510(k) Report

510(k) · Cleared

Recalled · Class I · 2023

Mega Soft Universal Patient Return Electrode

K133726510(k) SpecialCleared Jan 24, 201449-day reviewClass II21 CFR 878.4400Recalls 5 (5 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Megadyne Medical Products, Inc. · GEI Electrosurgical, Cutting & Coagulation & Accessories

Cleared on Jan 24, 2014 as substantially equivalent to K080741 (Megadyne Medical Products, Inc., 2008) after a 49-day review. Its family tree reaches back 6 generations to 1982. 6 of the 20 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it. It has been recalled 5 times, including 5 Class I recalls.

Decided
Jan 24, 2014
Review
49 days
Class
Class II
Product code
GEI
Recalls
5 (5 Class I)
Predicates
1

Its longest line of descent runs back 6 generations to K822572 (1982). 16 ancestors in all, the oldest K801694 (1980). 3 later devices descend from it, 3 citing it directly. 6 of the 20 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2000: 1 other ancestor2003: 1 other ancestor1995: 1 other ancestor1986: 1 other ancestor1981: 1 other ancestor1980: 1 other ancestor1999: 1 other ancestor1997: 1 other ancestor1985: 1 other ancestor2024: 1 descendant cleared, 1 with a Class I recall12020: 1 descendant cleared2022: 1 descendant clearedK822572 · Polyhesive Patient Return Electrode · Valleylab, Inc. · 1982K972273 · TBD · Megadyne Medical Products, Inc. · 1997K982826 · MEGA 2000 · Megadyne Medical Products, Inc. · 1998 · Class I recallK021077 · Mega 2000 Soft Patient Return Electrode Pad · Megadyne Medical Products, Inc. · 2002 · Class I recallK031285 · Mega 2000 Soft Dual Cord Patient Return Electrode Pad · Megadyne Medical Products, Inc. · 2003 · Class I recallK080741 · Mega Soft Reusable Patient Return Electrode · Megadyne Medical Products, Inc. · 2008 · Class I recall19801990200020102020
  1. Root · 6 generations backK822572 Polyhesive Patient Return ElectrodeValleylab, Inc. · 1982
  2. 5 generations backK972273 TBDMegadyne Medical Products, Inc. · 1997
  3. 4 generations backK982826 MEGA 2000Megadyne Medical Products, Inc. · 1998 · Recalled (Class I)
  4. 3 generations backK021077 Mega 2000 Soft Patient Return Electrode PadMegadyne Medical Products, Inc. · 2002 · Recalled (Class I)
  5. 2 generations backK031285 Mega 2000 Soft Dual Cord Patient Return Electrode PadMegadyne Medical Products, Inc. · 2003 · Recalled (Class I)
  6. 1 generation backK080741 Mega Soft Reusable Patient Return ElectrodeMegadyne Medical Products, Inc. · 2008 · Recalled (Class I)
  7. This deviceK133726 Mega Soft Universal Patient Return ElectrodeMegadyne Medical Products, Inc. · 2014
  8. Descendant · cites it directly · 0 of its ownK233644 Megadyne Patient Return Electrode Pad - Universal Dual (model 0846)…Megadyne Medical Products, Inc. · 2024 · Recalled (Class I)
  9. Descendant · cites it directly · 0 of its ownK210727 HotDog Warming Mattress + Return ElectrodeAugustine Temperature Management, LLC · 2022 · Recalled
  10. Descendant · cites it directly · 0 of its ownK183148 OKLand Patient Return Electrode PadAuckland Medical Polymer (Tianjin) Co., Ltd. · 2020

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K233644Megadyne Patient Return Electrode Pad - Universal Dual (model 0846), Megadyne Patient…Megadyne Medical Products, Inc. · 2024 · cites it directly0descendants
  • K210727HotDog Warming Mattress + Return ElectrodeAugustine Temperature Management, LLC · 2022 · cites it directly0descendants
  • K183148OKLand Patient Return Electrode PadAuckland Medical Polymer (Tianjin) Co., Ltd. · 2020 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K822572Polyhesive Patient Return ElectrodeValleylab, Inc.1982—
−5K972273TBDMegadyne Medical Products, Inc.1997—
−4K982826MEGA 2000Megadyne Medical Products, Inc.1998Class I
−3K021077Mega 2000 Soft Patient Return Electrode PadMegadyne Medical Products, Inc.2002Class I
−2K031285Mega 2000 Soft Dual Cord Patient Return Electrode PadMegadyne Medical Products, Inc.2003Class I
−1K080741Mega Soft Reusable Patient Return ElectrodeMegadyne Medical Products, Inc.2008Class I
0 (this device)K133726Mega Soft Universal Patient Return ElectrodeMegadyne Medical Products, Inc.2014—
+1K233644Megadyne Patient Return Electrode Pad - Universal Dual (model 0846), Megadyne…Megadyne Medical Products, Inc.2024Class I
+1K210727HotDog Warming Mattress + Return ElectrodeAugustine Temperature Management, LLC20222
+1K183148OKLand Patient Return Electrode PadAuckland Medical Polymer (Tianjin) Co., Ltd.2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
198110
198510
198610
199510
199710
199910
200010
200310
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
202210
202411
FIG. 1 — Lineage of K133726, 1982–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GEI device

