510(k) · Cleared
Recalled · Class I · 2026Achieva 1.5T & Inera 1.5T Family
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Philips Medical Systems Nederland B.V. · LNH System, Nuclear Magnetic Resonance Imaging
Cleared on Aug 2, 2005 as substantially equivalent to K043147 (Philips Medical Systems North America Co., 2005) after a 7-day review. Its family tree reaches back 2 generations to 2003. 2 of the 7 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 27 times, including 3 Class I recalls.
- Decided
- Aug 2, 2005
- Review
- 7 days
- Class
- Class II
- Product code
- LNH
- Recalls
- 27 (3 Class I)
- Predicates
- 1
Its longest line of descent runs back 2 generations to K031815 (2003). 4 later devices descend from it, 2 citing it directly. 2 of the 7 devices in this family, counting this one, have a Class I recall.
- Root · 2 generations backK031815 Intera-Sensation FamilyPhilips Medical Systems · 2003 · Recalled
- 1 generation backK043147 Achieva FamilyPhilips Medical Systems North America Co. · 2005 · Recalled (Class I)
- This deviceK052013 Achieva 1.5T & Inera 1.5T FamilyPhilips Medical Systems Nederland B.V. · 2005
- Descendant · cites it directly · 2 of its ownK080048 MSK Extreme MR Scanner, Model AA5000Oni Medical Systems, Inc. · 2008 · Recalled
- Descendant · 2 generations on · 1 of its ownK103238 1.5T Optima MR430SGe Medical Systems, LLC · 2011 · Recalled
- Descendant · cites it directly · 0 of its ownK131896 Supervan 1.5T Magnetic Resonance Imaging SystemSuzhou Anke Medical System Co., Ltd. · 2014
- Descendant · 3 generations on · 0 of its ownK123068 Optima MR 430SGe Medical Systems, LLC · 2012 · Recalled
Class I recalls in this family
- K043147 Achieva Familydirect predicate · 2005
Most influential descendants
Showing 4 of 4 · See all- K080048MSK Extreme MR Scanner, Model AA5000Oni Medical Systems, Inc. · 2008 · cites it directly2descendants
- K1032381.5T Optima MR430SGe Medical Systems, LLC · 2011 · 2 generations on1descendant
- K131896Supervan 1.5T Magnetic Resonance Imaging SystemSuzhou Anke Medical System Co., Ltd. · 2014 · cites it directly0descendants
- K123068Optima MR 430SGe Medical Systems, LLC · 2012 · 3 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K031815 | Intera-Sensation Family | Philips Medical Systems | 2003 | 3 |
| −1 | K043147 | Achieva Family | Philips Medical Systems North America Co. | 2005 | Class I |
| 0 (this device) | K052013 | Achieva 1.5T & Inera 1.5T Family | Philips Medical Systems Nederland B.V. | 2005 | — |
| +1 | K080048 | MSK Extreme MR Scanner, Model AA5000 | Oni Medical Systems, Inc. | 2008 | 3 |
| +2 | K103238 | 1.5T Optima MR430S | Ge Medical Systems, LLC | 2011 | 2 |
| +1 | K131896 | Supervan 1.5T Magnetic Resonance Imaging System | Suzhou Anke Medical System Co., Ltd. | 2014 | — |
| +3 | K123068 | Optima MR 430S | Ge Medical Systems, LLC | 2012 | 2 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2008 | 1 | 0 |
| 2011 | 1 | 0 |
| 2012 | 1 | 0 |
| 2014 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LNH device
- #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K043147 | Achieva FamilySummary page 1…e Device(s): The Philips Medical Systems ACHIEVA 1.5T & INTERA 1.5T are the successors of the already cleared (predicate device) ACHIEVA family release 1-series (ref. K043147). Indications for use: The ACHIEVA 1.5T & INTERA 1.5T consists of diagnostic devices that produce cross- sectional images,… | 2005 |
What descends from it? (4)
2 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (2 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 21, 2026 | Class II | Z-3240-2026 | Software design | A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. |
| Aug 21, 2026 | Class II | Z-3233-2026 | Software design | A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. |
| Aug 21, 2026 | Class II | Z-3232-2026 | Software design | A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory. |
| Jun 5, 2024 | Class I | Z-2165-2024 | Device Design | Potential for coils to heat up and harm patients (burn). |
| Jun 5, 2024 | Class I | Z-2163-2024 | Device Design | Potential for coils to heat up and harm patients (burn). |
| Jun 5, 2024 | Class I | Z-2164-2024 | Device Design | Potential for coils to heat up and harm patients (burn). |
| Oct 23, 2023 | Class II | Z-0393-2024 | Under Investigation by firm | Potential component failure in the Gradient Coil could product smoke and/or fire. |
