510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Universal Vial Access Spike

K080989510(k) SpecialCleared May 2, 200825-day reviewClass II21 CFR 880.5860Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Icu Medical, Inc. · FMF Syringe, Piston

Cleared on May 2, 2008 as substantially equivalent to K934561 (Icu Medical, Inc., 1994) and 1 other predicate after a 25-day review. Its family tree reaches back 1 generation to 1994. 2 of the 16 devices in its family, counting itself, have a Class I recall. 13 later devices descend from it. It has been recalled once.

Decided
May 2, 2008
Review
25 days
Class
Class II
Product code
FMF
Recalls
1
Predicates
2

Its longest line of descent runs back 1 generation to K934561 (1994). 2 ancestors in all. 13 later devices descend from it, 1 citing it directly. 2 of the 16 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1994: 1 other ancestor2008: 1 descendant cleared, 1 with a Class I recall12018: 2 descendants cleared2013: 1 descendant cleared, 1 with a Class I recall12015: 1 descendant cleared2017: 1 descendant cleared2020: 1 descendant cleared2021: 2 descendants cleared2023: 1 descendant cleared2025: 2 descendants cleared2024: 1 descendant clearedK934561 · One Time Vial Access Spike · Icu Medical, Inc. · 19941990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 2 descendants, 1 with a Class I recall13 generations forward: 1 descendant4 generations forward: 4 descendants5 generations forward: 5 descendants K934561 · One Time Vial Access Spike · Icu Medical, Inc. · 19940+5
  1. Root · 1 generation backK934561 One Time Vial Access SpikeIcu Medical, Inc. · 1994
  2. This deviceK080989 Universal Vial Access SpikeIcu Medical, Inc. · 2008
  3. Descendant · cites it directly · 12 of its ownK081361 Chemoclave Cytotoxic Medication Preparation and Delivery System…Icu Medical, Inc. · 2008 · Recalled (Class I)
  4. Descendant · 2 generations on · 10 of its ownK131549 Chemolock, Chemolock PortIcu Medical, Inc. · 2013 · Recalled (Class I)
  5. Descendant · 3 generations on · 9 of its ownK141448 Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2015
  6. Descendant · 4 generations on · 3 of its ownK192866 Chemfort CSTDSimplivia Healthcare, Ltd. · 2020 · Recalled
  7. Descendant · 4 generations on · 1 of its ownK201422 Arisure Closed System Drug Transfer Device (CSTD)Yukon Medical, LLC · 2021

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2008: 12013: 12015: 12017: 12018: 22020: 12021: 22023: 12024: 12025: 2
2008high 2 (2018) · low 0 (2009)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 13 · See all
  • K081361Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, CH-XXXXIcu Medical, Inc. · 2008 · cites it directly12descendants
  • K131549Chemolock, Chemolock PortIcu Medical, Inc. · 2013 · 2 generations on10descendants
  • K141448Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2015 · 3 generations on9descendants
  • K192866Chemfort CSTDSimplivia Healthcare, Ltd. · 2020 · 4 generations on3descendants
  • K201422Arisure Closed System Drug Transfer Device (CSTD)Yukon Medical, LLC · 2021 · 4 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K934561One Time Vial Access SpikeIcu Medical, Inc.1994—
0 (this device)K080989Universal Vial Access SpikeIcu Medical, Inc.2008—
+1K081361Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, CH-XXXXIcu Medical, Inc.2008Class I
+2K131549Chemolock, Chemolock PortIcu Medical, Inc.2013Class I
+3K141448Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant2015—
+4K192866Chemfort CSTDSimplivia Healthcare, Ltd.20201
+4K201422Arisure Closed System Drug Transfer Device (CSTD)Yukon Medical, LLC2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200811
201311
201510
201710
201820
202010
202120
202310
202410
202520
FIG. 1 — Lineage of K080989, 1994–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FMF device

  1. #1 K241821 Luer Lock Syringe with Safety Needle; Luer…score 16.2 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K934561One Time Vial Access Spike
Summary page 1
…on Name: Piston Syringe, Accessory 21 CFR 880.5860, Class II Device Product Code: FMF Legally Marketed Predicate Devices for Substantial Equivalence: *K934561 - One Time Vial Access - ICU Medical, Inc. *K934591 - Clave Vial Access Spike - ICU Medical, Inc. Rationale for SE: These combo…
Icu Medical, Inc.1994—
*K934591Clave Vial Access Spike
Summary page 1
…860, Class II Device Product Code: FMF Legally Marketed Predicate Devices for Substantial Equivalence: *K934561 - One Time Vial Access - ICU Medical, Inc. *K934591 - Clave Vial Access Spike - ICU Medical, Inc. Rationale for SE: These combo device is the vial access spike and the proprietary…
Icu Medical, Inc.1994—

Also cleared against K934561: 2 other devices →

What descends from it? (13)

See all 13

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K081361Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, CH-XXXX Class IIcu Medical, Inc.200812

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 31, 2020Class IIZ-0710-2021Process controlPotential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K081361Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, CH-XXXXIcu Medical, Inc. · 2008 · Cites it directly · ◆ in the 2008 descendants bar of FIG. 12 Class I recalls
  • K131549Chemolock, Chemolock PortIcu Medical, Inc. · 2013 · 2 generations on · ◆ in the 2013 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • None of the 2 ancestors has a recall.
  • Of its descendants, K081361 (Icu Medical, Inc., 2008) has gone furthest, with 12 descendants of its own.
  • 2 other devices also cite its predicate K934561.

How does its category compare?

All FMF devices

FMF Syringe, Piston: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · FMF

13in 2025

2007: 232008: 92009: 142010: 112011: 162012: 92013: 122014: 132015: 112016: 122017: 132018: 82019: 112020: 152021: 202022: 242023: 222024: 142025: 132026 (partial year): 9
2007high 24 (2022) · low 8 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FMF

114in 2025

2007: 90 days median2008: 74 days median2009: 94 days median2010: 130 days median2011: 109 days median2012: 86 days median2013: 154 days median2014: 84 days median2015: 87 days median2016: 146 days median2017: 135 days median2018: 87 days median2019: 146 days median2020: 107 days median2021: 175 days median2022: 211 days median2023: 173 days median2024: 130 days median2025: 114 days median2026 (partial year): 239 days median
2007high 239 (2026) · low 74 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FMF

19in 2023

2007: 62008: 112009: 152010: 122011: 32012: 62013: 12014: 12015: 22017: 32018: 22019: 12020: 22021: 12022: 22023: 19, 18 Class I2026 (partial year): 2, 1 Class I
2007high 19 (2023) · low 0 (2016)2026

Class I (19 since 2007) Class II/III not classified

Adverse event reports per year · FMF

4,181in 2025

2007: 6742008: 6912009: 9132010: 6792011: 1862012: 3072013: 3262014: 2442015: 2732016: 6842017: 1,8582018: 3,9522019: 5,8632020: 5,3342021: 5,8742022: 5,2332023: 5,2142024: 5,1222025: 4,181
2007high 5,874 (2021) · low 186 (2011)2025

MAUDE reports naming product code FMF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200723906674
200897411691
2009149415913
20101113012679
2011161093186
20129866307
2013121541326
201413841244
201511872273
2016121460684
20171313531,858
201888723,952
20191114615,863
20201510725,334
20212017515,874
20222421125,233
20232217319 (18)5,214
20241413005,122
20251311404,181
202692392 (1)—

What is it indicated for?

Accessory to a piston syringe for needleless access to sterile drug vials without the use of needles.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K080989: Universal Vial Access Spike.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K080989

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.