510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Modification to Automated Breast Ultrasound System, Model Abus

K080930510(k) TraditionalCleared Aug 7, 2008127-day reviewClass II21 CFR 892.1560Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

U-System, Inc. · IYO System, Imaging, Pulsed Echo, Ultrasonic

Cleared on Aug 7, 2008 as substantially equivalent to K052355 (U-Systems, Inc., 2005) and 1 other predicate after a 127-day review. Its family tree reaches back 4 generations to 1992. 3 later devices descend from it. It has been recalled once.

Decided
Aug 7, 2008
Review
127 days
Class
Class II
Product code
IYO
Recalls
1
Predicates
2

Its longest line of descent runs back 4 generations to K922470 (1992). 9 ancestors in all. 3 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 2 other ancestors2003: 1 other ancestor2000: 2 other ancestors2009: 1 descendant cleared2016: 1 descendant cleared2018: 1 descendant clearedK922470 · Ies, Image Enhancement System · Open Mri, Inc. · 1992K993802 · Sharpview Image Enhancement System · Context Vision AB · 2000K024028 · Sharp View Image Enhancement System · Context Vision AB · 2003K052355 · Automated Breast Ultrasound System, Model Abus · U-Systems, Inc. · 20051990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors3 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant K052355 · Automated Breast Ultrasound System, Model Abus · U-Systems, Inc. · 2005K024028 · Sharp View Image Enhancement System · Context Vision AB · 2003K993802 · Sharpview Image Enhancement System · Context Vision AB · 2000K922470 · Ies, Image Enhancement System · Open Mri, Inc. · 19920
  1. Root · 4 generations backK922470 Ies, Image Enhancement SystemOpen Mri, Inc. · 1992
  2. 3 generations backK993802 Sharpview Image Enhancement SystemContext Vision AB · 2000
  3. 2 generations backK024028 Sharp View Image Enhancement SystemContext Vision AB · 2003
  4. 1 generation backK052355 Automated Breast Ultrasound System, Model AbusU-Systems, Inc. · 2005
  5. This deviceK080930 Modification to Automated Breast Ultrasound System, Model AbusU-System, Inc. · 2008
  6. Descendant · cites it directly · 2 of its ownK092067 Abvs WorkplaceSiemens AG Medical Solutions · 2009
  7. Descendant · 2 generations on · 1 of its ownK151075 BR-ABVS Viewer 1.0Taihao Medical, Inc. · 2016
  8. Descendant · 3 generations on · 0 of its ownK180641 Invenia ABUS ViewerGE Healthcare · 2018

Most influential descendants

Showing 3 of 3 · See all
  • K092067Abvs WorkplaceSiemens AG Medical Solutions · 2009 · cites it directly2descendants
  • K151075BR-ABVS Viewer 1.0Taihao Medical, Inc. · 2016 · 2 generations on1descendant
  • K180641Invenia ABUS ViewerGE Healthcare · 2018 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K922470Ies, Image Enhancement SystemOpen Mri, Inc.1992—
−3K993802Sharpview Image Enhancement SystemContext Vision AB2000—
−2K024028Sharp View Image Enhancement SystemContext Vision AB2003—
−1K052355Automated Breast Ultrasound System, Model AbusU-Systems, Inc.2005—
0 (this device)K080930Modification to Automated Breast Ultrasound System, Model AbusU-System, Inc.2008—
+1K092067Abvs WorkplaceSiemens AG Medical Solutions2009—
+2K151075BR-ABVS Viewer 1.0Taihao Medical, Inc.2016—
+3K180641Invenia ABUS ViewerGE Healthcare2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200020
200220
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201610
201810
FIG. 1 — Lineage of K080930, 1992–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYO device

  1. #1 K243937 Accuro 3Sscore 20.5 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K052355Automated Breast Ultrasound System, Model Abus
Summary page 1
…ems.com A. Legally Marketed Predicate Device The Company believes that the modified ABUS is substantially equivalent to the previously cleared U-Systems - ABUS (K052355), as well as the Siemens - Antares DUS (K023720). The intended use and the technological characteristics of the device are…
U-Systems, Inc.2005—
*K023720Sonoline Antares Ultrasound System
Summary page 1
…The Company believes that the modified ABUS is substantially equivalent to the previously cleared U-Systems - ABUS (K052355), as well as the Siemens - Antares DUS (K023720). The intended use and the technological characteristics of the device are the same as the predicate devices. B. Device D…
Siemens Medical Solutions USA, Inc.20021

Also cleared against K052355: 4 other devices →

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K092067Abvs WorkplaceSiemens AG Medical Solutions20092

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 16, 2013Class IIZ-1587-2013Software designDevices labeled for an intended use not included in the existing 510(k) or PMA, in that, the user manual contains the Indication for Use of the soma v Platinum ABUS [a 510(k) product].

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 4 companies; a new generation was cleared every 3.0 years on median.
  • 2 of the 9 ancestors have a recall of any class.
  • Of its descendants, K092067 (Siemens AG Medical Solutions, 2009) has gone furthest, with 2 descendants of its own.
  • 4 other devices also cite its predicate K052355.

How does its category compare?

All IYO devices

IYO System, Imaging, Pulsed Echo, Ultrasonic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · IYO

7in 2025

2007: 202008: 142009: 132010: 152011: 112012: 152013: 172014: 142015: 192016: 102017: 192018: 142019: 132020: 112021: 152022: 52023: 112024: 102025: 72026 (partial year): 10
2007high 20 (2007) · low 5 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYO

171in 2025

2007: 17 days median2008: 36 days median2009: 79 days median2010: 113 days median2011: 69 days median2012: 68 days median2013: 50 days median2014: 63 days median2015: 46 days median2016: 36 days median2017: 72 days median2018: 56 days median2019: 59 days median2020: 57 days median2021: 64 days median2022: 190 days median2023: 177 days median2024: 220 days median2025: 171 days median2026 (partial year): 139 days median
2007high 220 (2024) · low 17 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYO

1in 2022

2007: 42008: 52009: 92010: 82012: 22013: 32014: 22015: 12018: 12019: 12021: 112022: 12026 (partial year): 1
2007high 11 (2021) · low 0 (2011)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYO

485in 2025

2007: 392008: 202009: 272010: 202011: 582012: 782013: 4362014: 3042015: 2682016: 2542017: 2382018: 3682019: 5372020: 4982021: 6172022: 5112023: 5272024: 6402025: 485
2007high 640 (2024) · low 20 (2008)2025

MAUDE reports naming product code IYO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072017439
20081436520
20091379927
201015113820
20111169058
20121568278
201317503436
201414632304
201519461268
201610360254
201719720238
201814561368
201913591537
202011570498
2021156411617
202251901511
2023111770527
2024102200640
202571710485
2026101391—

What is it indicated for?

An ultrasound pulsed echo imaging system is intended to project a pulsed sound beam into body tissue to determine the depth of location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts and accessories. The device is indicated for use…

Read the full statement (114 words)

An ultrasound pulsed echo imaging system is intended to project a pulsed sound beam into body tissue to determine the depth of location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts and accessories.

The device is indicated for use as an adjunct to mammography for B-mode ultrasonic imaging of a patient's breast when used with an automatic scanning linear array transducer or a handheld transducer. The device is not intended to be used as a replacement for screening mammography.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K080930: Modification to Automated Breast Ultrasound System, Model Abus.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K080930

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026