| Oct 23, 2023 | Class II | Z-0389-2024 | Under Investigation by firm | Potential component failure in the Gradient Coil could product smoke and/or fire. |
| Oct 23, 2023 | Class II | Z-0390-2024 | Under Investigation by firm | Potential component failure in the Gradient Coil could product smoke and/or fire. |
| Oct 23, 2023 | Class II | Z-0392-2024 | Under Investigation by firm | Potential component failure in the Gradient Coil could product smoke and/or fire. |
| Dec 21, 2018 | Class II | Z-0889-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 21, 2018 | Class II | Z-0890-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 21, 2018 | Class II | Z-0892-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 21, 2018 | Class II | Z-0894-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 21, 2018 | Class II | Z-0896-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 21, 2018 | Class II | Z-0909-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 21, 2018 | Class II | Z-0887-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 21, 2018 | Class II | Z-0888-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 21, 2018 | Class II | Z-0891-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 21, 2018 | Class II | Z-0895-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 21, 2018 | Class II | Z-0899-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 21, 2018 | Class II | Z-0915-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 21, 2018 | Class II | Z-0908-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Dec 21, 2018 | Class II | Z-0912-2019 | Unknown/Undetermined by firm | There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field. |
| Mar 16, 2018 | Class II | Z-1435-2018 | Nonconforming Material/Component | Potential risk for helium gas inside the MR examination room during a magnet quench |
| Mar 16, 2018 | Class II | Z-1433-2018 | Nonconforming Material/Component | Potential risk for helium gas inside the MR examination room during a magnet quench |
| Sep 30, 2008 | Not classified | Z-2461-2010 | Other | The cables of the coil may become too hot and cause burns when placed too close to the patient. While using the 1.5T Sense XL Torso coil, unwanted radiofrequencies (RF) interaction with the patient can occur. Especially in combination… |
The 5 latest of 27 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K043147Achieva FamilyPhilips Medical Systems North America Co. · 2005 · Direct predicate · ◆ on the line of descent in FIG. 13 Class I recalls
What does this lineage say?
- The line of descent spans 2 years and passes through 3 companies; a new generation was cleared every 1.4 years on median.
- 2 of the 2 ancestors have a recall of any class; 1 has a Class I recall.
- Of its descendants, K080048 (Oni Medical Systems, Inc., 2008) has gone furthest, with 2 descendants of its own.
- 5 other devices also cite its predicate K043147.
How does its category compare?
All LNH devicesLNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · LNH
26in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LNH
122in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LNH
8in 2025
Class I (8 since 2007) Class II/III not classified
Adverse event reports per year · LNH
1,263in 2025
MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 35 | 25 | 114 |
| 2008 | 19 | 53 | 12 | 167 |
| 2009 | 17 | 70 | 7 | 194 |
| 2010 | 12 | 94 | 17 | 165 |
| 2011 | 16 | 81 | 11 | 182 |
| 2012 | 22 | 89 | 5 | 170 |
| 2013 | 19 | 108 | 2 | 107 |
| 2014 | 17 | 108 | 13 | 119 |
| 2015 | 19 | 140 | 12 (2) | 113 |
| 2016 | 20 | 108 | 1 | 135 |
| 2017 | 18 | 98 | 0 | 121 |
| 2018 | 29 | 128 | 71 | 158 |
| 2019 | 26 | 62 | 5 | 158 |
| 2020 | 32 | 121 | 4 | 98 |
| 2021 | 27 | 87 | 41 | 84 |
| 2022 | 30 | 74 | 13 | 138 |
| 2023 | 30 | 124 | 10 (1) | 159 |
| 2024 | 26 | 104 | 4 | 281 |
| 2025 | 26 | 122 | 8 (5) | 1,263 |
| 2026 | 24 | 126 | 43 | — |
What is it indicated for?
ACHIEVA 1.5 & INTERA 1.ST family consists of diagnostic devices that produce cross-sectional images, spectroscopy images and/or spectra in any orientation of the internal structure of the whole body. These images when interpreted by a trained physician, yield information that may assist in diagnosis
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K052013 · printed Nov 6, 2026
Cite this page
510(k) Report. “K052013: Achieva 1.5T & Inera 1.5T Family.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K052013