  1. #1 K231790 The InMode System with the Morpheus8 Applicatorsscore 32.7 · 21 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K080741Mega Soft Reusable Patient Return Electrode
Summary page 2
…Page 2 of 3 . Predicate Devices The predicate device is Megadyne's Mega Soft Return Electrode Pad which was cleared for marketing via 51 0(k) #tK080741 by FDA's Office ofDevice Evaluation on December 12; -2008 F. Applicant Device Description The MegaSoff Universal…
Megadyne Medical Products, Inc.2008Class I

What descends from it? (3)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K233644Megadyne Patient Return Electrode Pad - Universal Dual (model 0846), Megadyne Patient… Class IMegadyne Medical Products, Inc.20240
K210727HotDog Warming Mattress + Return ElectrodeAugustine Temperature Management, LLC20220
K183148OKLand Patient Return Electrode PadAuckland Medical Polymer (Tianjin) Co., Ltd.20200

Has anything in its family been recalled? (6 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 8, 2023Class IZ-0597-2024Under Investigation by firmAffected product codes are now limited to use in patients age 12 years or older. Megadyne has received reports of patient burn injuries up to and including third-degree burns requiring intervention which may lead to prolonged hospital…
Jun 1, 2023Class IZ-2000-2023Under Investigation by firmFirm has received reports of patient burns in surgical procedures where device was used.
Jun 1, 2023Class IZ-2001-2023Under Investigation by firmFirm has received reports of patient burns in surgical procedures where device was used.
Jun 1, 2023Class IZ-1999-2023Under Investigation by firmFirm has received reports of patient burns in surgical procedures where device was used.
Jun 1, 2023Class IZ-1998-2023Under Investigation by firmFirm has received reports of patient burns in surgical procedures where device was used.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K080741Mega Soft Reusable Patient Return ElectrodeMegadyne Medical Products, Inc. · 2008 · Direct predicate · ◆ on the line of descent in FIG. 13 Class I recalls
  • K031285Mega 2000 Soft Dual Cord Patient Return Electrode PadMegadyne Medical Products, Inc. · 2003 · 2 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
  • K021077Mega 2000 Soft Patient Return Electrode PadMegadyne Medical Products, Inc. · 2002 · 2 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
  • K982826MEGA 2000Megadyne Medical Products, Inc. · 1998 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K233644Megadyne Patient Return Electrode Pad - Universal Dual (model 0846), Megadyne Patient Return Electrode Pad -…Megadyne Medical Products, Inc. · 2024 · Cites it directly · ◆ in the 2024 descendants bar of FIG. 13 Class I recalls

What does this lineage say?

  • The line of descent spans 32 years and passes through 2 companies; a new generation was cleared every 5.1 years on median.
  • 4 of the 16 ancestors have a recall of any class; 4 have a Class I recall.
  • Of its descendants, K233644 (Megadyne Medical Products, Inc., 2024) has gone furthest, with 0 descendants of its own.

How does its category compare?

All GEI devices

GEI Electrosurgical, Cutting & Coagulation & Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · GEI

52in 2025

2007: 502008: 612009: 422010: 462011: 352012: 392013: 482014: 492015: 472016: 642017: 522018: 472019: 482020: 592021: 462022: 552023: 652024: 592025: 522026 (partial year): 43
2007high 65 (2023) · low 35 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GEI

113in 2025

2007: 51 days median2008: 71 days median2009: 130 days median2010: 109 days median2011: 135 days median2012: 133 days median2013: 136 days median2014: 103 days median2015: 123 days median2016: 108 days median2017: 72 days median2018: 93 days median2019: 93 days median2020: 105 days median2021: 146 days median2022: 150 days median2023: 104 days median2024: 125 days median2025: 113 days median2026 (partial year): 183 days median
2007high 183 (2026) · low 51 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GEI

4in 2025

2007: 212008: 192009: 272010: 512011: 322012: 302013: 132014: 32015: 42016: 32017: 22018: 62019: 52020: 12021: 62022: 492023: 27, 4 Class I2024: 22, 10 Class I2025: 42026 (partial year): 8
2007high 51 (2010) · low 1 (2020)2026

Class I (14 since 2007) Class II/III not classified

Adverse event reports per year · GEI

7,592in 2025

2007: 2,7952008: 2,5562009: 3,5912010: 3,3772011: 4,4632012: 5,2162013: 4,8972014: 4,9082015: 4,9362016: 5,6162017: 6,0732018: 6,5452019: 7,0672020: 6,4902021: 6,3542022: 6,1032023: 7,1762024: 7,8972025: 7,592
2007high 7,897 (2024) · low 2,556 (2008)2025

MAUDE reports naming product code GEI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075051212,795
20086171192,556
200942130273,591
201046109513,377
201135135324,463
201239133305,216
201348136134,897
20144910334,908
20154712344,936
20166410835,616
2017527226,073
2018479366,545
2019489357,067
20205910516,490
20214614666,354
202255150496,103
20236510427 (4)7,176
20245912522 (10)7,897
20255211347,592
2026431838—

What is it indicated for?

This device is designed to be used whenever monopolar electrosurgery is indicated. The intended use of this device is to conduct monopolar electrosurgical energy from target tissue of a patient back to one or two electrosurgical units (ESU), or generators. Electrosurgical use is restricted to use with isolated monopolar electrosurgical generators. The device is not intended for RF ablation.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K133726: Mega Soft Universal Patient Return Electrode.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K133726

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026