Thor Anterior Plating System

K073437510(k) TraditionalCleared Feb 13, 200869-day reviewClass II21 CFR 888.3060Recalls 31

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Stryker Spine · KWQ Appliance, Fixation, Spinal Intervertebral Body

Cleared on Feb 13, 2008 as substantially equivalent to K022791 (Synthes (Usa), 2002) and 3 other predicates after a 69-day review. Its family tree reaches back 2 generations to 1998. 102 later devices descend from it. It has been recalled 31 times.

Decided
Feb 13, 2008
Review
69 days
Class
Class II
Product code
KWQ
Recalls
31
Predicates
4

Its longest line of descent runs back 2 generations to K983583 (1998). 5 ancestors in all. 102 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor2001: 1 other ancestor2000: 1 other ancestor2008: 1 descendant cleared2009: 1 descendant cleared2014: 1 descendant cleared2013: 1 descendant cleared2015: 1 descendant cleared2019: 7 descendants cleared2023: 20 descendants cleared2024: 16 descendants cleared2018: 4 descendants cleared2020: 4 descendants cleared2022: 11 descendants cleared2026: 16 descendants cleared2021: 6 descendants cleared2025: 13 descendants clearedK983583 · Depuy Acromed Moss and Moss Miami Spinal Systems · Depuy, Inc. · 1998K994347 · Centaur Spinal System · Howmedica Osteonics Corp. · 20001995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors 1 generation forward: 1 descendant2 generations forward: 2 descendants3 generations forward: 9 descendants4 generations forward: 15 descendants5 generations forward: 19 descendants6 generations forward: 19 descendants7 generations forward: 14 descendants8 generations forward: 10 descendants9 generations forward: 9 descendants10 generations forward: 3 descendants11 generations forward: 1 descendant K994347 · Centaur Spinal System · Howmedica Osteonics Corp. · 2000K983583 · Depuy Acromed Moss and Moss Miami Spinal Systems · Depuy, Inc. · 19980+5+10
  1. Root · 2 generations backK983583 Depuy Acromed Moss and Moss Miami Spinal SystemsDepuy, Inc. · 1998
  2. 1 generation backK994347 Centaur Spinal SystemHowmedica Osteonics Corp. · 2000
  3. This deviceK073437 Thor Anterior Plating SystemStryker Spine · 2008
  4. Descendant · cites it directly · 101 of its ownK080773 Thor Anterior Plating SystemStryker Spine · 2008
  5. Descendant · 2 generations on · 73 of its ownK091253 Thoracolumbar Plate, Model 1208-41FXX, Lumber Plate, Model…K2m, Inc. · 2009
  6. Descendant · 3 generations on · 72 of its ownK131533 Cayman MI Plate SystemK2m, Inc. · 2013
  7. Descendant · 4 generations on · 52 of its ownK172941 CASCADIA Interbody SystemK2m · 2018
  8. Descendant · 2 generations on · 46 of its ownK142699 LITe Plate SystemStryker Corporation · 2014

Showing 5 of 102 descendants. See all descendants

Descendants cleared per year

2008: 12009: 12013: 12014: 12015: 12018: 42019: 72020: 42021: 62022: 112023: 202024: 162025: 132026: 16
2008high 20 (2023) · low 0 (2010)2026

Most influential descendants

Showing 5 of 102 · See all
  • K080773Thor Anterior Plating SystemStryker Spine · 2008 · cites it directly101descendants
  • K091253Thoracolumbar Plate, Model 1208-41FXX, Lumber Plate, Model 1208-41FXXL, Sacral Plate…K2m, Inc. · 2009 · 2 generations on73descendants
  • K131533Cayman MI Plate SystemK2m, Inc. · 2013 · 3 generations on72descendants
  • K172941CASCADIA Interbody SystemK2m · 2018 · 4 generations on52descendants
  • K142699LITe Plate SystemStryker Corporation · 2014 · 2 generations on46descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K983583Depuy Acromed Moss and Moss Miami Spinal SystemsDepuy, Inc.1998—
−1K994347Centaur Spinal SystemHowmedica Osteonics Corp.2000—
0 (this device)K073437Thor Anterior Plating SystemStryker Spine2008—
+1K080773Thor Anterior Plating SystemStryker Spine2008—
+2K091253Thoracolumbar Plate, Model 1208-41FXX, Lumber Plate, Model 1208-41FXXL, Sacral…K2m, Inc.2009—
+3K131533Cayman MI Plate SystemK2m, Inc.2013—
+4K172941CASCADIA Interbody SystemK2m2018—
+2K142699LITe Plate SystemStryker Corporation2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200110
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
200910
201310
201410
201510
201840
201970
202040
202160
2022110
2023200
2024160
2025130
2026160
FIG. 1 — Lineage of K073437, 1998–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KWQ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KWQ device

