510(k) Report

510(k) · Cleared

Cleared

BD Phoenix Automated Microbiology System-Piperacillin (GN) -128 UG/ML

K073433510(k) TraditionalCleared Jan 14, 200839-day reviewClass II21 CFR 866.1645

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Becton, Dickinson & CO · LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation

Cleared on Jan 14, 2008 as substantially equivalent to K020321 (Becton, Dickinson & CO, 2002) and 1 other predicate after a 39-day review. Its family tree reaches back 2 generations to 2002.

Decided
Jan 14, 2008
Review
39 days
Class
Class II
Product code
LON
Recalls
None
Predicates
3

Its longest line of descent runs back 2 generations to K020322 (2002). 4 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor K041572 · BD Phoenix Automated Microbiology System-Pipercillin (GN) 0.5-128 UG/ML · Becton, Dickinson and Co. · 2004K020322 · BD Phoenix Automated Microbiology System · Becton, Dickinson & CO · 20020
  1. Root · 2 generations backK020322 BD Phoenix Automated Microbiology SystemBecton, Dickinson & CO · 2002
  2. 1 generation backK041572 BD Phoenix Automated Microbiology System-Pipercillin (GN) 0.5-128…Becton, Dickinson and Co. · 2004
  3. This deviceK073433 BD Phoenix Automated Microbiology System-Piperacillin (GN) -128 UG/MLBecton, Dickinson & CO · 2008
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K020322BD Phoenix Automated Microbiology SystemBecton, Dickinson & CO2002—
−1K041572BD Phoenix Automated Microbiology System-Pipercillin (GN) 0.5-128 UG/MLBecton, Dickinson and Co.2004—
0 (this device)K073433BD Phoenix Automated Microbiology System-Piperacillin (GN) -128 UG/MLBecton, Dickinson & CO2008—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200220
FIG. 1 — Lineage of K073433, 2002–2008Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LON? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LON device

  1. #1 K260619 VITEK 2 AST-Streptococcus Cefepime (<=0.06 –…score 13.3 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
Kept off the family treeP050510VITEK System
Not in FDA’s 510(k), De Novo or PMA lists as we hold them (a misreading, a typo, or a number FDA does not list), so it is kept off the family tree
*K020321BD Phoenix Automated Microbiology SystemBecton, Dickinson & CO20024
*K041572BD Phoenix Automated Microbiology System-Pipercillin (GN) 0.5-128 UG/ML
Summary page 1
…Automated Microbiology System with Gatifioxacin (K02032 l May 23, 2002) and Piperacillin (K04 1572, August 12, 2004). INTENDED USE: The BD Phoenix",' Automated Microbiology System is…
Becton, Dickinson and Co.2004—

Also cleared against K020321: 102 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 6 years and passes through 2 companies; a new generation was cleared every 3.4 years on median.
  • 1 of the 4 ancestors has a recall of any class.
  • 102 other devices also cite its predicate K020321.

How does its category compare?

All LON devices

LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · LON

5in 2025

2007: 192008: 72009: 82010: 12011: 82012: 102013: 72014: 42016: 32017: 102018: 42019: 82020: 52021: 52022: 82023: 112024: 22025: 52026 (partial year): 4
2007high 19 (2007) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LON

90in 2025

2007: 50 days median2008: 66 days median2009: 76 days median2011: 85 days median2012: 200 days median2013: 115 days median2014: 90 days median2016: 242 days median2017: 89 days median2018: 88 days median2019: 84 days median2020: 81 days median2021: 268 days median2022: 189 days median2023: 210 days median2025: 90 days median2026 (partial year): 146 days median
2007high 268 (2021) · low 50 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LON

1in 2025

2008: 12010: 1162011: 32012: 32014: 2, 2 Class I2016: 22017: 12018: 62019: 12020: 82022: 152024: 12025: 12026 (partial year): 6
2007high 116 (2010) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · LON

690in 2025

2007: 32008: 12010: 32011: 42012: 92013: 32014: 102015: 412016: 942017: 1972018: 5122019: 3152020: 2182021: 2322022: 1802023: 4532024: 4182025: 690
2007high 690 (2025) · low 0 (2009)2025

MAUDE reports naming product code LON, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007195003
200876611
200987600
20101—1163
201188534
20121020039
2013711503
20144902 (2)10
20150—041
20163242294
201710891197
20184886512
20198841315
20205818218
202152680232
2022818915180
2023112100453
20242—1418
20255901690
202641466—

What is it indicated for?

The BD PhoenixTM Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus and Streptococcus. This…

Read the full statement (155 words)

The BD PhoenixTM Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus and Streptococcus.

This premarket notification is for Piperacillin at concentrations of 2-128 }zg/mL to Gram-negative ID/AST or AST only Phoenix panels with the removal of the limitations for Proteus mirabilis, Providencia species and Stenotrophomonas mcaltophilia. Piperacillin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Acinetobacter species Klebsiella species Providencia rettgeri

Enterobacter species Morganella morganii Pseudoironas aeruginosa

Escherichia coli Proteus mirabilis Serratia species

Active In Vitro Against:

Burkholderia cepacia Citrobacter fi'eundii Stenotrophomonas maltophi/ia

Citrobacter diversus Pseudomonas fluorescens Yersinia enterocolitica

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K073433: BD Phoenix Automated Microbiology System-Piperacillin (GN) -128 UG/ML.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K073433

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026