510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

Innercool Therapies, Inc., Rapidblue System

K080908510(k) SpecialCleared Oct 17, 2008199-day reviewClass II21 CFR 870.5900Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Innercool Therapies, Inc.- A Delaware Corporation · NCX System, Hypothermia, Intravenous, Cooling

Cleared on Oct 17, 2008 as substantially equivalent to K033623 (Innercool Therapies, Inc., 2004) after a 199-day review. Its family tree reaches back 5 generations to 1981. 3 of the 27 devices in its family, counting itself, have a Class I recall. 1 later device descends from it. It has been recalled once.

Decided
Oct 17, 2008
Review
199 days
Class
Class II
Product code
NCX
Recalls
1
Predicates
1

Its longest line of descent runs back 5 generations to K811742 (1981). 25 ancestors in all, the oldest K811743 (1981). 1 later device descends from it, 1 citing it directly. 3 of the 27 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor2001: 1 other ancestor2000: 5 other ancestors1997: 2 other ancestors, 1 with a Class I recall11993: 1 other ancestor1991: 2 other ancestors1989: 2 other ancestors1990: 2 other ancestors, 1 with a Class I recall11986: 1 other ancestor, 1 with a Class I recall11985: 1 other ancestor1982: 1 other ancestor1981: 1 other ancestor2016: 1 descendant clearedK811742 · Hemotherm Model 400 · Cincinnati Sub-Zero · 1981 · recalledK012512 · Setpoint Endovascular Temperature Management System · Radiant Medical · 2002K022366 · Celsius Control System · Innercool Therapies · 2003K030421 · Alsius Coolgard 3000 & Alsius Catheter System · Alsius Corp. · 2003 · recalledK033623 · Celsius Control System · Innercool Therapies, Inc. · 200419801990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 10 other ancestors, 1 with a Class I recall14 generations back: 9 other ancestors, 2 with a Class I recall2 1 generation forward: 1 descendant K033623 · Celsius Control System · Innercool Therapies, Inc. · 2004K030421 · Alsius Coolgard 3000 & Alsius Catheter System · Alsius Corp. · 2003 · recalledK022366 · Celsius Control System · Innercool Therapies · 2003K012512 · Setpoint Endovascular Temperature Management System · Radiant Medical · 2002K811742 · Hemotherm Model 400 · Cincinnati Sub-Zero · 1981 · recalled−50
  1. Root · 5 generations backK811742 Hemotherm Model 400Cincinnati Sub-Zero · 1981 · Recalled
  2. 4 generations backK012512 Setpoint Endovascular Temperature Management SystemRadiant Medical · 2002
  3. 3 generations backK022366 Celsius Control SystemInnercool Therapies · 2003
  4. 2 generations backK030421 Alsius Coolgard 3000 & Alsius Catheter SystemAlsius Corp. · 2003 · Recalled
  5. 1 generation backK033623 Celsius Control SystemInnercool Therapies, Inc. · 2004
  6. This deviceK080908 Innercool Therapies, Inc., Rapidblue SystemInnercool Therapies, Inc.- A Delaware Corporation · 2008
  7. Descendant · cites it directly · 0 of its ownK152946 TempedySeiratherm GmbH · 2016

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K152946TempedySeiratherm GmbH · 2016 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K811742Hemotherm Model 400Cincinnati Sub-Zero19812
−4K012512Setpoint Endovascular Temperature Management SystemRadiant Medical2002—
−3K022366Celsius Control SystemInnercool Therapies2003—
−2K030421Alsius Coolgard 3000 & Alsius Catheter SystemAlsius Corp.20034
−1K033623Celsius Control SystemInnercool Therapies, Inc.2004—
0 (this device)K080908Innercool Therapies, Inc., Rapidblue SystemInnercool Therapies, Inc.- A Delaware Corporation2008—
+1K152946TempedySeiratherm GmbH2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198210
198510
198611
198920
199021
199120
199310
199721
200050
200110
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
FIG. 1 — Lineage of K080908, 1981–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NCX device

  1. #1 K213031 Solex 7 Intravascular Heat Exchange Catheter…score 11.0 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K033623Celsius Control System
Summary page 1
…mal Regulating System Predicate Devices The subject device is substantially equivalent, in whole or in part, to predicate devices manufactured by INNERCOOL therapies (K033623). Device Description The subject device is a thermal regulating system consisting of three (3) parts: 1. A console contain…
Innercool Therapies, Inc.2004—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K152946TempedySeiratherm GmbH20160

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 13, 2014Class IIZ-0845-2014Nonconforming Material/ComponentPhillips Healthcare initiated this action because the Main Control Board (MCB) may produce an inaccurate low patient temperature reading and low temperature alarm.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K970864Arrow-Howes Large Bore Multi-Lumen Central Venous CathetersArrow Intl., Inc. · 1997 · 3 generations back · ◆ in the 1997 ancestors bar of FIG. 11 Class I recall
  • K895417Arrow Large-Bore Dual Lumen Hemodialysis KitArrow Intl., Inc. · 1990 · 4 generations back · ◆ in the 1990 ancestors bar of FIG. 11 Class I recall
  • K862056[Central Venous][Multi-Lumen] Catheter W/Flex TipArrow Intl., Inc. · 1986 · 4 generations back · ◆ in the 1986 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 27 years and passes through 6 companies; a new generation was cleared every 9 months on median.
  • 13 of the 25 ancestors have a recall of any class; 3 have a Class I recall.
  • Of its descendants, K152946 (Seiratherm GmbH, 2016) has gone furthest, with 0 descendants of its own.

How does its category compare?

All NCX devices

NCX System, Hypothermia, Intravenous, Cooling: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · NCX

1in 2023

2007: 62008: 32010: 12014: 12015: 12016: 32021: 12022: 12023: 1
2007high 6 (2007) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NCX

204in 2016

2007: 28 days median2008: 129 days median2016: 204 days median
2007high 204 (2016) · low 28 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NCX

28in 2021

2007: 22014: 12021: 28
2007high 28 (2021) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NCX

273in 2025

2007: 42008: 82009: 42010: 92011: 72012: 142013: 312014: 412015: 322016: 3112017: 4502018: 5092019: 4362020: 4152021: 4612022: 5112023: 3732024: 3432025: 273
2007high 511 (2022) · low 4 (2007)2025

MAUDE reports naming product code NCX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200762824
2008312908
20090—04
20101—09
20110—07
20120—014
20130—031
20141—141
20151—032
201632040311
20170—0450
20180—0509
20190—0436
20200—0415
20211—28461
20221—0511
20231—0373
20240—0343
20250—0273
20260—0—

What is it indicated for?

The RapidBlue TM System is a thermal regulating system intended to induce, maintain and reverse mild hypothermia in neurosurgical patients in surgery and in recovery/intensive care, to achieve and/or maintain normothermia in cardiac surgery patients in surgery and in recovery/intensive care, and for use in fever reduction, as an adjunct to other antipyretic therapy, in patients with cerebral infarction and intracerebral hemorrhage who required access to the central venous circulation and who are intubated and sedated.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K080908: Innercool Therapies, Inc., Rapidblue System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K080908

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026