510(k) · Cleared
Recalled · Class II · 2016MSK Extreme MR Scanner, Model AA5000
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Oni Medical Systems, Inc. · LNH System, Nuclear Magnetic Resonance Imaging
Cleared on Feb 6, 2008 as substantially equivalent to K001773 (Oni, Inc., 2000) and 2 other predicates after a 29-day review. Its family tree reaches back 4 generations to 1997. 2 of the 15 devices in its family, counting itself, have a Class I recall. 2 later devices descend from it. It has been recalled 3 times.
- Decided
- Feb 6, 2008
- Review
- 29 days
- Class
- Class II
- Product code
- LNH
- Recalls
- 3
- Predicates
- 3
Its longest line of descent runs back 4 generations to K971684 (1997). 12 ancestors in all. 2 later devices descend from it, 1 citing it directly. 2 of the 15 devices in this family, counting this one, have a Class I recall.
- Root · 4 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
- 3 generations backK032428 Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc. · 2003 · Recalled
- 2 generations backK041112 Magnetom Espree SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
- 1 generation backK052164 Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc. · 2005
- This deviceK080048 MSK Extreme MR Scanner, Model AA5000Oni Medical Systems, Inc. · 2008
- Descendant · cites it directly · 1 of its ownK103238 1.5T Optima MR430SGe Medical Systems, LLC · 2011 · Recalled
- Descendant · 2 generations on · 0 of its ownK123068 Optima MR 430SGe Medical Systems, LLC · 2012 · Recalled
Class I recalls in this family
Most influential descendants
Showing 2 of 2 · See allView FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −4 | K971684 | Magnetom Project 047 System (Numaris 3.5 Software) | Siemens Medical Solutions USA, Inc. | 1997 | 4 |
| −3 | K032428 | Magnetom Avanto MR System | Siemens Medical Solutions USA, Inc. | 2003 | 4 |
| −2 | K041112 | Magnetom Espree System | Siemens Medical Solutions USA, Inc. | 2004 | 4 |
| −1 | K052164 | Magnetom Systems with Software Syngo MR 2006A | Siemens Medical Solutions USA, Inc. | 2005 | — |
| 0 (this device) | K080048 | MSK Extreme MR Scanner, Model AA5000 | Oni Medical Systems, Inc. | 2008 | — |
| +1 | K103238 | 1.5T Optima MR430S | Ge Medical Systems, LLC | 2011 | 2 |
| +2 | K123068 | Optima MR 430S | Ge Medical Systems, LLC | 2012 | 2 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1999 | 1 | 0 |
| 2000 | 1 | 0 |
| 2001 | 1 | 0 |
| 2003 | 1 | 0 |
| 2005 | 4 | 2 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2011 | 1 | 0 |
| 2012 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LNH device
- #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K001773 | OrthoneSummary page 1…mon Name: Magnetic Resonance Diagnostic Device Classification Magnetic Resonance Diagnostic Device Name: Predicate Device: OrthOne Extremity MRI Scanner (K00 1773) Philips Medical Systems Achieva 1.5T(K052013) MAGNETOM Systems with Software syngo MR 20… | 2000 |
| * | K052013 | Achieva 1.5T & Inera 1.5T FamilySummary page 1…on Magnetic Resonance Diagnostic Device Name: Predicate Device: OrthOne Extremity MRI Scanner (K00 1773) Philips Medical Systems Achieva 1.5T(K052013) MAGNETOM Systems with Software syngo MR 2006A (l.5T Symphony version) (K052164) Device… | 2005 |
| * | K052164 | Magnetom Systems with Software Syngo MR 2006ASummary page 1…Philips Medical Systems Achieva 1.5T(K052013) MAGNETOM Systems with Software syngo MR 2006A (l.5T Symphony version) (K052164) Device The MSK ExtreMeTM MR Scanner utilizes a Description: superconducting magnet to acquire 2D single-sli… | 2005 |
What descends from it? (2)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K103238 | 1.5T Optima MR430S | 2011 | 1 |
Has anything in its family been recalled? (2 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 9, 2016 | Class II | Z-1927-2016 | Nonconforming Material/Component | The gas venting may not occur properly. A site reported a magnet quenching with subsequent venting of the cryogen gas into the magnet room. Cryogen gas that vents into the magnet room could cause critical deprivation of oxygen to users… |
| Mar 28, 2013 | Class II | Z-1211-2013 | Process control | Coolant supply lines may leak liquid onto the floor resulting in a slippery surface and fall hazard. |
| Aug 29, 2011 | Not classified | Z-0107-2012 | Nonconforming Material/Component | GE Healthcare has recently become aware that an exposed high voltage gradient wire internal to the MSK 1.5T Extreme system equipment may cause electrical shock, which may impact service personnel safety. This issue does not impact patient… |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K052013Achieva 1.5T & Inera 1.5T FamilyPhilips Medical Systems Nederland B.V. · 2005 · Direct predicate · ◆ in the 2005 ancestors bar of FIG. 13 Class I recalls
- K043147Achieva FamilyPhilips Medical Systems North America Co. · 2005 · 2 generations back · ◆ in the 2005 ancestors bar of FIG. 13 Class I recalls
What does this lineage say?
- The line of descent spans 11 years and passes through 2 companies; a new generation was cleared every 2.3 years on median.
- 10 of the 12 ancestors have a recall of any class; 2 have a Class I recall.
- Of its descendants, K103238 (Ge Medical Systems, LLC, 2011) has gone furthest, with 1 descendant of its own.
- 1 other device also cites its predicate K001773.
How does its category compare?
All LNH devicesLNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.
FDA decisions per year · LNH
26in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LNH
122in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LNH
8in 2025
Class I (8 since 2007) Class II/III not classified
Adverse event reports per year · LNH
1,263in 2025
MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 35 | 25 | 114 |
| 2008 | 19 | 53 | 12 | 167 |
| 2009 | 17 | 70 | 7 | 194 |
| 2010 | 12 | 94 | 17 | 165 |
| 2011 | 16 | 81 | 11 | 182 |
| 2012 | 22 | 89 | 5 | 170 |
| 2013 | 19 | 108 | 2 | 107 |
| 2014 | 17 | 108 | 13 | 119 |
| 2015 | 19 | 140 | 12 (2) | 113 |
| 2016 | 20 | 108 | 1 | 135 |
| 2017 | 18 | 98 | 0 | 121 |
| 2018 | 29 | 128 | 71 | 158 |
| 2019 | 26 | 62 | 5 | 158 |
| 2020 | 32 | 121 | 4 | 98 |
| 2021 | 27 | 87 | 41 | 84 |
| 2022 | 30 | 74 | 13 | 138 |
| 2023 | 30 | 124 | 10 (1) | 159 |
| 2024 | 26 | 104 | 4 | 281 |
| 2025 | 26 | 122 | 8 (5) | 1,263 |
| 2026 | 24 | 126 | 43 | — |
What is it indicated for?
The ONI MSK ExtremeTM MR Scanner is indicated to use as a magnetic resonance imaging device of the leg (excluding the thigh), knee, ankle, foot, elbow, forearm, wrist, and hand. The device produces transverse, sagittal, coronal, and oblique cross-sectional images, displaying the internal structure of the limbs and joints being imaged. If interpreted by a medical expert, these images can provide diagnostically useful information.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K080048 · printed Nov 4, 2026
Cite this page
510(k) Report. “K080048: MSK Extreme MR Scanner, Model AA5000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K080048
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026