510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Aquilion Precision (TSX-304A/2) V8.6 with FIRST 3.0

K173468510(k) TraditionalCleared Feb 23, 2018107-day reviewClass II21 CFR 892.1750Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Toshibamedical Systems Corporation · JAK System, X-Ray, Tomography, Computed

Cleared on Feb 23, 2018 as substantially equivalent to K141741 (Toshibamedical Systems Corporation, 2014) after a 107-day review. Its family tree reaches back 3 generations to 2012. 4 later devices descend from it. It has been recalled once.

Decided
Feb 23, 2018
Review
107 days
Class
Class II
Product code
JAK
Recalls
1
Predicates
1

Its longest line of descent runs back 3 generations to K120710 (2012). 4 later devices descend from it, 1 citing it directly.

Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 1 descendant K141741 · Aquilion Prime, V6.00 · Toshibamedical Systems Corporation · 2014 · recalledK130645 · Aquilion Prime, V5.00 · Toshibamedical Systems Corporation · 2013K120710 · Aquilion Prime · Toshibamedical Systems Corporation · 2012 · recalled0
  1. Root · 3 generations backK120710 Aquilion PrimeToshibamedical Systems Corporation · 2012 · Recalled
  2. 2 generations backK130645 Aquilion Prime, V5.00Toshibamedical Systems Corporation · 2013
  3. 1 generation backK141741 Aquilion Prime, V6.00Toshibamedical Systems Corporation · 2014 · Recalled
  4. This deviceK173468 Aquilion Precision (TSX-304A/2) V8.6 with FIRST 3.0Toshibamedical Systems Corporation · 2018
  5. Descendant · cites it directly · 3 of its ownK182901 Aquilion Precision (TSX-304A/1 and /2) V8.8 with AiCECanon Medical Systems Corporation · 2019
  6. Descendant · 2 generations on · 2 of its ownK220986 Aquilion Precision (TSX-304A/4) V10.10 with AiCECanon Medical Systems Corporation · 2022
  7. Descendant · 3 generations on · 1 of its ownK223726 Aquilion Precision (TSX-304A/4) V10.14 with AiCECanon Medical Systems Corporation · 2023
  8. Descendant · 4 generations on · 0 of its ownK243398 CT Scanner TSX-501R/1 V11.1Canon Medical Systems Corporation · 2025

Most influential descendants

Showing 4 of 4 · See all
  • K182901Aquilion Precision (TSX-304A/1 and /2) V8.8 with AiCECanon Medical Systems Corporation · 2019 · cites it directly3descendants
  • K220986Aquilion Precision (TSX-304A/4) V10.10 with AiCECanon Medical Systems Corporation · 2022 · 2 generations on2descendants
  • K223726Aquilion Precision (TSX-304A/4) V10.14 with AiCECanon Medical Systems Corporation · 2023 · 3 generations on1descendant
  • K243398CT Scanner TSX-501R/1 V11.1Canon Medical Systems Corporation · 2025 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K120710Aquilion PrimeToshibamedical Systems Corporation20122
−2K130645Aquilion Prime, V5.00Toshibamedical Systems Corporation2013—
−1K141741Aquilion Prime, V6.00Toshibamedical Systems Corporation20142
0 (this device)K173468Aquilion Precision (TSX-304A/2) V8.6 with FIRST 3.0Toshibamedical Systems Corporation2018—
+1K182901Aquilion Precision (TSX-304A/1 and /2) V8.8 with AiCECanon Medical Systems Corporation2019—
+2K220986Aquilion Precision (TSX-304A/4) V10.10 with AiCECanon Medical Systems Corporation2022—
+3K223726Aquilion Precision (TSX-304A/4) V10.14 with AiCECanon Medical Systems Corporation2023—
+4K243398CT Scanner TSX-501R/1 V11.1Canon Medical Systems Corporation2025—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202210
202310
202510
FIG. 1 — Lineage of K173468, 2012–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141741Aquilion Prime, V6.00
Summary page 5
…America 21 CFR System, X‐ray, v6.00 Tomography K141741 11/11/2014 Medical 892.1750 Tomography, (Primary Predicate…
Toshibamedical Systems Corporation20142
†K170177Aquilion ONE (TSX-305A/3) V8.3 with FIRST 2.1Toshibamedical Systems Corporation20171

Also cleared against K141741: 3 other devices →

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K182901Aquilion Precision (TSX-304A/1 and /2) V8.8 with AiCECanon Medical Systems Corporation20193

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 25, 2019Class IIZ-1725-2020Software designScanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 6 years and passes through 1 company; a new generation was cleared every 1.5 years on median.
  • 2 of the 3 ancestors have a recall of any class.
  • Of its descendants, K182901 (Canon Medical Systems Corporation, 2019) has gone furthest, with 3 descendants of its own.
  • 3 other devices also cite its predicate K141741.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head. The Aquilion Precision has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and gqualified physician.

FIRST 3.0 is an iterative reconstruction algorithm intended to reduce exposure dose and improve high contrast spatial resolution for abdomen, pelvis, chest, cardiac, extremities and head applications.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K173468: Aquilion Precision (TSX-304A/2) V8.6 with FIRST 3.0.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K173468

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026