510(k) Report

510(k) · Cleared

Cleared

Angiotech CVC

K073520510(k) TraditionalCleared Apr 11, 2008119-day reviewClass II21 CFR 880.5200

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Angiotech Pharmaceuticals, Inc. · FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days

Cleared on Apr 11, 2008 as substantially equivalent to K993691 (Arrow Intl., Inc., 2000) and 1 other predicate after a 119-day review. Its family tree reaches back 1 generation to 1990. 2 of the 3 devices in its family, counting itself, have a Class I recall.

Decided
Apr 11, 2008
Review
119 days
Class
Class II
Product code
FOZ
Recalls
None
Predicates
2

Its longest line of descent runs back 1 generation to K900263 (1990). 2 ancestors in all. No later device cites it as a predicate yet. 2 of the 3 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2000: 1 other ancestor, 1 with a Class I recall1K900263 · Antimicrobial Multi-Lumen Central Venous Catheter · Arrow Intl., Inc. · 1990 · Class I recall19901995200020052010
Other ancestors, by generation 1 generation back: 1 other ancestor, 1 with a Class I recall1 K900263 · Antimicrobial Multi-Lumen Central Venous Catheter · Arrow Intl., Inc. · 1990 · Class I recall0
  1. Root · 1 generation backK900263 Antimicrobial Multi-Lumen Central Venous CatheterArrow Intl., Inc. · 1990 · Recalled (Class I)
  2. This deviceK073520 Angiotech CVCAngiotech Pharmaceuticals, Inc. · 2008

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K900263Antimicrobial Multi-Lumen Central Venous CatheterArrow Intl., Inc.1990Class I
0 (this device)K073520Angiotech CVCAngiotech Pharmaceuticals, Inc.2008—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200011
FIG. 1 — Lineage of K073520, 1990–2008Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FOZ device

  1. #1 K243403 BD Nexiva™ Closed IV Catheter Systemscore 14.8 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K993691Arrowg +Ard Blue Plus Antimicrobial Multi-Lumen Central Venous Catheter
Summary page 2
…days. Product Code FOZ, 21 CFR 880.5200, Class II Legally marketed ARROWg+ard Blue® Central Venous Catheter devices to which the K993691 and K900263 device is substantially equivalent: Description of the The Angiotech® CVC Multi-lumen Central Venous Catheter…
Arrow Intl., Inc.2000Class I
*K900263Antimicrobial Multi-Lumen Central Venous Catheter
Summary page 2
…Product Code FOZ, 21 CFR 880.5200, Class II Legally marketed ARROWg+ard Blue® Central Venous Catheter devices to which the K993691 and K900263 device is substantially equivalent: Description of the The Angiotech® CVC Multi-lumen Central Venous Catheter with device…
Arrow Intl., Inc.1990Class I

Also cleared against K993691: 5 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (2 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K993691Arrowg +Ard Blue Plus Antimicrobial Multi-Lumen Central Venous CatheterArrow Intl., Inc. · 2000 · Direct predicate · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall
  • K900263Antimicrobial Multi-Lumen Central Venous CatheterArrow Intl., Inc. · 1990 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls

What does this lineage say?

  • 2 of the 2 ancestors have a recall of any class; 2 have a Class I recall.
  • 5 other devices also cite its predicate K993691.

How does its category compare?

All FOZ devices

FOZ Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · FOZ

11in 2025

2007: 62008: 92009: 92010: 32011: 62012: 132013: 42014: 92015: 102016: 92017: 92018: 12019: 82020: 102021: 42022: 82023: 92024: 62025: 112026 (partial year): 7
2007high 13 (2012) · low 1 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FOZ

91in 2025

2007: 159 days median2008: 103 days median2009: 30 days median2010: 28 days median2011: 132 days median2012: 122 days median2013: 104 days median2014: 99 days median2015: 107 days median2016: 63 days median2017: 51 days median2019: 140 days median2020: 191 days median2021: 242 days median2022: 183 days median2023: 129 days median2024: 190 days median2025: 91 days median2026 (partial year): 203 days median
2007high 242 (2021) · low 28 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FOZ

4in 2023

2007: 162009: 232010: 1152013: 22014: 42016: 3, 1 Class I2017: 222018: 42019: 32020: 32021: 12022: 4, 2 Class I2023: 4, 2 Class I2026 (partial year): 2
2007high 115 (2010) · low 0 (2008)2026

Class I (5 since 2007) Class II/III not classified

Adverse event reports per year · FOZ

5,110in 2025

2007: 3742008: 4222009: 5082010: 3702011: 7862012: 5602013: 6402014: 8152015: 7292016: 9252017: 1,9662018: 3,3052019: 4,3652020: 3,7972021: 4,6742022: 4,1332023: 4,4302024: 4,7722025: 5,110
2007high 5,110 (2025) · low 370 (2010)2025

MAUDE reports naming product code FOZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007615916374
200891030422
200993023508
2010328115370
201161320786
2012131220560
201341042640
20149994815
2015101070729
20169633 (1)925
2017951221,966
20181—43,305
2019814034,365
20201019133,797
2021424214,674
202281834 (2)4,133
202391294 (2)4,430
2024619004,772
2025119105,110
202672032—

What is it indicated for?

The multiple-lumen catheter permits venous access to the central circulation. The Angiotech antimicrobial surface is intended to help provide protection against bacterial colonization on the catheter surface and it may help reduce catheter-related infections. The catheter is not intended to be used as a treatment for existing infections, nor is it indicated for longer than 30 days use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K073520: Angiotech CVC.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K073520

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026