510(k) · Cleared
Recalled · Class I · 2024Mega Soft Reusable Patient Return Electrode
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Megadyne Medical Products, Inc. · GEI Electrosurgical, Cutting & Coagulation & Accessories
Cleared on Dec 16, 2008 as substantially equivalent to K021077 (Megadyne Medical Products, Inc., 2002) and 3 other predicates after a 274-day review. Its family tree reaches back 5 generations to 1982. 6 of the 20 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 3 times, including 3 Class I recalls.
- Decided
- Dec 16, 2008
- Review
- 274 days
- Class
- Class II
- Product code
- GEI
- Recalls
- 3 (3 Class I)
- Predicates
- 4
Its longest line of descent runs back 5 generations to K822572 (1982). 15 ancestors in all, the oldest K801694 (1980). 4 later devices descend from it, 1 citing it directly. 6 of the 20 devices in this family, counting this one, have a Class I recall.
- Root · 5 generations backK822572 Polyhesive Patient Return ElectrodeValleylab, Inc. · 1982
- 4 generations backK972273 TBDMegadyne Medical Products, Inc. · 1997
- 3 generations backK982826 MEGA 2000Megadyne Medical Products, Inc. · 1998 · Recalled (Class I)
- 2 generations backK021077 Mega 2000 Soft Patient Return Electrode PadMegadyne Medical Products, Inc. · 2002 · Recalled (Class I)
- 1 generation backK031285 Mega 2000 Soft Dual Cord Patient Return Electrode PadMegadyne Medical Products, Inc. · 2003 · Recalled (Class I)
- This deviceK080741 Mega Soft Reusable Patient Return ElectrodeMegadyne Medical Products, Inc. · 2008
- Descendant · cites it directly · 3 of its ownK133726 Mega Soft Universal Patient Return ElectrodeMegadyne Medical Products, Inc. · 2014 · Recalled (Class I)
- Descendant · 2 generations on · 0 of its ownK233644 Megadyne Patient Return Electrode Pad - Universal Dual (model 0846)…Megadyne Medical Products, Inc. · 2024 · Recalled (Class I)
- Descendant · 2 generations on · 0 of its ownK210727 HotDog Warming Mattress + Return ElectrodeAugustine Temperature Management, LLC · 2022 · Recalled
- Descendant · 2 generations on · 0 of its ownK183148 OKLand Patient Return Electrode PadAuckland Medical Polymer (Tianjin) Co., Ltd. · 2020
Class I recalls in this family
Most influential descendants
Showing 4 of 4 · See all- K133726Mega Soft Universal Patient Return ElectrodeMegadyne Medical Products, Inc. · 2014 · cites it directly3descendants
- K233644Megadyne Patient Return Electrode Pad - Universal Dual (model 0846), Megadyne Patient…Megadyne Medical Products, Inc. · 2024 · 2 generations on0descendants
- K210727HotDog Warming Mattress + Return ElectrodeAugustine Temperature Management, LLC · 2022 · 2 generations on0descendants
- K183148OKLand Patient Return Electrode PadAuckland Medical Polymer (Tianjin) Co., Ltd. · 2020 · 2 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −5 | K822572 | Polyhesive Patient Return Electrode | Valleylab, Inc. | 1982 | — |
| −4 | K972273 | TBD | Megadyne Medical Products, Inc. | 1997 | — |
| −3 | K982826 | MEGA 2000 | Megadyne Medical Products, Inc. | 1998 | Class I |
| −2 | K021077 | Mega 2000 Soft Patient Return Electrode Pad | Megadyne Medical Products, Inc. | 2002 | Class I |
| −1 | K031285 | Mega 2000 Soft Dual Cord Patient Return Electrode Pad | Megadyne Medical Products, Inc. | 2003 | Class I |
| 0 (this device) | K080741 | Mega Soft Reusable Patient Return Electrode | Megadyne Medical Products, Inc. | 2008 | — |
| +1 | K133726 | Mega Soft Universal Patient Return Electrode | Megadyne Medical Products, Inc. | 2014 | Class I |
| +2 | K233644 | Megadyne Patient Return Electrode Pad - Universal Dual (model 0846), Megadyne… | Megadyne Medical Products, Inc. | 2024 | Class I |
| +2 | K210727 | HotDog Warming Mattress + Return Electrode | Augustine Temperature Management, LLC | 2022 | 2 |
| +2 | K183148 | OKLand Patient Return Electrode Pad | Auckland Medical Polymer (Tianjin) Co., Ltd. | 2020 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1980 | 1 | 0 |
| 1981 | 1 | 0 |
| 1985 | 1 | 0 |
| 1986 | 1 | 0 |
| 1995 | 1 | 0 |
| 1997 | 1 | 0 |
| 1999 | 1 | 0 |
| 2000 | 1 | 0 |
| 2003 | 1 | 0 |
| 2006 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2014 | 1 | 1 |
| 2020 | 1 | 0 |
| 2022 | 1 | 0 |
| 2024 | 1 | 1 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new GEI device
- #1 K231790 The InMode System with the Morpheus8 Applicatorsscore 32.7 · 21 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (4)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K021077 | Mega 2000 Soft Patient Return Electrode PadSummary page 2…ices The predicate devices include: Megadyne's Mega 2000TM Soft Patient Return Electrode Pad which was cleared for marketing via 510(k) # K021077 by FDA's Office of Device Evaluation on April 17, 2002; Megadyne's Mega 2000T Soft Dual Cord Patient Re… | 2002 |