  1. #1 K233640 Segmental Plating System (SPS);IdentiTi SPS…score 21.9 · 31 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K022791Synthes Anterior Tension Band (Atb) System
Summary page 1
…Body Fixation Orthosis and Number 21 CFR 888.3060 Product Code KWQ Predicate Devices Synthes Anterior Tension Band System: 510(k) #K022791 Stryker Spine CENTAUR Spinal System: 510(k) #K994347,#K001844 Stryker Spine X…
Synthes (Usa)2002—
*K994347Centaur Spinal System
Summary page 1
…KWQ Predicate Devices Synthes Anterior Tension Band System: 510(k) #K022791 Stryker Spine CENTAUR Spinal System: 510(k) #K994347,#K001844 Stryker Spine Xia Stainless Steel System: 510(k) #K012870 Device Description The Stry…
Howmedica Osteonics Corp.2000—
*K001844Modification to Centaur Spinal System
Summary page 1
…KWQ Predicate Devices Synthes Anterior Tension Band System: 510(k) #K022791 Stryker Spine CENTAUR Spinal System: 510(k) #K994347,#K001844 Stryker Spine Xia Stainless Steel System: 510(k) #K012870 Device Description The Stryker Spine…
Howmedica Osteonics Corp.2000—
*K012870Xia Stainless Steel System
Summary page 1
…22791 Stryker Spine CENTAUR Spinal System: 510(k) #K994347,#K001844 Stryker Spine Xia Stainless Steel System: 510(k) #K012870 Device Description The Stryker Spine THORTm Anterior Plating System is designed for anterior a…
Howmedica Osteonics Corp.2001—

Also cleared against K022791: 14 other devices →

What descends from it? (102)

See all 102

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K080773Thor Anterior Plating SystemStryker Spine2008101

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 2, 2009Not classifiedZ-0426-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0427-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0430-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0435-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0441-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0442-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0447-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0453-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0431-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0432-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0433-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0434-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0440-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0454-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0424-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0425-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0436-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0439-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0443-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0445-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0446-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0449-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0428-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0450-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0451-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0452-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0429-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0437-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0438-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0444-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…
Oct 2, 2009Not classifiedZ-0448-2010Device DesignSurgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 3 companies; a new generation was cleared every 7.9 years on median.
  • None of the 5 ancestors has a recall.
  • Of its descendants, K080773 (Stryker Spine, 2008) has gone furthest, with 101 descendants of its own.
  • 14 other devices also cite its predicate K022791.

How does its category compare?

All KWQ devices

KWQ Appliance, Fixation, Spinal Intervertebral Body: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · KWQ

7in 2025

2007: 272008: 322009: 232010: 232011: 222012: 262013: 222014: 172015: 252016: 152017: 202018: 222019: 262020: 262021: 172022: 182023: 162024: 172025: 72026 (partial year): 10
2007high 32 (2008) · low 7 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KWQ

64in 2025

2007: 101 days median2008: 70 days median2009: 90 days median2010: 104 days median2011: 95 days median2012: 90 days median2013: 88 days median2014: 124 days median2015: 102 days median2016: 121 days median2017: 67 days median2018: 95 days median2019: 81 days median2020: 55 days median2021: 59 days median2022: 85 days median2023: 59 days median2024: 64 days median2025: 64 days median2026 (partial year): 76 days median
2007high 124 (2014) · low 55 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KWQ

1in 2018

2007: 22008: 332009: 352010: 22011: 52012: 72013: 62015: 22016: 92017: 82018: 1
2007high 35 (2009) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KWQ

248in 2025

2007: 2882008: 3482009: 3552010: 6912011: 4152012: 5442013: 7962014: 5462015: 5622016: 5022017: 4922018: 3412019: 2982020: 4002021: 5912022: 3922023: 3202024: 4192025: 248
2007high 796 (2013) · low 248 (2025)2025

MAUDE reports naming product code KWQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007271012288
2008327033348
2009239035355
2010231042691
201122955415
201226907544
201322886796
2014171240546
2015251022562
2016151219502
201720678492
201822951341
201926810298
202026550400
202117590591
202218850392
202316590320
202417640419
20257640248
202610760—

What is it indicated for?

The Stryker Spine THOR TM Anterior Plating System is indicated for use via a lateral or anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach below the bifurcation of the great vessels. The Styker Spine THORTM Anterior Plating System is intended for use in the lumbar and lumbosacral spine (L1-Si). THOR"4 Anterior Plating System is intended to provide additional support during fusion using…

Read the full statement (137 words)

The Stryker Spine THOR TM Anterior Plating System is indicated for use via a lateral or anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach below the bifurcation of the great vessels.

The Styker Spine THORTM Anterior Plating System is intended for use in the lumbar and lumbosacral spine (L1-Si). THOR"4 Anterior Plating System is intended to provide additional support during fusion using autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities:

* Degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies);

* Pseudoarthrosis;

* Spondylolysis;

* Spondylolisthesis;

* Trauma (i.e., fracture or dislocation);

* Deformities (i.e., scoliosis or lordosis)

* Spinal Stenosis; and

* Failed Previous Fusion.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K073437: Thor Anterior Plating System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K073437

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026