MAUDE reports naming product code ITX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071026016
20081273439
20091728423
20101042430
20111016134
2012725129
20133105038
2014441143
20151—5183
201641510269
201761101412
20182—15130
20191—0119
202031203116
2021313910 (10)146
20222—687
202362131138
20243811988
202541262571
202661580—

What is it indicated for?

The Xario XG is intended to be used for the following type of studies; fetal, abdominal, intraoperative, pediatric, small organs, neonatal cephalic, adult cephalic, cardiac, transrectal, transvaginal, transesophageal, peripheral vascular and musculo-skeletal (both conventional and superficial).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K072918: Model Ssa-680A Xario XG Diagnostic Ultrasound System, Version 1.00.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K072918

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026