510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

Achieva 1.5T & Inera 1.5T Family

K052013510(k) TraditionalCleared Aug 2, 20057-day reviewClass II21 CFR 892.1000Recalls 27 (3 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems Nederland B.V. · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Aug 2, 2005 as substantially equivalent to K043147 (Philips Medical Systems North America Co., 2005) after a 7-day review. Its family tree reaches back 2 generations to 2003. 2 of the 7 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 27 times, including 3 Class I recalls.

Decided
Aug 2, 2005
Review
7 days
Class
Class II
Product code
LNH
Recalls
27 (3 Class I)
Predicates
1

Its longest line of descent runs back 2 generations to K031815 (2003). 4 later devices descend from it, 2 citing it directly. 2 of the 7 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2014: 1 descendant cleared2008: 1 descendant cleared2011: 1 descendant cleared2012: 1 descendant clearedK031815 · Intera-Sensation Family · Philips Medical Systems · 2003 · recalledK043147 · Achieva Family · Philips Medical Systems North America Co. · 2005 · Class I recall200020052010201520202025
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 1 descendant3 generations forward: 1 descendant K043147 · Achieva Family · Philips Medical Systems North America Co. · 2005 · Class I recallK031815 · Intera-Sensation Family · Philips Medical Systems · 2003 · recalled0
  1. Root · 2 generations backK031815 Intera-Sensation FamilyPhilips Medical Systems · 2003 · Recalled
  2. 1 generation backK043147 Achieva FamilyPhilips Medical Systems North America Co. · 2005 · Recalled (Class I)
  3. This deviceK052013 Achieva 1.5T & Inera 1.5T FamilyPhilips Medical Systems Nederland B.V. · 2005
  4. Descendant · cites it directly · 2 of its ownK080048 MSK Extreme MR Scanner, Model AA5000Oni Medical Systems, Inc. · 2008 · Recalled
  5. Descendant · 2 generations on · 1 of its ownK103238 1.5T Optima MR430SGe Medical Systems, LLC · 2011 · Recalled
  6. Descendant · cites it directly · 0 of its ownK131896 Supervan 1.5T Magnetic Resonance Imaging SystemSuzhou Anke Medical System Co., Ltd. · 2014
  7. Descendant · 3 generations on · 0 of its ownK123068 Optima MR 430SGe Medical Systems, LLC · 2012 · Recalled

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K080048MSK Extreme MR Scanner, Model AA5000Oni Medical Systems, Inc. · 2008 · cites it directly2descendants
  • K1032381.5T Optima MR430SGe Medical Systems, LLC · 2011 · 2 generations on1descendant
  • K131896Supervan 1.5T Magnetic Resonance Imaging SystemSuzhou Anke Medical System Co., Ltd. · 2014 · cites it directly0descendants
  • K123068Optima MR 430SGe Medical Systems, LLC · 2012 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K031815Intera-Sensation FamilyPhilips Medical Systems20033
−1K043147Achieva FamilyPhilips Medical Systems North America Co.2005Class I
0 (this device)K052013Achieva 1.5T & Inera 1.5T FamilyPhilips Medical Systems Nederland B.V.2005—
+1K080048MSK Extreme MR Scanner, Model AA5000Oni Medical Systems, Inc.20083
+2K1032381.5T Optima MR430SGe Medical Systems, LLC20112
+1K131896Supervan 1.5T Magnetic Resonance Imaging SystemSuzhou Anke Medical System Co., Ltd.2014—
+3K123068Optima MR 430SGe Medical Systems, LLC20122

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
201110
201210
201410
FIG. 1 — Lineage of K052013, 2003–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K043147Achieva Family
Summary page 1
…e Device(s): The Philips Medical Systems ACHIEVA 1.5T & INTERA 1.5T are the successors of the already cleared (predicate device) ACHIEVA family release 1-series (ref. K043147). Indications for use: The ACHIEVA 1.5T & INTERA 1.5T consists of diagnostic devices that produce cross- sectional images,…
Philips Medical Systems North America Co.2005Class I

Also cleared against K043147: 5 other devices →

What descends from it? (4)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K080048MSK Extreme MR Scanner, Model AA5000Oni Medical Systems, Inc.20082
K131896Supervan 1.5T Magnetic Resonance Imaging SystemSuzhou Anke Medical System Co., Ltd.20140

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 21, 2026Class IIZ-3240-2026Software designA software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Aug 21, 2026Class IIZ-3233-2026Software designA software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Aug 21, 2026Class IIZ-3232-2026Software designA software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Jun 5, 2024Class IZ-2165-2024Device DesignPotential for coils to heat up and harm patients (burn).
Jun 5, 2024Class IZ-2163-2024Device DesignPotential for coils to heat up and harm patients (burn).
Jun 5, 2024Class IZ-2164-2024Device DesignPotential for coils to heat up and harm patients (burn).
Oct 23, 2023Class IIZ-0393-2024Under Investigation by firmPotential component failure in the Gradient Coil could product smoke and/or fire.
Oct 23, 2023Class IIZ-0389-2024Under Investigation by firmPotential component failure in the Gradient Coil could product smoke and/or fire.
Oct 23, 2023Class IIZ-0390-2024Under Investigation by firmPotential component failure in the Gradient Coil could product smoke and/or fire.
Oct 23, 2023Class IIZ-0392-2024Under Investigation by firmPotential component failure in the Gradient Coil could product smoke and/or fire.
Dec 21, 2018Class IIZ-0889-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Dec 21, 2018Class IIZ-0890-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Dec 21, 2018Class IIZ-0892-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Dec 21, 2018Class IIZ-0894-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Dec 21, 2018Class IIZ-0896-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Dec 21, 2018Class IIZ-0909-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Dec 21, 2018Class IIZ-0887-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Dec 21, 2018Class IIZ-0888-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Dec 21, 2018Class IIZ-0891-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Dec 21, 2018Class IIZ-0895-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Dec 21, 2018Class IIZ-0899-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Dec 21, 2018Class IIZ-0915-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Dec 21, 2018Class IIZ-0908-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Dec 21, 2018Class IIZ-0912-2019Unknown/Undetermined by firmThere is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Mar 16, 2018Class IIZ-1435-2018Nonconforming Material/ComponentPotential risk for helium gas inside the MR examination room during a magnet quench
Mar 16, 2018Class IIZ-1433-2018Nonconforming Material/ComponentPotential risk for helium gas inside the MR examination room during a magnet quench
Sep 30, 2008Not classifiedZ-2461-2010OtherThe cables of the coil may become too hot and cause burns when placed too close to the patient. While using the 1.5T Sense XL Torso coil, unwanted radiofrequencies (RF) interaction with the patient can occur. Especially in combination…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K043147Achieva FamilyPhilips Medical Systems North America Co. · 2005 · Direct predicate · ◆ on the line of descent in FIG. 13 Class I recalls

What does this lineage say?

  • The line of descent spans 2 years and passes through 3 companies; a new generation was cleared every 1.4 years on median.
  • 2 of the 2 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K080048 (Oni Medical Systems, Inc., 2008) has gone furthest, with 2 descendants of its own.
  • 5 other devices also cite its predicate K043147.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

ACHIEVA 1.5 & INTERA 1.ST family consists of diagnostic devices that produce cross-sectional images, spectroscopy images and/or spectra in any orientation of the internal structure of the whole body. These images when interpreted by a trained physician, yield information that may assist in diagnosis

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K052013: Achieva 1.5T & Inera 1.5T Family.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K052013

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.