Dimension Exl with LM Clinical Chemistry System, Dimension FT4L Flex Reagent Cartridge, and Loci Thyroid Calibrator, Mod

K073604510(k) TraditionalCleared Mar 13, 200883-day reviewClass II21 CFR 862.1695Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Dade Behring, Inc. · CEC Radioimmunoassay, Free Thyroxine

Cleared on Mar 13, 2008 as substantially equivalent to K053531 (Dade Behring, Inc., 2006) after an 83-day review. Its family tree reaches back 2 generations to 1997. 5 later devices descend from it. It has been recalled 7 times.

Decided
Mar 13, 2008
Review
83 days
Class
Class II
Product code
CEC
Recalls
7
Predicates
1

Its longest line of descent runs back 2 generations to K970389 (1997). 5 later devices descend from it, 4 citing it directly.

Descendants, by year cleared 2010: 1 descendant cleared2008: 1 descendant cleared2013: 1 descendant cleared2025: 1 descendant cleared2014: 1 descendant clearedK970389 · Thyroid Calibrator · Dade Chemistry Systems, Inc. · 1997K053531 · Dimension Vista Free Thyroxine Flex Reagent Cartridge and Loci 1 Calibrator · Dade Behring, Inc. · 20061995200020052010201520202025
Descendants, by generation 1 generation forward: 4 descendants2 generations forward: 1 descendant K053531 · Dimension Vista Free Thyroxine Flex Reagent Cartridge and Loci 1 Calibrator · Dade Behring, Inc. · 2006K970389 · Thyroid Calibrator · Dade Chemistry Systems, Inc. · 19970
  1. Root · 2 generations backK970389 Thyroid CalibratorDade Chemistry Systems, Inc. · 1997
  2. 1 generation backK053531 Dimension Vista Free Thyroxine Flex Reagent Cartridge and Loci 1…Dade Behring, Inc. · 2006
  3. This deviceK073604 Dimension Exl with LM Clinical Chemistry System, Dimension FT4L Flex…Dade Behring, Inc. · 2008
  4. Descendant · cites it directly · 1 of its ownK130276 Dimension Exl with LM SystemSiemens Healthcare Diagnostics, Inc. · 2013 · Recalled
  5. Descendant · cites it directly · 0 of its ownK243570 Dimension LOCI Thyroid Stimulating Hormone Flex reagent cartridge…Siemens Healthcare Diagnostics · 2025
  6. Descendant · 2 generations on · 0 of its ownK140842 Dimension Loci Free Thyroxine Flex Reagent Cartridge, FT4L…Siemens Healthcare Diagnostics, Inc. · 2014 · Recalled
  7. Descendant · cites it directly · 0 of its ownK100778 Dimension Exl Loci Thyroid Calibrator, Model RC 610ASiemens Healthcare Diagnostics · 2010
  8. Descendant · cites it directly · 0 of its ownK081103 Loci Thyroid Calibrator with Model RC610Siemens Healthcare Diagnostics, Inc. · 2008

Most influential descendants

Showing 5 of 5 · See all
  • K130276Dimension Exl with LM SystemSiemens Healthcare Diagnostics, Inc. · 2013 · cites it directly1descendant
  • K243570Dimension LOCI Thyroid Stimulating Hormone Flex reagent cartridge (TSHL); Dimension LOCI…Siemens Healthcare Diagnostics · 2025 · cites it directly0descendants
  • K140842Dimension Loci Free Thyroxine Flex Reagent Cartridge, FT4L, Dimension Loci Thyroid…Siemens Healthcare Diagnostics, Inc. · 2014 · 2 generations on0descendants
  • K100778Dimension Exl Loci Thyroid Calibrator, Model RC 610ASiemens Healthcare Diagnostics · 2010 · cites it directly0descendants
  • K081103Loci Thyroid Calibrator with Model RC610Siemens Healthcare Diagnostics, Inc. · 2008 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K970389Thyroid CalibratorDade Chemistry Systems, Inc.1997—
−1K053531Dimension Vista Free Thyroxine Flex Reagent Cartridge and Loci 1 CalibratorDade Behring, Inc.2006—
0 (this device)K073604Dimension Exl with LM Clinical Chemistry System, Dimension FT4L Flex Reagent…Dade Behring, Inc.2008—
+1K130276Dimension Exl with LM SystemSiemens Healthcare Diagnostics, Inc.20136
+1K243570Dimension LOCI Thyroid Stimulating Hormone Flex reagent cartridge (TSHL)…Siemens Healthcare Diagnostics2025—
+2K140842Dimension Loci Free Thyroxine Flex Reagent Cartridge, FT4L, Dimension Loci…Siemens Healthcare Diagnostics, Inc.20142
+1K100778Dimension Exl Loci Thyroid Calibrator, Model RC 610ASiemens Healthcare Diagnostics2010—
+1K081103Loci Thyroid Calibrator with Model RC610Siemens Healthcare Diagnostics, Inc.2008—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
201010
201310
201410
202510
FIG. 1 — Lineage of K073604, 1997–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CEC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CEC device