| * | K031285 | Mega 2000 Soft Dual Cord Patient Return Electrode PadSummary page 2…Megadyne's Mega 2000T Soft Dual Cord Patient Return Electrode Pad M which was cleared for marketing via 5 10(k) # K031285 by FDA's Office of Device Evaluation on May 19, 2003; Valleylabs E7512 Neonatal REM Polyhesive II Patie… | 2003 |
| * | K994428 | E7512 Premie Rem Polyhesive II Patient Return ElectrodeSummary page 2…of Device Evaluation on May 19, 2003; Valleylabs E7512 Neonatal REM Polyhesive II Patient Return which was cleared for marketing via 5 10(k) #K994428 by FDA's Office of Device Evaluation on March 7, 2000; and Leonard Lang's Skintact®Cool Contact Electro… | 2000 |
| * | K063161 | Skintact Cool Contact Electrosurgical Grounding Plate with NH 04 GelSummary page 2…7, 2000; and Leonard Lang's Skintact®Cool Contact Electrosurgical Grounding Plates with NH 04 gel which was cleared for marketing via 510(k) # K063161 by FDA's Office of Device Evaluation on November 26, 2006. F. Applicant Device Description The Mega Soft P… | 2006 |
What descends from it? (4)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K133726 | Mega Soft Universal Patient Return Electrode Class I | 2014 | 3 |
Has anything in its family been recalled? (6 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 17, 2024 | Class I | Z-2248-2024 | Unknown/Undetermined by firm | To limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads. |
| May 8, 2024 | Class I | Z-1920-2024 | Device Design | Reports of patient burns. |
| Jun 1, 2023 | Class I | Z-1997-2023 | Under Investigation by firm | Firm has received reports of patient burns in surgical procedures where device was used. |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K031285Mega 2000 Soft Dual Cord Patient Return Electrode PadMegadyne Medical Products, Inc. · 2003 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
- K021077Mega 2000 Soft Patient Return Electrode PadMegadyne Medical Products, Inc. · 2002 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
- K982826MEGA 2000Megadyne Medical Products, Inc. · 1998 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
- K133726Mega Soft Universal Patient Return ElectrodeMegadyne Medical Products, Inc. · 2014 · Cites it directly · ◆ in the 2014 descendants bar of FIG. 15 Class I recalls
- K233644Megadyne Patient Return Electrode Pad - Universal Dual (model 0846), Megadyne Patient Return Electrode Pad -…Megadyne Medical Products, Inc. · 2024 · 2 generations on · ◆ in the 2024 descendants bar of FIG. 13 Class I recalls
What does this lineage say?
- The line of descent spans 26 years and passes through 2 companies; a new generation was cleared every 3.4 years on median.
- 3 of the 15 ancestors have a recall of any class; 3 have a Class I recall.
- Of its descendants, K133726 (Megadyne Medical Products, Inc., 2014) has gone furthest, with 3 descendants of its own.
- 1 other device also cites its predicate K021077.
How does its category compare?
All GEI devicesGEI Electrosurgical, Cutting & Coagulation & Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.
FDA decisions per year · GEI
52in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · GEI
113in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · GEI
4in 2025
Class I (14 since 2007) Class II/III not classified
Adverse event reports per year · GEI
7,592in 2025
MAUDE reports naming product code GEI, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 50 | 51 | 21 | 2,795 |
| 2008 | 61 | 71 | 19 | 2,556 |
| 2009 | 42 | 130 | 27 | 3,591 |
| 2010 | 46 | 109 | 51 | 3,377 |
| 2011 | 35 | 135 | 32 | 4,463 |
| 2012 | 39 | 133 | 30 | 5,216 |
| 2013 | 48 | 136 | 13 | 4,897 |
| 2014 | 49 | 103 | 3 | 4,908 |
| 2015 | 47 | 123 | 4 | 4,936 |
| 2016 | 64 | 108 | 3 | 5,616 |
| 2017 | 52 | 72 | 2 | 6,073 |
| 2018 | 47 | 93 | 6 | 6,545 |
| 2019 | 48 | 93 | 5 | 7,067 |
| 2020 | 59 | 105 | 1 | 6,490 |
| 2021 | 46 | 146 | 6 | 6,354 |
| 2022 | 55 | 150 | 49 | 6,103 |
| 2023 | 65 | 104 | 27 (4) | 7,176 |
| 2024 | 59 | 125 | 22 (10) | 7,897 |
| 2025 | 52 | 113 | 4 | 7,592 |
| 2026 | 43 | 183 | 8 | — |
What is it indicated for?
This device is designed to be used whenever monopolar electrosurgy is indicated. The intended use of this device is to conduct monopolar electrosurgical energy from target tissue of a patient back to one or two electrosurgical units (ESU), or generators. At the same time the product reduces the risk of pressure related injury due to immobility during surgery.
Electrosurgical use is restricted to use with isolated monopolar electrosurgical generators. The device is not intended for RF ablation.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K080741 · printed Nov 3, 2026
Cite this page
510(k) Report. “K080741: Mega Soft Reusable Patient Return Electrode.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K080741
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026