  1. #1 K240273 Access Free T4score 7.0 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K053531Dimension Vista Free Thyroxine Flex Reagent Cartridge and Loci 1 Calibrator
Summary page 1
…L Max is a family member of the Dimension® XL clinical chemistry system, which was cleared under K944093) Dimension VistaTM FT4 Flex® reagent cartridge, K053531 Dimension Vista TM LOCI I Calibrator, K053531…
Dade Behring, Inc.2006—
*—Dimension® RxL Max® clinical chemistry system
Named without a submission number
—

Also cleared against K053531: 1 other device →

What descends from it? (5)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K130276Dimension Exl with LM SystemSiemens Healthcare Diagnostics, Inc.20131
K243570Dimension LOCI Thyroid Stimulating Hormone Flex reagent cartridge (TSHL); Dimension LOCI…Siemens Healthcare Diagnostics20250
K100778Dimension Exl Loci Thyroid Calibrator, Model RC 610ASiemens Healthcare Diagnostics20100
K081103Loci Thyroid Calibrator with Model RC610Siemens Healthcare Diagnostics, Inc.20080

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 2, 2015Class IIZ-1615-2015Under Investigation by firmCertain Dimension Integrated Chemistry Systems Instrument Power Configurations may have been installed with a non-standard instrument power plug.The use of a non-standard plug configuration as described may lead to a potential electrical…
May 27, 2014Class IIIZ-2133-2014Employee errorThere is a misassembly issue with the photometer installed on Dimension¿ Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension…
Feb 27, 2013Class IIIZ-1046-2013Under Investigation by firmThrough investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these…
Feb 27, 2013Class IIIZ-1047-2013Under Investigation by firmThrough investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these…
Feb 27, 2013Class IIIZ-1048-2013Under Investigation by firmThrough investigation of customer complaints, Siemens Healthcare Diagnostics has confirmed an increase in System Check failures due to high System Check LOCI results. Upon further investigation, Siemens has determined that many of these…
Apr 12, 2010Not classifiedZ-0458-2011OtherSome LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high elevations, causing the cartridge to leak. Use of leaking cartridges may result in inaccurate test results.
Sep 30, 2008Not classifiedZ-0581-2009Software designError messages, "Wash Aspirator Failure" and "Wash station unable to recover" are not generated when one or both HM wash probes are totally occluded.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 2 companies; a new generation was cleared every 9.0 years on median.
  • None of the 2 ancestors has a recall.
  • Of its descendants, K130276 (Siemens Healthcare Diagnostics, Inc., 2013) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K053531.

How does its category compare?

All CEC devices

CEC Radioimmunoassay, Free Thyroxine: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · CEC

1in 2024

2007: 12008: 22009: 12012: 12013: 22014: 32015: 12016: 12017: 12018: 22023: 12024: 1
2007high 3 (2014) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CEC

238in 2014

2014: 238 days median
2007high 238 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CEC

1in 2018

2010: 32013: 12018: 1
2007high 3 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CEC

213in 2025

2009: 22010: 72011: 152012: 62013: 62014: 402015: 1662016: 1482017: 1962018: 1992019: 1672020: 1062021: 962022: 812023: 992024: 2762025: 213
2007high 276 (2024) · low 0 (2007)2025

MAUDE reports naming product code CEC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20082—00
20091—02
20100—37
20110—015
20121—06
20132—16
20143238040
20151—0166
20161—0148
20171—0196
20182—1199
20190—0167
20200—0106
20210—096
20220—081
20231—099
20241—0276
20250—0213
20260—0—

What is it indicated for?

The FT4L method is an in vitro diagnostic test for the quantitative measurement of Free Thyroxine in human serum and plasma on the Dimension® EXLTM with LM system. Measurements of free thyroxine are used in the diagnosis and monitoring of thyroid disease. The LOCI® Thyroid Calibrator is an in vitro diagnostic product for the calibration of the FT4L method on the Dimension® EXLTM with LM system. The Dimension® EXLTM with LM system is an in vitro…

Read the full statement (111 words)

The FT4L method is an in vitro diagnostic test for the quantitative measurement of Free Thyroxine in human serum and plasma on the Dimension® EXLTM with LM system. Measurements of free thyroxine are used in the diagnosis and monitoring of thyroid disease.

The LOCI® Thyroid Calibrator is an in vitro diagnostic product for the calibration of the FT4L method on the Dimension® EXLTM with LM system.

The Dimension® EXLTM with LM system is an in vitro diagnostic device that is intended to measure a variety of analytes in human body fluids. The system utilizes photometric, turbidimetric, chemiluminescence and integrated ion selective multisensor technology for chemical and immunochemical applications for clinical use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K073604: Dimension Exl with LM Clinical Chemistry System, Dimension FT4L Flex Reagent Cartridge, and Loci Thyroid Calibrator, Mod.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K073604

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026